Medos International Sàrl · Peneterating Grasper - Straight · Class 1

Quick summary

This Medos straight penetrating arthroscopic grasper is a surgical instrument used to grasp and pass tissue, such as during rotator cuff or meniscus repair. The 43 reports mostly describe the instrument failing to open or close properly, or breaking at the tip during surgery, with surgeons switching to a spare instrument. No deaths and no injuries were reported. There are no recalls naming this device, and reports have been trending down in recent years.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 10886705003839), received between Jul 19, 2018 and Aug 12, 2026 — classified as 42 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close27
  • Break11
  • Material Deformation4
  • Material Twisted/Bent2
  • Degraded1
  • Mechanical Jam1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Consequences Or Impact To Patient21

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2026Break; Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50224400 ↗
  • MalfunctionFeb 20, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THAT THE ARTHRO GRASPER PEN STR DEVICE JAWS FAILED TO STAY CLOSED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405374 ↗
  • MalfunctionSep 5, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE ARTHRO GRASPER PEN STR EA DEVICE WAS BROKEN. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22982932 ↗
  • MalfunctionJul 11, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE ARTHRO GRASPER PEN STR DEVICE WOULD NO LONGER OPEN OR CLOSE. ANOTHER DEVICE HAD TO BE OPENED TO FINISH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22480098 ↗
  • MalfunctionJun 9, 2025Difficult to Open or Close; Mechanical Jam; Difficult to Open or Close; Mechanical Jam

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS FOUND THAT THE ARTHRO GRASPER PEN STR *EA DEVICE HAD A STUCK CAPTURE POINT; WHEN THE TRIGGER WAS PULLED, THE GRASPER DID NOT MOVE IN AND OUT CONSISTENTLY AND FAILED TO CLOSE COMPLETELY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168168 ↗
  • MalfunctionFeb 20, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS DISCOVERED THAT THE ARTHRO GRASPER PEN STR DEVICE WAS MALFUNCTIONING AND WOULD NOT CLOSE PROPERLY. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21420550 ↗
  • MalfunctionDec 27, 2024Degraded; Material Twisted/Bent; Degraded; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE ARTHRO GRASPER PEN STR DEVICE BLUE HOOK NOT PASSING THROUGH TENDON. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE TIP WAS DEFORMED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21025845 ↗
  • MalfunctionFeb 1, 2024Break; Difficult to Open or Close; Break; Difficult to Open or Close

    IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER SCOPE PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT WAS BROKEN; AND THAT THE CAPTURE POINT OF THE BIRD BEAK WOULD NOT CLOSE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18622977 ↗
  • MalfunctionAug 24, 2023Break; Difficult to Open or Close; Break; Difficult to Open or Close

    IT WAS REPORTED BY SALES REP THAT AT THE END OF A SHOULDER BANKART REPAIR PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE BROKE AND WAS UNABLE TO OPEN/CLOSE PROPERLY. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17615565 ↗
  • MalfunctionAug 3, 2023Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE SALES REP THAT DURING SHOULDER ARTHROSCOPY PROCEDURE ON (B)(6)2023, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE WAS NOT CLOSING; AND THAT IT SEEMED LIKE THE HANDLE MECHANISM HAD DISENGAGED AND HAD NO TENSION. A DIFFERENT ANGLE WAS USED TO COMPLETE PROCEDURE. THERE WERE NO SURGICAL DELAY NOR PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17452784 ↗
  • MalfunctionJul 10, 2023Break; Break

    IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE SPRING ON THE PENETRATING GRASPER STRAIGHT DEVICE BROKE; AND THEREFORE, WAS NOT CLOSING. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17291089 ↗
  • MalfunctionApr 19, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER SCOPE PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE WAS FOUND BENT AND NOT FLUSHED WITH THE NOSE OF THE DEVICE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16772934 ↗
  • MalfunctionNov 21, 2022Material Deformation; Material Deformation

    IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE WAS BENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCES OR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15837444 ↗
  • MalfunctionNov 10, 2022Material Deformation; Material Deformation

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT DURING IN THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE SURGEON BENT TWO PENETRATING GRASPER DEVICES WHILE TRYING TO REPAIR THE LABRUM. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15775250 ↗
  • MalfunctionNov 1, 2022Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE AFFILIATE IN JAPAN THAT DURING AN UNKNOWN SURGERY ON (B)(6)2022 , IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE WAS DIFFICULT TO OPEN OR CLOSE; AND THEREFORE, UNABLE TO USE. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15708955 ↗
  • MalfunctionAug 3, 2022Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE JAW ON THE PENETRATING GRASPER STRAIGHT DEVICE WAS NOT OPENING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15156796 ↗
  • MalfunctionApr 22, 2022Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE AFFILIATE IN COLOMBIA THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE TIP OPENING MECHANISM ON THE ARTHRO GRASPER PEN STR DEVICE WAS NOT WORKING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS PIECE OF INSTRUMENTATION IS BAD BECAUSE THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14187227 ↗
  • MalfunctionApr 13, 2022Material Deformation; Material Deformation

    IT WAS REPORTED BY THE SALES REP THAT DURING A SHOULDER ARTHROSCOPY PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE WAS BENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14106112 ↗
  • MalfunctionDec 7, 2021Break; Break

    IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE TIP ON THE ARTHRO GRASPER PEN STR DEVICE BROKE. THERE WERE NO FRAGMENTS LEFT IN THE BODY OF THE PATIENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12955348 ↗
  • MalfunctionOct 19, 2021Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE SALES REP THAT PREOPERATIVELY TO AN UNKNOWN PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE JAWS ON THE ARTHRO GRASPER PEN STR DEVICE WAS NOT ARTICULATING WITH THE HANDPIECE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12658517 ↗
  • MalfunctionMar 2, 2021Break; Break

    IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE PENETRATING GRASPER STRAIGHT DEVICE BROKE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11398815 ↗
  • MalfunctionOct 8, 2020Difficult to Open or Close

    IT WAS REPORTED VIA COMPLAINT SUBMISSION TOOL THAT DURING A SHOULDER REPAIR, THE SURGEON WAS USING THE PENETRATING GRASPER STRAIGHT TO PASS SUTURE THROUGH THE ROTATOR CUFF. HE TRIED TO OPEN THE JAWS, HEARD A POP, AND THE JAWS WOULD NOT OPEN. UPON INVESTIGATION, THE JAWS WOULD NOT OPEN ANY LONGER. THE INSTRUMENT WAS REPLACED WITH A READILY AVAILABLE REPLACEMENT AND THE CASE WAS SUCCESSFULLY COMPLETED WITHOUT PATIENT HARM OR SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10652824 ↗
  • MalfunctionOct 8, 2020Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED VIA COMPLAINT SUBMISSION TOOL THAT DURING A SHOULDER REPAIR, THE SURGEON WAS USING THE PENETRATING GRASPER STRAIGHT TO PASS SUTURE THROUGH THE ROTATOR CUFF. HE TRIED TO OPEN THE JAWS, HEARD A POP, AND THE JAWS WOULD NOT OPEN. UPON INVESTIGATION, THE JAWS WOULD NOT OPEN ANY LONGER. THE INSTRUMENT WAS REPLACED WITH A READILY AVAILABLE REPLACEMENT AND THE CASE WAS SUCCESSFULLY COMPLETED WITHOUT PATIENT HARM OR SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10652785 ↗
  • MalfunctionJan 14, 2020Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE SALES REP VIA PHONE THAT DURING A ROTATOR CUFF REPAIR PROCEDURE BOTH THEIR PENETRATING GRASPER STRAIGHT AND PENETRATING GRASPER 35 DEGREE UP WILL NOT OPEN OR CLOSE. THE PROCEDURE WAS COMPLETED WITH ANOTHER ANGLE DEVICE WITH NO PATIENT HARM OR SURGICAL DELAY TO THE CASE. THE DEVICES WILL BE RETURNING FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9587209 ↗
  • MalfunctionNov 22, 2019Break; Break

    IT WAS REPORTED BY THE SALES REP VIA PHONE THAT THIS PENETRATING GRASPER STRAIGHT BROKE DURING AN UNKNOWN PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCE OR SURGICAL DELAY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9364516 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →