NA 214642Recalls & reported problems

Medos International Sàrl · SIXTER - RIGHT · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886705004232), received between Sep 25, 2019 and Apr 10, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Material Deformation3
  • Break1
  • Device-Device Incompatibility1
  • Difficult to Open or Close1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2024Material Deformation; Appropriate Device Problem Term/Code Not Available; Material Deformation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE AFFILIATE IN IRELAND THAT DURING SERVICE AND REPAIR, IT WAS DETERMINED THAT ARTHRO SIXTER-RIGHT *EA DEVICE WAS DEFECTIVE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DISTAL PART OF THE JAW IS BENT. THE DEICE DID NOT OPEN AND CLOSE DUE TO THE JAW CONDITION. THERE WAS NO PROCEDURE INVOLVED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083431 ↗
  • MalfunctionApr 10, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE AFFILIATE IN IRELAND THAT DURING SERVICE AND REPAIR, IT WAS DETERMINED THAT ARTHRO SIXTER-RIGHT *EA DEVICE WAS DEFECTIVE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE SHARP TIP WAS DEFORMED. THERE WAS NO PROCEDURE INVOLVED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083264 ↗
  • MalfunctionApr 10, 2024Material Deformation; Appropriate Device Problem Term/Code Not Available; Material Deformation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE AFFILIATE IN IRELAND THAT DURING SERVICE AND REPAIR, IT WAS DETERMINED THAT ARTHRO SIXTER-RIGHT *EA DEVICE WAS DEFECTIVE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE SHARP TIP WAS DEFORMED. THERE WAS NO PROCEDURE INVOLVED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19082808 ↗
  • MalfunctionOct 5, 2022Material Deformation; Material Deformation

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY A FIELD SERVICE ENGINEER IN AUSTRALIA THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE SIXTER - RIGHT DEVICE WAS BENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15545263 ↗
  • MalfunctionJun 29, 2022Device-Device Incompatibility; Device-Device Incompatibility

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN UNITED KINGDOM THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE SIXTER - RIGHT DEVICE WAS ABLE TO CAPTURE THE SUTURE INITIALLY BUT WHEN THE SUTURE WAS PASSED THROUGH THE TISSUE, IT WAS LOST. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14856321 ↗
  • MalfunctionJun 14, 2021Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED BY THE AFFILIATE IN GERMANY THAT POSTOPERATIVELY TO AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE ARTHRO SIXTER-RIGHT DEVICE DID NOT CLOSE AND WOULD OPEN POORLY AFTER EACH USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE SHAFT ON THE DEVICE WAS BENT AND THE JAW AT THE DISTAL TIP DID NOT FUNCTION AS IT WAS LOOSE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11991110 ↗
  • MalfunctionSep 25, 2019Break; Break

    IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING A ROTATOR CUFF REPAIR, IT WAS OBSERVED THAT THERE WAS A SNAP SOUND THAT WAS HEARD WHEN THE SURGEON WAS PUSHING THE ARTHRO SIXTER-RIGHT *EA DEVICE INTO THE INCISION. ACCORDING TO THE REPORT, THE EVENT OCCURRED WHEN THE SURGEON PICKED UP THE DEVICE AND ATTEMPTED TO USE IT ON THE PATIENT BUT DUE TO THE SWELLING OF THE PATIENT IT WAS DIFFICULT TO GET THE DEVICE IN THE INCISION. IT WAS FURTHER REPORTED THAT WHEN THE SURGEON PULLED THE SIXTER OUT, HE REALIZED THAT HE MUST HAVE BROKEN THE MECHANISM THAT OPENS AND CLOSES THE JAWS OF THE DEVICE. IT WAS REPORTED THAT ANOTHER TRAY WAS OPENED TO USE THE CLEVER HOOKS INSTEAD FOR THE REST OF THE CASE. THE SURGERY WAS DELAYED ONLY ABOUT ONE MINUTE WHILE WE GRABBED THE TRAY TO OPEN THE CLEVER HOOKS. THERE WAS A DELAY IN THE SURGICAL PROCEDURE. THERE WAS A SPARE DEVICE AVAILABLE FOR USE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9117649 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →