NA 219392Recalls & reported problems
Medos International Sàrl · TIBIAL GUIDE ARM - ELBOW AIMER · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886705006212), received between Nov 5, 2018 and Jan 6, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Failure to Align1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 6, 2023Mechanical Jam; Mechanical Jam
THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE RESTORE TIBIAL GUIDE/RATCHET, ±10% DEVICE AND THE TIBIAL GUIDE ARM - ELBOW AIMER DEVICE BECAME STUCK WHEN THE SURGEON WAS USING WIRE DRIVER TO INSERT PASSING PIN THROUGH THE DEVICE. ACCORDING TO THE REPORT, THE PASSING PIN WOULD NOT ADVANCE AND WAS STUCK IN THE REAMER BULLET. IT WAS REPORTED THAT THE SURGEON REMOVED THE DEVICE AND WAS UNABLE TO REMOVE THE PIN. IT WAS REPORTED THAT THE SURGEON HAD TO FREEHAND DRILLED THE PASSING PIN TO COMPLETE THE PROCEDURE AS THERE WAS NO OTHER TIBIAL AIMER AVAILABLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16117419 ↗MalfunctionOct 28, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING THE ROUTINE PROCESSING AND STERILIZATION FROM WEAR AND TEAR, IT WAS OBSERVED THAT THE TIBIAL GUIDE ARM - ELBOW AIMER DEVICE WAS BROKEN. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12713168 ↗MalfunctionOct 28, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING THE ROUTINE PROCESSING AND STERILIZATION FROM WEAR AND TEAR, IT WAS OBSERVED THAT THE TIBIAL GUIDE ARM - ELBOW AIMER DEVICE WAS BROKEN. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12713096 ↗MalfunctionNov 5, 2018Failure to Align; Failure to Align
THIS IS REPORT 1 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP THAT DURING ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE TIBIAL GUIDE ARM - ELBOW AIMER WAS NOT LINING UP WITH THE TIBIAL GUIDE RAIL. THE CASE WAS COMPLETED WITH ANOTHER LIKE DEVICE. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS. THE SALES REP WAS NOT PRESENT DURING THE CASE AND COULD NOT PROVIDE ANY MORE DETAILS. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8038798 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.