Skip to content

NA 251720Recalls & reported problems

Medos International Sàrl · Suture Grasper 15 Degrees · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886705012282), received between Feb 19, 2019 and Nov 26, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 26, 2024Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED DURING AN UNSPECIFIED PROCEDURE, THE TIP OF THE IDEAL SUTGRASPER 15 DEG *EA DEVICE STOPPED OPENING AFTER ONLY A COUPLE OF PASSES. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20777142 ↗
  • MalfunctionFeb 19, 2019Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED BY THE AFFILIATE THAT DR. XXXXX WAS DOING A REMPLISSAGE PROCEDURE USING THE HEALIX BR 3.4 MM ANCHOR AND PASSING SUTURE RETROGRADE WITH THE IDEAL SUTURE GRASPERS. THE FIRST HEALIX BR 3.4MM ANCHOR HE WENT TO IMPLANT AND THE TIP OF THE ANCHOR SNAPPED JUST TRYING TO PUSH THE ANCHOR THROUGH THE TISSUE TO GET TO THE BONE HOLE. I SUGGESTED WE OPEN AND USE THE PERCANNULA SYSTEM TO MAKE THIS PROCEDURE AND IMPLANTING THE ANCHORS EASIER AND HE AGREED SO WE OPENED THE PERCANNULA AND IMPLANTED A HEALIX BR 3.4MM ANCHOR WITH NO ISSUES. WE THEN OPENED A 30 DEGREE IDEAL SUTURE GRASPER AND WHEN WE HAD THE SUTURE HOOKED IN THE RETRIEVAL CLIP AND WENT TO PULL IT THROUGH THE TISSUE THE CLIP DID NOT HANG ON TO THE STITCH AND STRAIGHTENED OUT MAKING THE TOOL UNUSABLE. WE OPENED A 15 DEGREE IDEAL SUTURE GRASPER TO TRY AS HE THOUGHT THAT ANGLE MIGHT BE BETTER ANYWAY AND THE SAME THING HAPPENED. HE REALIZES THE CLIPS ON THESE ARE FRAGILE, BUT FOR THAT TO HAPPEN ON ONE PASS EACH WAS NOT ACCEPTABLE TO HIM. WE OPENED ONE MORE 15 DEGREE IDEAL SUTURE GRASPER, I GOT HIM TO SLOW DOWN AND LISTEN TO MY ADVICE AND ONCE HE HAD THE SUTURE IN THE GRASPER TO PUT IT IN ¿SLIDE¿ MODE AND GENTLY TOGGLE THE TI