NA 242005Recalls & reported problems

Medos International Sàrl · HD EP Arthroscope/Sinuscope compatible with Storz style 4.0mm x 0 deg x 175mm · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10886705028764), received between Jan 22, 2024 and Apr 12, 2024 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Break” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 14 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break15
  • Use of Device Problem15
  • Device Contaminated During Manufacture or Shipping13
  • Material Deformation5

Patient problems

  • No Clinical Signs, Symptoms or Conditions15

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 12, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19101514 ↗
  • MalfunctionApr 12, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19101505 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM, A COSMETIC ISSUE WITH RUST/DISCOLORATION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084186 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084056 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM DISTAL COVER GLASS/NEGATIVE DAMAGED SCRATCHES/RESIDUE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084030 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083975 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM, AND THERE WAS A COSMETIC ISSUE WITH RUST/DISCOLORATION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083901 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE LENSES WERE BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083833 ↗
  • MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19083782 ↗
  • MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation

    THIS IS REPORT 1 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075665 ↗
  • MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation

    THIS IS REPORT 3 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075657 ↗
  • MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation

    THIS IS REPORT 2 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075654 ↗
  • MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation

    THIS IS REPORT 5 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075645 ↗
  • MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation

    THIS IS REPORT 4 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075644 ↗
  • MalfunctionJan 22, 2024Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED BY THE J&J EMPLOYEE THAT DURING A LARGE TRAINING EVENT FOR AN ENDOSCOPIC SINUS SURGERY ON (B)(6) 2023, IT WAS DISCOVERED THAT 3 OF THE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING PROPERLY. DURING AN IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES WAS FRACTURED. THE DEVICES WERE NOT FOR HUMAN USE, THEREFORE THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18552075 ↗

Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices

All NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices →