NA 242005Recalls & reported problems
Medos International Sàrl · HD EP Arthroscope/Sinuscope compatible with Storz style 4.0mm x 0 deg x 175mm · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10886705028764), received between Jan 22, 2024 and Apr 12, 2024 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break15
- Use of Device Problem15
- Device Contaminated During Manufacture or Shipping13
- Material Deformation5
Patient problems
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19101514 ↗MalfunctionApr 12, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19101505 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM, A COSMETIC ISSUE WITH RUST/DISCOLORATION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19084186 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19084056 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM DISTAL COVER GLASS/NEGATIVE DAMAGED SCRATCHES/RESIDUE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19084030 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19083975 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM, AND THERE WAS A COSMETIC ISSUE WITH RUST/DISCOLORATION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19083901 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE LENSES WERE BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19083833 ↗MalfunctionApr 10, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE HD EPSCP,4.0,0,175,CW_STORZ SCOPE WAS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE OPTICS OF THE SCOPE PARTICULATE UNDER THE DISTAL LENS, AND PARTICULATE UNDER THE PROXIMAL LENS. ALSO, THE LENS WAS BROKEN IN THE OPTICAL SYSTEM. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19083782 ↗MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation
THIS IS REPORT 1 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19075665 ↗MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation
THIS IS REPORT 3 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19075657 ↗MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation
THIS IS REPORT 2 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19075654 ↗MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation
THIS IS REPORT 5 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19075645 ↗MalfunctionApr 9, 2024Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation; Break; Use of Device Problem; Device Contaminated During Manufacture or Shipping; Material Deformation
THIS IS REPORT 4 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING A CADAVERIC TRAINING EVENT, IT WAS DISCOVERED THAT FIVE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING UPON ATTEMPTED USE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WAS DEVICE FRACTURED. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE DEVICES WERE USED EXCLUSIVELY ON CADAVERS. THERE WERE NO REPORTS OF USER INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19075644 ↗MalfunctionJan 22, 2024Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED BY THE J&J EMPLOYEE THAT DURING A LARGE TRAINING EVENT FOR AN ENDOSCOPIC SINUS SURGERY ON (B)(6) 2023, IT WAS DISCOVERED THAT 3 OF THE HD EP ARTHROSCOPE/SINUSCOPE COMPATIBLE WITH STORZ STYLE 4.0MM X 0 DEG X 175MM DEVICES WERE NOT WORKING PROPERLY. DURING AN IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT ONE OF THE DEVICES WAS FRACTURED. THE DEVICES WERE NOT FOR HUMAN USE, THEREFORE THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18552075 ↗
Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page