NA 242009Recalls & reported problems
Medos International Sàrl · HD EP Arthroscope/Sinuscope compatible with Storz style 4.0mm x 70 deg x 175mm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886705028788), received between Mar 3, 2020 and Jun 30, 2022 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Display or Visual Feedback Problem3
- Scratched Material3
- Break2
- Poor Quality Image2
- Use of Device Problem2
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2022Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED BY THE SALES REP THAT DURING AN ARTHROSCOPY PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE HD EP ARTHROSCOPE/SINUSCOPE 4.0MM X 70 DEG X 175MM (STORZ LOCK) DEVICE WAS DROPPED DURING REPROCESSING AND HAD AN UNSPECIFIED MALFUNCTION. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD BROKEN LENSES IN ITS OPTICAL SYSTEM. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14879806 ↗MalfunctionMar 9, 2022Break; Crack; Scratched Material; Break; Crack; Scratched Material
IT WAS REPORTED BY THE SALES REP THAT THE DISTAL TIP ON THE HD EPSCP,4.0,70,175,CW_STORZ DEVICE WAS SCRATCHED AND ITS INTERNAL LENS CRACKED IN HALF. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD BROKEN LENSES IN THE OPTICAL SYSTEM. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13710951 ↗MalfunctionDec 2, 2021Poor Quality Image; Scratched Material; Poor Quality Image; Scratched Material
IT WAS REPORTED BY THE CUSTOMER THAT DURING AN ARTHROSCOPIC HIP SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE DISTAL TIP-ON THE HD EPSCP, (B)(4 )STORZ ENDOSCOPE DEVICE CAME IN CONTACT WITH THE SHAVER BLADE THEN SCRATCHING THE DISTAL TIP. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD AN IMAGE ERROR; AND THAT THE IMAGE ON THE CAMERA WAS CLOUDY/BLURRED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12915903 ↗MalfunctionJul 13, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED BY THE SALES REP THAT DURING STERILE PROCESSING FIELD INSPECTION PREOPERATIVELY TO AN UNKNOWN SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE HD EPSCP,4.0,70,175,CW_STORZ DEVICE APPEARED VERY FOGGY WHEN LOOKING THROUGH IT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158993 ↗MalfunctionJul 9, 2021Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY THE SALES REP THAT DURING HIP PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE DISTAL TIP ON THE HD EPSCP,4.0,70,175,CW_STORZ DEVICE WAS DAMAGED. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD BROKEN LENSES IN ITS OPTICAL SYSTEM. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12143038 ↗MalfunctionApr 30, 2021Scratched Material; Scratched Material
IT WAS REPORTED BY THE SALES REP THAT DURING REPROCESSING FOR A HIP PROCEDURE ON (B)(6) 2020, IT WAS OBSERVED THAT THE DISTAL LENS ON THE ENDOSCOPE DEVICE WAS SCRATCHED. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE IMAGE ON THE CAMERA WAS CLOUDY AND BLURRY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11752636 ↗MalfunctionApr 27, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT THE ENDOSCOPE DEVICE HAD A SHINY ASPECTS IN THE OPTIC. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11733172 ↗MalfunctionFeb 24, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING INSPECTION ON (B)(6) 2020, IT WAS OBSERVED THAT THE LENS ON THE HD EPSCP,4.0,70,175,CW_STORZ DEVICE WAS PHYSICALLY DAMAGED. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11373547 ↗MalfunctionMar 3, 2020Poor Quality Image; Poor Quality Image
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING AN UNKNOWN PROCEDURE THE HD EPSCP, 4.0, 70 , 175, CW_STORZ HAS STAINS AND SCRATCHES. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9780863 ↗
Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.