NA 242388Recalls & reported problems

Medos International Sàrl · LIGHT GUIDE, OLYMPUS - ACMI 5.0MM X 4.0M · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886705028832), received between Jul 27, 2022 and May 3, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 3, 2024Break; Break

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE LIGHT GUIDE, OLYMPUS - ACMI 5.0MM X 4.0M DEVICE BROKE. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19243245 ↗
  • MalfunctionJul 27, 2022Break; Display or Visual Feedback Problem; Break; Display or Visual Feedback Problem

    IT WAS REPORTED BY SALES REP THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE LIGHT GUIDE OLYM-ACMI_5.0MMX4.0M DEVICE WAS NOT PRODUCING ENOUGH LIGHT TO SEE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE EPOXY POTTED END WAS CRASHED AND BROKEN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH ONE MINUTE OF DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15114426 ↗

Other LIGHT, SURGICAL, FIBEROPTIC devices

All LIGHT, SURGICAL, FIBEROPTIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.