NA 02.231.260Recalls & reported problems
Synthes GmbH · 5.0MM VARIABLE ANGLE LOCKNG SCREW/SLF-TPNG/STRDRV/60MM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886982062192), received between Jul 23, 2020 and Jul 21, 2022 — classified as 10 injuries.
The most frequently reported specific device problem is “Device Slipped” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K083025, K110354); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem9
- Device Slipped1
- Device-Device Incompatibility1
Patient problems
- Nonunion/Delayed-union Bone Fracture9
- Unspecified Infection8
- No Code Available1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJul 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GREAT BRITAIN AS FOLLOWS: PATIENT SUSTAINED PERIPROSTHETIC FRACTURE (PPF) BELOW FEMORAL NAIL IN MARCH AND WAS FIXED WITH VA CONDYLAR PLATE ON AN UNKNOWN DATE IN MARCH 2022. PATIENT WAS READMITTED APPROXIMATELY 3 MONTHS AFTER IMPLANT WITH PLATE FRACTURE. NO NEW TRAUMA. HAD BEEN MOBILIZING IN RESIDENTIAL HOME. NORMAL TO LOW BMI. FRACTURE NON-UNITED, BUT NO EVIDENCE OF INFECTION. REVISION OCCURRED (B)(6) 2022. THIS IS REPORT 7 OF 11 FOR (B)(4). THIS REPORT IS FOR A LOCKING SCREW. THIS COMPLAINT IS LINKED TO (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 15077703 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839775 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839774 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839772 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839771 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839769 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839766 ↗InjuryMar 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022 AFTER THE SURGERY DURING THE REMOVAL OF VALCP CURVED CONDYLAR PLATE, SCREWS, AND A ZIMMER RETROGRADE NAIL DUE TO INFECTION AND NONUNION. PATIENT¿S PREVIOUS BELOW KNEE AMPUTATION WAS REVISED TO AN ABOVE KNEE AMPUTATION. THE PATIENT EXPERIENCED NO POST-OP DEVICE MALFUNCTION. THIS COMPLAINT INVOLVES TWELVE (12) DEVICES.
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 13839757 ↗InjuryAug 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL EVENT DESCRIPTION: ON 10/04/2019, A CLINIC REPORTED THAT PATIENT HAD A RECENT FALL, AND SUSTAINED A DISPLACED FRACTURE DISTAL TIP. PAIN LEVEL IS 4/10. ON (B)(6) 2019, PROGRESS NOTE REPORTED THAT A FRACTURE PLATE AT THE SITE OF PREVIOUS FRACTURE.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 12308609 ↗InjuryJul 23, 2020Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 THAT THE PATIENT UNDERWENT THE REMOVAL OF OSTEOSYNTHESIS MATERIAL AND A CONDYLAR SUPPORT PLATE THAT HAD FAILED. THE DISTAL LOCKING SCREWS OF THE PLATE WERE LOOSE AND ONE WAS ALREADY EXPOSED OUTSIDE OF THE SKIN. AN ADDITIONAL REDUCTION WAS PERFORMED BUT THE SCREWS DID NOT BLOCK THE PLATE AND IT WAS DECIDED TO REMOVE ALL THE MATERIAL AND PUT A SPLINT IN ONE OF THE WOUNDS THAT WAS FOUND TO BE INFECTED. THE FIRST INTERVENTION WAS PERFORMED BECAUSE THE DISTAL SCREWS OF THE PLATE CAME OFF. THE SECOND OPERATION WAS PERFORMED TO EXTRACT A SCREW THAT WAS EXPOSED THROUGH THE SKIN AND THE THIRD INTERVENTION WAS WHEN ALL OF THE MATERIAL WAS REMOVED. THE FIRST TWO INTERVENTIONS WERE PERFORMED TO REMOVE THE IMPLANT AND TRY TO CORRECT THE NEW FRACTURE, IN THE LAST SURGERY IT WAS SHOWN THAT THERE WAS AN INFECTION. THIS REPORT IS FOR ONE (1) 5.0MM VARIABLE ANGLE LOCKNG SCREW/SLF-TPNG/STRDRV/60MM. THIS IS REPORT 4 OF 9 FOR (B)(4). THIS REPORT IS LINKED TO (B)(4).
Patient problems: No Code Available; Unspecified Infection
MAUDE report 10312728 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083025, K110354. One submission often covers many models, so check the product and lot details in each record.
- Z-1125-2014 · Feb 5, 2014 · TerminatedSynthes 5.0mm Variable Locking Screw This device is part of the Variable Angle Locking Compression Plate family and this technology included variable angle Combi holes in the plat
- Z-0401-2014 · May 29, 2013 · TerminatedSynthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER Synthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER is used for buttressing multi fragmentary diatal femur fra
Other condylar plate fixation Implant devices
- VA-LCP 03.231.007Synthes GmbH · condylar plate fixation Implant56 reports
- VA-LCP 03.231.005Synthes GmbH · condylar plate fixation Implant37 reports
- VA-LCP 02.124.416Synthes GmbH · condylar plate fixation Implant24 reports
- VA-LCP 02.124.413Synthes GmbH · condylar plate fixation Implant14 reports
- VA-LCP 02.124.408Synthes GmbH · condylar plate fixation Implant12 reports
- NA 02.231.675Synthes GmbH · condylar plate fixation Implant11 reports
- NA 02.231.275Synthes GmbH · condylar plate fixation Implant10 reports
- NA 02.231.670Synthes GmbH · condylar plate fixation Implant6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page