NA 02.231.275Recalls & reported problems
Synthes GmbH · 5.0MM VARIABLE ANGLE LOCKNG SCREW/SLF-TPNG/STRDRV/75MM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886982062222), received between Apr 9, 2018 and Jul 21, 2022 — classified as 10 injuries.
The most frequently reported specific device problem is “Break” (5 reports).
2 other recalls cover devices cleared under the same FDA submission (K083025, K110354); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Break5
Patient problems
- Insufficient Information5
- Nonunion/Delayed-union Bone Fracture5
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJul 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: PATIENT SUSTAINED PERIPROSTHETIC FRACTURE (PPF) BELOW FEMORAL NAIL IN (B)(6) AND WAS FIXED WITH VA CONDYLAR PLATE ON AN UNKNOWN DATE IN (B)(6) 2022. PATIENT WAS READMITTED APPROXIMATELY 3 MONTHS AFTER IMPLANT WITH PLATE FRACTURE. NO NEW TRAUMA. HAD BEEN MOBILIZING IN RESIDENTIAL HOME. NORMAL TO LOW BMI. FRACTURE NON-UNITED, BUT NO EVIDENCE OF INFECTION. REVISION OCCURRED (B)(6) 2022. THIS IS REPORT 2 OF 11 FOR (B)(4). THIS REPORT IS FOR A VA-LOCKING SCREW. THIS COMPLAINT IS LINKED TO (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 15075275 ↗InjuryJun 24, 2021Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT A ORIF DISTAL FEMUR PROCEDURE. IT WAS NOTED THAT THE PLATE BROKE AT HOLE NUMBER 8 AND ALSO THE 3 - 4.5MM CORTICAL SCREWS BROKE. IT IS UNKNOWN IF FRAGMENTS WERE GENERATED. IT IS UNKNOWN IF THE PROCEDURE IS COMPLETED. NO PATIENT CONSEQUENCES NOTED. DEVICES WERE INVOLVED: 214.836 (4.5MM CORTEX SCREW SELF-TAPPING 36MM); QUANTITY: 3; 02.231.275 (5.0 VA LOCKNG SCR SLF-TPNG/SD/75); QUANTITY: 5-6; A PLATE. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/75. THIS REPORT IS 9 OF 10 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 12063895 ↗InjuryJun 24, 2021Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT A ORIF DISTAL FEMUR PROCEDURE. IT WAS NOTED THAT THE PLATE BROKE AT HOLE NUMBER 8 AND ALSO THE 3 - 4.5MM CORTICAL SCREWS BROKE. IT IS UNKNOWN IF FRAGMENTS WERE GENERATED. IT IS UNKNOWN IF THE PROCEDURE IS COMPLETED. NO PATIENT CONSEQUENCES NOTED. DEVICES WERE INVOLVED: 214.836 (4.5MM CORTEX SCREW SELF-TAPPING 36MM), QUANTITY: 3, 02.231.275 (5.0 VA LOCKNG SCR SLF-TPNG/SD/75), QUANTITY: 5-6, A PLATE. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/75. THIS REPORT IS 10 OF 10 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 12063875 ↗InjuryJun 24, 2021Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT A ORIF DISTAL FEMUR PROCEDURE. IT WAS NOTED THAT THE PLATE BROKE AT HOLE NUMBER 8 AND ALSO THE 3 - 4.5MM CORTICAL SCREWS BROKE. IT IS UNKNOWN IF FRAGMENTS WERE GENERATED. IT IS UNKNOWN IF THE PROCEDURE IS COMPLETED. NO PATIENT CONSEQUENCES NOTED. DEVICES WERE INVOLVED: 214.836 (4.5MM CORTEX SCREW SELF-TAPPING 36MM), QUANTITY: 3, 02.231.275 (5.0 VA LOCKNG SCR SLF-TPNG/SD/75), QUANTITY: 5-6, A PLATE. THIS COMPLAINT INVOLVES (10) DEVICES. THIS REPORT IS FOR (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/75. THIS REPORT IS 8 OF 10 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 12063873 ↗InjuryJun 24, 2021Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT A ORIF DISTAL FEMUR PROCEDURE. IT WAS NOTED THAT THE PLATE BROKE AT HOLE NUMBER 8 AND ALSO THE 3 - 4.5MM CORTICAL SCREWS BROKE. IT IS UNKNOWN IF FRAGMENTS WERE GENERATED. IT IS UNKNOWN IF THE PROCEDURE IS COMPLETED. NO PATIENT CONSEQUENCES NOTED. DEVICES WERE INVOLVED: 214.836 (4.5MM CORTEX SCREW SELF-TAPPING 36MM); QUANTITY: 3; 02.231.275 (5.0 VA LOCKNG SCR SLF-TPNG/SD/75); QUANTITY: 5-6; A PLATE. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/75. THIS REPORT IS 7 OF 10 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 12063868 ↗InjuryJun 24, 2021Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT A ORIF DISTAL FEMUR PROCEDURE. IT WAS NOTED THAT THE PLATE BROKE AT HOLE NUMBER 8 AND ALSO THE 3 - 4.5MM CORTICAL SCREWS BROKE. IT IS UNKNOWN IF FRAGMENTS WERE GENERATED. IT IS UNKNOWN IF THE PROCEDURE IS COMPLETED. NO PATIENT CONSEQUENCES NOTED. DEVICES WERE INVOLVED: 214.836 (4.5MM CORTEX SCREW SELF-TAPPING 36MM); QUANTITY: 3; 02.231.275 (5.0 VA LOCKNG SCR SLF-TPNG/SD/75); QUANTITY: 5-6; A PLATE. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR ONE (1) 5.0 VA LOCKNG SCR SLF-TPNG/SD/75. THIS REPORT IS 2 OF 10 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 12063845 ↗InjuryApr 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, A PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE DISTAL FEMUR USING SYNTHES VA CONDYLAR PLATE AND SCREWS. THE OUTCOME OF THE SURGERY WAS A NONUNION. ON (B)(6) 2018 THE SURGEON PERFORMED A REVISION ORIF OF THE DISTAL FEMUR TO ADDRESS THE NONUNION. THE PLATES AND SCREWS WERE INTACT AND SUCCESSFULLY REMOVED. A BONE GRAFT WAS INSERTED AND A NEW SYNTHES VA CONDYLAR PLATE AND SCREWS WERE INSERTED IN THE PATIENT. THE REVISION WAS COMPLETED SUCCESSFULLY WITH NO HARM TO THE PATIENT. THIS IS REPORT 8 OF 8 (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 7408033 ↗InjuryApr 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, A PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE DISTAL FEMUR USING SYNTHES VA CONDYLAR PLATE AND SCREWS. THE OUTCOME OF THE SURGERY WAS A NONUNION. ON (B)(6) 2018 THE SURGEON PERFORMED A REVISION ORIF OF THE DISTAL FEMUR TO ADDRESS THE NONUNION. THE PLATES AND SCREWS WERE INTACT AND SUCCESSFULLY REMOVED. A BONE GRAFT WAS INSERTED AND A NEW SYNTHES VA CONDYLAR PLATE AND SCREWS WERE INSERTED IN THE PATIENT. THE REVISION WAS COMPLETED SUCCESSFULLY WITH NO HARM TO THE PATIENT. THIS IS REPORT 7 OF 8 (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 7408005 ↗InjuryApr 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, A PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE DISTAL FEMUR USING SYNTHES VA CONDYLAR PLATE AND SCREWS. THE OUTCOME OF THE SURGERY WAS A NONUNION. ON (B)(6) 2018 THE SURGEON PERFORMED A REVISION ORIF OF THE DISTAL FEMUR TO ADDRESS THE NONUNION. THE PLATES AND SCREWS WERE INTACT AND SUCCESSFULLY REMOVED. A BONE GRAFT WAS INSERTED AND A NEW SYNTHES VA CONDYLAR PLATE AND SCREWS WERE INSERTED IN THE PATIENT. THE REVISION WAS COMPLETED SUCCESSFULLY WITH NO HARM TO THE PATIENT. THIS IS REPORT 6 OF 8 (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 7407344 ↗InjuryApr 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, A PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE DISTAL FEMUR USING SYNTHES VA CONDYLAR PLATE AND SCREWS. THE OUTCOME OF THE SURGERY WAS A NONUNION. ON (B)(6) 2018 THE SURGEON PERFORMED A REVISION ORIF OF THE DISTAL FEMUR TO ADDRESS THE NONUNION. THE PLATES AND SCREWS WERE INTACT AND SUCCESSFULLY REMOVED. A BONE GRAFT WAS INSERTED AND A NEW SYNTHES VA CONDYLAR PLATE AND SCREWS WERE INSERTED IN THE PATIENT. THE REVISION WAS COMPLETED SUCCESSFULLY WITH NO HARM TO THE PATIENT. THIS IS REPORT 2 OF 8 (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 7406656 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083025, K110354. One submission often covers many models, so check the product and lot details in each record.
- Z-1125-2014 · Feb 5, 2014 · TerminatedSynthes 5.0mm Variable Locking Screw This device is part of the Variable Angle Locking Compression Plate family and this technology included variable angle Combi holes in the plat
- Z-0401-2014 · May 29, 2013 · TerminatedSynthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER Synthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER is used for buttressing multi fragmentary diatal femur fra
Other condylar plate fixation Implant devices
- VA-LCP 03.231.007Synthes GmbH · condylar plate fixation Implant56 reports
- VA-LCP 03.231.005Synthes GmbH · condylar plate fixation Implant37 reports
- VA-LCP 02.124.416Synthes GmbH · condylar plate fixation Implant24 reports
- VA-LCP 02.124.413Synthes GmbH · condylar plate fixation Implant14 reports
- VA-LCP 02.124.408Synthes GmbH · condylar plate fixation Implant12 reports
- NA 02.231.675Synthes GmbH · condylar plate fixation Implant11 reports
- NA 02.231.260Synthes GmbH · condylar plate fixation Implant10 reports
- NA 02.231.670Synthes GmbH · condylar plate fixation Implant6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page