NA 03.010.050Recalls & reported problems
Synthes GmbH · AIMING ARM FOR RETROGRADE STANDARD LOCKING · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982067227), received between Apr 5, 2018 and Aug 14, 2023 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility2
- Failure to Align2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 14, 2023Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CONNECTING SCREW IS CROSS THREADED AND AIMING ARM IS BROKEN IN HALF. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) AIMING ARM FOR RETROGRADE STANDARD LOCKING THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17541265 ↗MalfunctionFeb 10, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON AN UNKNOWN DATE, THE AIMING ARM AND GUIDE MISSED NAIL. THERE WAS NO PATIENT OUTCOMES REPORTED. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR ONE (1) AIMING ARM FOR RETROGRADE STD LOCKING. THIS IS REPORT 1 OF 4 FOR COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16358367 ↗MalfunctionFeb 10, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON AN UNKNOWN DATE, THE AIMING ARM AND GUIDE MISSED NAIL. THERE WAS NO PATIENT OUTCOMES REPORTED. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR ONE (1) AIMING ARM FOR RETROGRADE STD LOCKING. THIS IS REPORT 2 OF 4 FOR COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16358362 ↗MalfunctionMay 24, 2018Failure to Align; Failure to Align
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, THE 4.2MM DRILL BIT WAS HITTING THE NAIL WHEN DRILLING FOR THE TARGETED DISTAL INTERLOCK. SURGEON WAS ABLE TO PASS THE DRILL BIT THROUGH, BUT IT STILL HIT THE NAIL. NO PATIENT CONSEQUENCE OR SURGICAL DELAY WAS REPORTED. THIS REPORT IS FOR ONE (1) AIMING ARM FOR RETROGRADE STANDARD LOCKING THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7541536 ↗InjuryApr 5, 2018Failure to Align; Failure to Align
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT A PATIENT WAS IMPLANTED WITH A RETROGRADE/ANTEGRADE FEMORAL NAIL (RAFN) TO TREAT THE DISTAL FEMUR FRACTURE ON (B)(6) 2018. DURING THE PROCEDURE THE TWO (2) DISTAL SCREWS MISSED THE NAIL HOLE (RAFN) WHICH CAUSED THE SCREWS INSERTED POSTERIORLY. THE CONNECTING SCREW WAS TIGHTLY SECURED TO THE INSERTION HANDLE AND PRE-CHECKING OF THE NAIL AND JIG ALIGNMENT WAS DONE AND IT WAS ALIGNED PERFECTLY. SURGEON DOESN¿T KNOW WHY THE JIG WOULD HAVE MISSED THE NAIL, GIVEN THAT THE CONNECTING SCREW WAS TIGHT TOO. THE SURGEON HAD TO REMOVE THE 2 SCREWS AND REPLACE IT WITH A SPIRAL BLADE FOR THE NAIL. THERE WAS A THREE (3) MINUTE SURGICAL DELAY TO THE PROCEDURE. NO PATIENT CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) AIMING ARM FOR RETROGRADE STANDARD LOCKING. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Code Available
MAUDE report 7399644 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page