NA 03.010.415Recalls & reported problems
Synthes GmbH · CONNECTING SCREW F/GUIDE WIRE AIMING DEVICE F/TFN · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982068590), received between Jul 3, 2018 and Apr 22, 2022 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Device Operates Differently Than Expected2
- Device-Device Incompatibility1
Patient problems
- No Code Available2
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED 03.010.412 SEEMS BROKEN. CLIENT WANTS TO HAVE A REPLACEMENT. THIS REPORT IS FOR ONE (1) CONNECSCR F/TFN F/03.010.412. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14189404 ↗MalfunctionOct 1, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING A TRAINING SESSION IT WAS NOTICED THAT THE AIMING ARM HANDLE CANNOT BE CONNECTED TO THE AIMING ARM WITH THE CONNECTION SCREW BECAUSE THE PROVIDED CONNECTION SCREW IS A WRONG SCREW. IT IS A POSSIBLE MISMATCH WHILE ASSEMBLING THE TRAINING SET. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS REPORT IS FOR ONE (1) CONNECTING SCREW F/GUIDE WIRE AIMING DEVICE F/TFN. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10617229 ↗MalfunctionSep 1, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, PATIENT UNDERWENT HIP INTRAMEDULLARY NAILING, DURING THE PROCEDURE THE CONNECTING NUT OF 130 DEGREE TUNA TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) AIMING ARM BROKE DURING ASSEMBLY. THE CONNECTING NUT WAS REFERRING TO THE CONNECTING SCREW FOR 130 AIMING GUIDE. THE AIMING ARM WAS NOT PROPERLY SECURED TO TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) INSERTION HANDLE. THIS MADE IT DIFFICULT TO INSERT GUIDE WIRE PROPERLY. PLIERS WERE USED TO SECURE THE AIMING ARM. THE GUIDEWIRE WAS EVENTUALLY INSERTED WHERE IT WAS NEEDED. THERE WAS A SURGICAL DELAY OF FIVE (5) MINUTES. THERE WERE FRAGMENTS GENERATED FROM BROKEN DEVICE, THE SCREW BROKE PERFECTLY IN HALF. THE TWO FRAGMENTS WERE EASILY REMOVED FROM THE FIELD WITH NO ADDITIONAL INTERVENTION. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: NAILS: TFNA (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1), 130 DEG AIMING ARM (PART NUMBER 03.037.013, LOT 9282818, QUANTITY 1), NAIL INSERTION HANDLES (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1), GUIDE/COMPRESSION/K-WIRES (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) CONNECTING
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10476275 ↗MalfunctionJul 3, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE. DURING THE PROCEDURE, THE GUIDE WIRE AIMING DEVICE FOR THE TROCHANTERIC FIXATION NAIL (TFN) AND TWO (2) CONNECTING SCREWS FOR GUIDE WIRE AIMING DEVICE WERE NOT INTERFACING CORRECTLY WITH 125 DEGREE AND 130 DEGREE AIMING ARMS. THE TOOLS WERE ABANDONED FROM USAGE. VISUAL TECHNIQUE WAS USED TO PROCEED WITH NAIL DEPTH ASSESSMENT. A USER ERROR WAS SUSPECTED AND FOLLOW-UP EQUIPMENT INSPECTION WAS NEEDED. THE CONNECTING SCREWS WERE FOUND BENT AND THE THREADS WERE DAMAGED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 2 MINUTES OF SURGICAL DELAY DUE TO SWAPPING OF THE SECOND CONNECTING SCREW. THE PATIENT WAS REPORTEDLY FINE AT THE END OF THE PROCEDURE. IN ADDITION, ON JUNE 11, 2018, A NON-STERILE INSPECTION WAS DONE WHICH VALIDATED THE CONFIRMATION OF A POOR INTERFACE. CONCOMITANT DEVICE/S REPORTED: 125 DEGREE AIMING ARM (PART: 03.037.014, LOT: UNKNOWN, QUANTITY: 1); 130 DEGREE AIMING ARM (PART: 03.037.013, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A CONNECTING SCREW F/GUIDE WIRE AIMING DEVICE F/TFN. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7658093 ↗MalfunctionJul 3, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE. DURING THE PROCEDURE, THE GUIDE WIRE AIMING DEVICE FOR THE TROCHANTERIC FIXATION NAIL (TFN) AND TWO (2) CONNECTING SCREWS FOR GUIDE WIRE AIMING DEVICE WERE NOT INTERFACING CORRECTLY WITH 125 DEGREE AND 130 DEGREE AIMING ARMS. THE TOOLS WERE ABANDONED FROM USAGE. VISUAL TECHNIQUE WAS USED TO PROCEED WITH NAIL DEPTH ASSESSMENT. A USER ERROR WAS SUSPECTED AND FOLLOW-UP EQUIPMENT INSPECTION WAS NEEDED. THE CONNECTING SCREWS WERE FOUND BENT AND THE THREADS WERE DAMAGED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 2 MINUTES OF SURGICAL DELAY DUE TO SWAPPING OF THE SECOND CONNECTING SCREW. THE PATIENT WAS REPORTEDLY FINE AT THE END OF THE PROCEDURE. IN ADDITION, ON JUNE 11, 2018, A NON-STERILE INSPECTION WAS DONE WHICH VALIDATED THE CONFIRMATION OF A POOR INTERFACE. CONCOMITANT DEVICE/S REPORTED: 125 DEGREE AIMING ARM (PART: 03.037.014, LOT: UNKNOWN, QUANTITY: 1); 130 DEGREE AIMING ARM (PART: 03.037.013, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A CONNECTING SCREW F/GUIDE WIRE AIMING DEVICE F/TFN. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7658081 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page