EXPert 03.010.483Recalls & reported problems
Synthes GmbH · RADIOLUCENT RECON AIMING ARM FOR EXPERT ALFN · Class 2
All models of EXPert (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886982069016), received between Apr 3, 2018 and Jan 14, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility5
- Break2
- Failure to Align2
- Crack1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 14, 2026Break
USED AN AIMING ARM AND THE CLIPS THAT HOLD THE TROCARS IN PLACE BROKE OFF. THERE WAS NO PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24067063 ↗MalfunctionDec 17, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PRODUCT HAS MISSED THE NAIL IN THE PAST 2 CASES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23829067 ↗MalfunctionDec 17, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PRODUCT HAS MISSED THE NAIL IN THE PAST 2 CASES THEY HAVE DONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23828947 ↗MalfunctionJun 4, 2025Break; Crack; Device-Device Incompatibility; Break; Crack; Device-Device Incompatibility
IT WAS REPORTED THAT USER WAS UNABLE TO STERILIZE/ASSEMBLE. IT WAS CHIPPING AND CRACKING. USER WAS UNABLE TO USE AND NEEDED REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22143778 ↗MalfunctionOct 11, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON SEPTEMBER 15, 2023, THE AIMING ARM WAS DAMAGED DURING REMOVAL. THE AIMING ARM WAS NOT LINING UP; THE SCREW WAS HITTING THE NAILS. THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE RADIOLUCENT RECON AIMING ARM FOR EXPERT ALFN. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17911314 ↗MalfunctionMay 26, 2021Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT THE AIMING ARM REQUIRED REPLACEMENT AT THE JUNCTION WHERE THE CLIPS LOCK. IT WAS WORN DOWN AND DID NOT RETAIN THE CLIPS. THIS REPORT IS FOR ONE (1) RADIOLUCENT RECON AIMING ARM FOR EXPERT ALFN. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11888718 ↗MalfunctionMay 9, 2018Failure to Align; Failure to Align
IT WAS REPORTED DURING FEMUR OPEN REDUCTION AND INTERNAL FIXATION (ORIF) PERFORMED ON (B)(6) 2018 FOR AN UNKNOWN PATIENT THE SURGEON WAS ASSEMBLING THE ADOLESCENT LATERAL FEMORAL NAIL (ALFN) WHEN IT WAS NOTED THE PROXIMAL INTERLOCKS WEREN¿T LINING UP IN THE NAIL THROUGH THE AIMING ASSEMBLY. THE DRILL WAS COMPLETELY MISSING THE NAIL THROUGH THE AIMING ASSEMBLY. AFTER THE CASE, THE SALES CONSULTANT INSPECTED THE DEVICES AND NOTED THE AIMING ARM TO BE LOOSE. THE AIMING ARM AND CONNECTING ARM HAD SOME PLAY WHEN FULLY SEATED DUE TO WEAR AND TEAR AND THE CONNECTING ARM TO BE CHIPPED ON THE METAL CLIPS THAT CONNECT TO THE NAIL. THESE DEVICES WERE NOT USED ON THE PATIENT DURING THE SURGERY. THE SURGEON COMPLETED THE PROCEDURE BY DOING PERFECT CIRCLES. THERE WAS NO REPORTED SURGICAL DELAY AND NO REPORTED PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: INSERTION INSTRUMENT: COUPLING SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1); ALFN NAIL (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) RADIOLUCENT RECON AIMING ARM FOR EXPERT ALFN. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7498364 ↗InjuryApr 3, 2018Failure to Align; Failure to Align
IT WAS REPORTED THAT THE SCREWS WOULD NOT GO IN THE AIMING ARM WHEN THE SURGEON WAS ATTEMPTING TO INSERT THE PROXIMAL LOCKING SCREWS IN AN ADOLESCENT FEMORAL NAIL DURING AN INTRAMEDULLARY NAIL OF A FEMUR ON (B)(6) 2018. THE SALES CONSULTANT FURTHER EXPLAINED THAT THE AIMING ARM WOULD ACCEPT THE DRILL BIT, BUT NOT THE SCREWS. THE INSERTION HANDLE AND/OR THE AIMING ARM KEPT THE PROTECTION SLEEVE FROM LINING UP PROPERLY, WHICH EVENTUALLY LED TO THE TRANSVERSE INTERLOCK NOT WORKING PROPERLY AND THEY HAD TO SWITCH THE HOLE AND USE THE ANTEGRADE INTERLOCK TO LOCK THE SCREW AND COMPLETE THE PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THE PATIENT WAS FINE AT THE END OF THE PROCEDURE. CONCOMITANT DEVICE (S) REPORTED:12.0MM/8.0MM PROTECTION SLEEVE (PART # 03.010.063, UNKNOWN LOT #, QTY # 1). THIS REPORT IS FOR ONE (1) RADIOLUCENT RECON AIMING ARM. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7393711 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page