Skip to content

NA 03.019.014Recalls & reported problems

Synthes GmbH · 10.0MM/3.8MM DRILL SLEEVE FOR 03.019.013 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982069511), received between May 16, 2019 and Jul 3, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility3
  • No Apparent Adverse Event1

Patient problems

  • No Patient Involvement3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT ON JUNE 7, 2024 THE TROCAR IS BENT AND NEEDS TO BE REPLACED. THERE WAS NO SURGICAL DELAY. PROCEDURE SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) 10.0MM/3.8MM DRILL SLEEVE FOR 03.019.013. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19668831 ↗
  • MalfunctionMay 16, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED ON APRIL 17, 2019 DURING AN INSPECTION AT STERILE PROCESSING DEPARTMENT, THE INSERTION HANDLE, AIMING ARM, CONNECTING SCREW, AIMING ARM KNOB, THREE (3) DRILL SLEEVE AND THREE (3) PROTECTION SLEEVE OF THE MULTILOC HUMERAL NAIL INSTRUMENTS DID NOT ACCURATELY TARGET HOLES IN NAIL WITH DRILL BITS. THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8615669 ↗
  • MalfunctionMay 16, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED ON (B)(6) 2019 DURING AN INSPECTION AT STERILE PROCESSING DEPARTMENT, THE INSERTION HANDLE, AIMING ARM, CONNECTING SCREW, AIMING ARM KNOB, THREE (3) DRILL SLEEVE AND THREE (3) PROTECTION SLEEVE OF THE MULTILOC HUMERAL NAIL INSTRUMENTS DID NOT ACCURATELY TARGET HOLES IN NAIL WITH DRILL BITS. THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8615649 ↗
  • MalfunctionMay 16, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED ON APRIL 17, 2019 DURING AN INSPECTION AT STERILE PROCESSING DEPARTMENT, THE INSERTION HANDLE, AIMING ARM, CONNECTING SCREW, AIMING ARM KNOB, THREE (3) DRILL SLEEVE AND THREE (3) PROTECTION SLEEVE OF THE MULTILOC HUMERAL NAIL INSTRUMENTS DID NOT ACCURATELY TARGET HOLES IN NAIL WITH DRILL BITS. THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8615563 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page