NA 03.111.024Recalls & reported problems
Synthes GmbH · COMPRESSION/DISTRACTION HOLDING SLEEVE/3.0MM K-WIRE · Class 1
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10886982072238), received between Dec 10, 2020 and Nov 3, 2025 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility5
- Failure to Cut5
- Material Deformation4
- Break2
- Contamination /Decontamination Problem1
- Device Slipped1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2025Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility
IT WAS REPORTED THE COMPRESSION/DISTRACTION SHEATH DIDN'T GRIP PROPERLY, MOVED AND DIDN'T STAY IN PLACE. ANOTHER DEVICE WAS AVAILABLE TO USE SO THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452042 ↗MalfunctionJun 17, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE INSTRUMENTS WERE DULL AND NEED TO BE REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22236145 ↗MalfunctionJun 17, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT INSTRUMENTS WERE DULL AND NEED REPLACEMENT. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY DUE TO THE REPORTED EVENT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22235608 ↗MalfunctionJun 17, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT INSTRUMENTS WERE DULL AND NEED REPLACEMENT. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY DUE TO THE REPORTED EVENT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22235345 ↗MalfunctionJun 17, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT INSTRUMENTS WERE DULL AND NEED REPLACEMENT. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY DUE TO THE REPORTED EVENT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22235204 ↗MalfunctionJun 17, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THAT INSTRUMENTS WERE DULL AND NEED REPLACEMENT. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY DUE TO THE REPORTED EVENT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22235048 ↗MalfunctionJan 27, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING DEVICE WAS RECEIVED FOR ANALYSIS: 03.111.024, LOT NO: 821P559. THERE WAS NO PRODUCT MALFUNCTION OR PATIENT HARM REPORTED WITH THE RETURNED DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21241913 ↗MalfunctionJul 24, 2024Material Deformation; Material Deformation
IT WAS REPORTED ON AN UNKNOWN DATE, IT WAS DETERMINED AN INSTRUMENT IN A LOANER SET WAS BENT / CROOKED. NO PATIENT IMPACT OR IMPACT ON THE PROCEDURE. THERE WAS NO DELAY. THE OTHER DEVICE IN THE SET WAS USED. THIS REPORT IS FOR HOLD-SL Ø3 F/COMPRESSION/DISTRACTION DEV. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19819111 ↗MalfunctionMay 8, 2023Break; Device-Device Incompatibility; Material Deformation; Break; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED, THAT ON UNKNOWN DATE. DURING THE CASE, THE COMPRESSION/DISTRACTOR BROKE, ALONG WITH THE 3.0MM SLEEVES USED FOR IT. THE PIECE SEEMS TO BE MISSING INSIDE THE DISTRACTOR, SO THERE IS NO ¿RATCHET¿ WHEN YOU SPIN THE DISTRACTOR. ALSO, THE SLEEVES SEEM BENT AND A 3.0MM WIRE CAN'T FIT THROUGH THEM. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. THIS COMPLAIN INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) COMP/DISTRACTION HOLDING SLV/3.0 K-WIRE. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16891631 ↗MalfunctionMay 8, 2023Break; Device-Device Incompatibility; Material Deformation; Break; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT ON UNKNOWN DATE, DURING THE CASE, THE COMPRESSION/DISTRACTOR BROKE, ALONG WITH THE 3.0MM SLEEVES USED FOR IT. THE PIECE SEEMS TO BE MISSING INSIDE THE DISTRACTOR, SO THERE IS NO ¿RATCHET¿ WHEN YOU SPIN THE DISTRACTOR. ALSO, THE SLEEVES SEEM BENT AND A 3.0MM WIRE CAN'T FIT THROUGH THEM. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. THIS COMPLAIN INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) COMP/DISTRACTION HOLDING SLV/3.0 K-WIRE THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16891623 ↗MalfunctionSep 14, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SLEEVES FAILED TO LOCK ON TO THE WIRES DURING A SURGERY. THERE WAS NO SURGICAL DELAY AND THERE WAS NO PATIENT HARM. THIS REPORT IS FOR ONE (1) COMP/DISTRACTION HOLDING SLV/3.0 K-WIRE. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15420450 ↗MalfunctionSep 14, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SLEEVES FAILED TO LOCK ON TO THE WIRES DURING A SURGERY. THERE WAS NO SURGICAL DELAY AND THERE WAS NO PATIENT HARM. THIS REPORT IS FOR ONE (1) COMP/DISTRACTION HOLDING SLV/3.0 K-WIRE THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15420438 ↗MalfunctionDec 10, 2020Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT 3MM K-WIRE DISTRACTION HOLDING SLEEVE HAD METAL STUCK IN THE CANNULATION SO WAS UNUSABLE DURING A DISTRACTION OF THE ULNA - DIAPHYSEAL ACLASIA CASE PERFORMED ON (B)(6) 2020. THE CANNULATION OF THE INSTRUMENT WAS BLOCKED. SURGEON HAD TO USE 2.5MM K-WIRE AND SLEEVE AS AN ALTERNATIVE. THE SURGEON WAS NOT ABLE TO USE HIS DESIRED CHOICE OF DIAMETER FOR THE K-WIRES DUE TO THIS ISSUE HOWEVER IT DID NOT RESULT IN ANY ADVERSE EVENTS FOR THE PATIENT. THERE WAS A FIFTEEN (15) MINUTE SURGICAL DELAY HOWEVER THE CASE WAS FINISHED SUCCESSFULLY. THE PATIENT OUTCOME IS REPORTED AS WELL. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED TWO (2) DISTRACTION HOLDING SLEEVES/3.0MM K-WIRE IT WAS OBSERVED THAT THE SURFACE OF THE INSTRUMENTS ARE WORN AND DEFORMED FROM FREQUENT USING AS WELL AS THE BORE ARE WORN AND DEFORMED. THIS DEVICE CONDITION WAS EVALUATED AND DETERMINED TO BE REPORTABLE ON (B)(6) 2020. THIS REPORT IS FOR ONE (1) COMPRESSION/DISTRACTION HOLDING SLEEVE/3.0MM K-WIRE. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10985947 ↗MalfunctionDec 10, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT 3MM K-WIRE DISTRACTION HOLDING SLEEVE HAD METAL STUCK IN THE CANNULATION SO WAS UNUSABLE DURING A DISTRACTION OF THE ULNA - DIAPHYSEAL ACLASIA CASE PERFORMED ON (B)(6) 2020. THE CANNULATION OF THE INSTRUMENT WAS BLOCKED. SURGEON HAD TO USE 2.5MM K-WIRE AND SLEEVE AS AN ALTERNATIVE. THE SURGEON WAS NOT ABLE TO USE HIS DESIRED CHOICE OF DIAMETER FOR THE K-WIRES DUE TO THIS ISSUE HOWEVER IT DID NOT RESULT IN ANY ADVERSE EVENTS FOR THE PATIENT. THERE WAS A FIFTEEN (15) MINUTE SURGICAL DELAY HOWEVER THE CASE WAS FINISHED SUCCESSFULLY. THE PATIENT OUTCOME IS REPORTED AS WELL. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED TWO (2) DISTRACTION HOLDING SLEEVES/3.0MM K-WIRE IT WAS OBSERVED THAT THE SURFACE OF THE INSTRUMENTS ARE WORN AND DEFORMED FROM FREQUENT USING AS WELL AS THE BORE ARE WORN AND DEFORMED. THIS DEVICE CONDITION WAS EVALUATED AND DETERMINED TO BE REPORTABLE ON NOVEMBER 16, 2020. THIS REPORT IS FOR ONE (1) COMPRESSION/DISTRACTION HOLDING SLEEVE/3.0MM K-WIRE. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10985930 ↗
Other INSTRUMENT, COMPRESSION devices
- ATTUNE 2545-01-042Depuy (Ireland) · INSTRUMENT, COMPRESSION153 reports
- ATTUNE 2545-11-041Depuy (Ireland) · INSTRUMENT, COMPRESSION78 reports
- ATTUNE 2545-05-550Depuy (Ireland) · INSTRUMENT, COMPRESSION54 reports
- ATTUNE 2545-05-551Depuy (Ireland) · INSTRUMENT, COMPRESSION34 reports
- NA 03.111.021Synthes GmbH · INSTRUMENT, COMPRESSION32 reports
- ATTUNE 2545-05-552Depuy (Ireland) · INSTRUMENT, COMPRESSION27 reports
- NA 03.211.400Synthes GmbH · INSTRUMENT, COMPRESSION25 reports
- VIPER PRIME 286750083Medos International Sàrl · INSTRUMENT, COMPRESSION18 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.