The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886982072535), received between Mar 5, 2019 and Apr 22, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility5
  • Break2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient3
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Break

    DROP IN GUIDE HAS BROKEN AND IS UNUSABLE. GUIDE BLOCKS HAVE ALSO BROKEN WHERE THREADS ARE FOR THE SCREW. ISSUE OCCURRED DURING CONTINUED USE INTRAOPERATIVELY OVER TIME; WAS NOTICED POST-OP DURING REASSEMBLY. DEVICES HAVE BEEN DISCARDED. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957603 ↗
  • MalfunctionNov 1, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON 10/23/2024, THE GUIDE LOCK WOULD NOT ALLOW THE CORTICAL SCREWS TO LOCK.

    Patient problems: Insufficient Information

    MAUDE report 20592324 ↗
  • MalfunctionJan 29, 2024Material Deformation; Material Deformation

    DEVICE REPORT FROM SYNTHESE REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON JANUARY 11, 2024, WHEN THE J&J¿S DISTRIBUTOR WAS CHECKING THE VATCP INSTRUMENTS BEFORE SURGERY, IT NOTICED THAT THE SCREW THAT CONNECTS THE GUIDING BLOCK TO THE PLATE WAS BENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) GUIDING BLOCK F/2-COLUMN DIST-RAD-PL2.4 7 THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18602436 ↗
  • MalfunctionAug 31, 2023Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE AND MONTH IN 2023, THE TRAYS AND INSTRUMENTS WERE BROKEN. NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) GUIDE BLOCK FOR 2 COLUMN PLATE 7 HOLE HEAD/LEFT. THIS IS REPORT 3 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17662965 ↗
  • MalfunctionJul 17, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2020, DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE DISTAL RADIUS SYSTEM IMPLANTATION, WHILE INSERTING AN UNKNOWN VOLAR VARIABLE ANGLE (VA) PLATE, THE SURGEON WAS USING A GUIDE BLOCK ON THE UNKNOWN PLATE. THE SURGEON PREFERRED INSERTING PROVISIONAL 2.4MM CORTEX SCREWS IN THE DISTAL PORTION OF THE UNKNOWN PLATE THROUGH THE GUIDE BLOCK. HOWEVER, TWO (2) OF THE CORTEX SCREWS WOULD NOT FIT THROUGH THE GUIDE BLOCK. ONE OF THEM DID. THE SURGEON UNDERSTOOD THAT THE UNKNOWN VARIABLE ANGLE (VA) LOCKING SCREWS ARE TO BE USED IN THE DISTAL PORTION, CORTEX SCREWS HAVE BEEN USED IN THE PAST TO COMPRESS THE UNKNOWN PLATE PROVISIONALLY BEFORE REPLACING WITH AN UNKNOWN VA VARIABLE ANGLE (VA) LOCKING SCREWS. THE SURGEON INSERTED AN UNKNOWN VARIABLE ANGLE (VA) LOCKING SCREWS INSTEAD OF A CORTEX SCREWS. THE ISSUE WAS THAT THE 2.4MM CORTEX SCREWS HEADS ARE DIFFERENT DIAMETERS AND SOME WOULD NOT FIT THROUGH THE GUIDE BLOCK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF FIVE (5) MINUTES. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICE REPORTED: PLATE (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1), 2.4MM CORTEX SCREW SLF-TPNG WIT

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10289884 ↗
  • MalfunctionMay 21, 2019Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, SIX (6) GUIDING BLOCKS FOR A VARIABLE ANGLE LOCKING COMPRESSION PLATE (VA-LCP) TWO COLUMN DISTAL RADIUS PLATE DID NOT FIT CORRECTLY ON TWO (2) PLATES. A SCREW TO FIX THE GUIDING BLOCKS ON THE PLATE WAS TOO LONG RESULTING IN THE PLATE NOT BEING PLACED PROPERLY ON THE BONE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A PLATE WITH NO GUIDING BLOCK. THERE WAS NO PATIENT HARM OR INTERVENTION REQUIRED THIS REPORT IS FOR A VA-LCP PLATE. THIS IS REPORT 8 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8629138 ↗
  • MalfunctionMay 21, 2019Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, SIX (6) GUIDING BLOCKS FOR A VARIABLE ANGLE LOCKING COMPRESSION PLATE (VA-LCP) TWO COLUMN DISTAL RADIUS PLATE DID NOT FIT CORRECTLY ON TWO (2) PLATES. A SCREW TO FIX THE GUIDING BLOCKS ON THE PLATE WAS TOO LONG RESULTING IN THE PLATE NOT BEING PLACED PROPERLY ON THE BONE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A PLATE WITH NO GUIDING BLOCK. THERE WAS NO PATIENT HARM OR INTERVENTION REQUIRED THIS REPORT IS FOR A VA-LCP PLATE. THIS IS REPORT 7 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8629119 ↗
  • MalfunctionMar 5, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2019, THE SURGEON WAS UNABLE TO SCREW THE AIMING BLOCK FOR DISTAL RADIUS PLATE ONTO THE PLATE. THE ATTEMPT WAS MADE USING DIFFERENT PLATES. THE SURGEON SUGGESTED THAT THE SCREW MIGHT BE STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE: DISTAL RADIUS PLATE (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: UNKNOWN). THIS REPORT IS FOR A GUIDE BLOCK FOR 2 COLUMN PLATE 7 HOLE HEAD/LEFT. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Patient Involvement