NA 03.113.025Recalls & reported problems
Synthes GmbH · INSERTN HANDLE F/3.5MM LOW BND MEDIAL DISTAL TIB PLATE/RIGHT · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886982072917), received between May 1, 2017 and Jun 28, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Component Missing” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Component Missing4
- Device-Device Incompatibility4
- Material Deformation1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Patient Involvement4
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2024Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
IT WAS REPORTED THAT ON JUNE 3, 2024, THE SALES CONSULTANT WAS NOTIFIED BY CPD STAFF THAT THE SCREW THAT GOES INTO THE CONNECTING ARM WAS NOT SCREWING BACK IN. THE THREADS APPEAR TO BE STRIPPED AND WAS PROBABLY A RESULT OF MULTIPLE ATTEMPTS TO FIX. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR AN INSERTN HANDLE F/3.5MM LOW BND MEDIAL DISTAL TIB PLATE/RIGHT THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19638540 ↗MalfunctionApr 2, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A PROCEDURE, THE AIMING ARM FOR THE LOCKING COMPRESSION (LCP) LOW BEND MEDIAL DISTAL TIBIA PLATE WAS NOT LINING UP WITH THE PLATE PROPERLY. THERE WAS NO SURGICAL DELAY REPORTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) INSERTN HANDLE F/3.5MM LOW BND MEDIAL DISTAL TIB PLATE/RIGHT. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11610360 ↗MalfunctionMar 12, 2021Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
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
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11475469 ↗MalfunctionDec 31, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 DURING AN LCP MEDIAL DISTAL TIBIA INTERNAL FIXATION WITH JIG PROCEDURE, THE AIMING ARM AND INSERTION HANDLE DID NOT ALIGN WITH LCP HOLES. THE SURGEON WAS UNABLE TO USE THESE INSTRUMENTS FOR PERCUTANEOUS INSERTION. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN TIBIAL PLATE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) INSERTN HANDLE F/3.5MM LOW BND MEDIAL DISTAL TIB PLATE/RIGHT. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11102229 ↗MalfunctionJun 18, 2018Component Missing; Component Missing
IT WAS REPORTED THAT DURING SERVICE EVALUATION, THE SCREW OF A PERCUTANEOUS INSERTION HANDLE FOR AIMING ARM (RIGHT) WAS FOUND MISSING AND MIGHT REQUIRE REPLACEMENT. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7610733 ↗MalfunctionMay 24, 2018Component Missing; Component Missing
(B)(6) REPORTED FOLLOWING: IT WAS REPORTED THAT WHILE GOING THROUGH THE WASHING MACHINE DURING STERILIZATION, THE SCREW THAT TIGHTENS THE INSERTION HANDLE TO THE AIMING ARM MUST HAVE LOOSENED. IT WAS LOST AND COULD NOT BE LOCATED. THERE WAS NO SURGERY OR PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) INSERTN HANDLE F/3.5MM LOW BND MEDIAL DISTAL TIB PLATE/RIGHT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7542471 ↗MalfunctionJun 7, 2017Component Missing; Component Missing
IT WAS REPORTED THAT AN INSERTION SCREW FOR AN INSERTION HANDLE IS MISSING. NO PATIENT INVOLVEMENT REPORTED. THIS REPORT IS FOR ONE (1) INSERTION HANDLE. THIS IS REPORT 1 OF 1 FOR COMPLAINT COM-(B)(4).
Patient problems: No Patient Involvement
MAUDE report 6619838 ↗MalfunctionMay 1, 2017Component Missing; Component Missing
IT IS REPORTED THE INSERTION HANDLE IS MISSING THE CONNECTION SCREW COMPONENT. ISSUE WAS DISCOVERED DURING INSPECTION OF THE DEVICE. NO PATIENT OR SURGICAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) INSERTION HANDLE THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 6533556 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 03.113.025
Has the NA 03.113.025 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982072917), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 03.113.025?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between May 1, 2017 and Jun 28, 2024. In those reports, the most common specific device problems are “Component Missing” (4), “Device-Device Incompatibility” (4) and “Material Deformation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 03.113.025?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 03.113.025 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 03.113.025, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.113.025). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.091Synthes GmbH · GAUGE, DEPTH64 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page