Skip to content

LCP 03.120.001Recalls & reported problems

Synthes GmbH · PERIARTICULAR AIMING ARM FOR 4.5MM LCP PROX TIBIA PL-RT · Class 1

All models of LCP (367)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982073990), received between Sep 24, 2021 and Jun 27, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2024 THE PRIOR TO THE PROCEDURE, THE SCRUB TECH WAS ASSEMBLING THE AIMING ARM INSTRUMENTS. AFTER DETERMINING THE PLATE TO USE, THE SCRUB TECH CONNECTED THE AIMING ARM INSTRUMENTS TO THE PLATE IN ORDER TO "BUILD THE BOX". THIS WAS UNSUCCESSFUL. AFTER MULTIPLE ATTEMPTS OF DISASSEMBLING AND REASSEMBLING THE AIMING INSTRUMENTS, WE WERE UNABLE TO PROPERLY ALIGN THE AIMING ARM INSTRUMENTS WITH THE PLATE. WE PROCEEDED WITH THE PROCEDURE WITHOUT USING THE AIMING ARM INSTRUMENTATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAYS, NEGATIVE OUTCOMES, OR INTERVENTION REQUIRED. AFTER THE PROCEDURE WAS OVER, IT WAS DETERMINED THAT THE AIMING ARM INSTRUMENTATION WAS DEFECTIVE. THIS REPORT IS FOR ONE (1) PERIARTICULAR AIMING ARM FOR 4.5MM LCP PROX TIBIA PL-RT THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630161 ↗
  • MalfunctionJan 23, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A CASE, THE SURGEON WAS USING AN AIMING ARM, BUT IT DID NOT WORK. IT WAS NOT CONNECTING TO THE TWO PIECES AND IS A BOLT CONNECTING BOLT FOR THE AIMING ARM OF THE PERIARTIC AIM ARM INSTRUMENT THAT CONNECTS BOTH PIECES TOGETHER. IT WAS UNKNOWN IF THERE WERE ANY PATIENT CONSEQUENCES OR OUTCOMES. THIS REPORT IS FOR ONE (1) PERIARTIC AIM ARM 4.5 LCP PROX TIB PL-RT. THIS IS REPORT 1 OF 3 (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18566766 ↗
  • MalfunctionNov 15, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PERFORMING A PROXIMAL TIBIA, THE SURGEON FELT THERE WAS ISSUES WITH THE AIMING ARM BECAUSE THE THREADED TOWERS WERE NOT ABLE TO REACH THE PLATE. ONCE THE SURGEON PUSHED HARDER, THE TOWERS WERE ABLE TO ENGAGE INTO THE PLATE. IT APPEARS THAT WHEN THE PLATE WAS PULLED TO THE BONE WITH A SCREWS, IT PULLED THE PLATE AWAY FROM THE HANDLE AND FURTHER AWAY FROM THE TOWER. THE SURGERY WAS SUCCESSFULLY COMPLETED. SALES REPRESENTATIVE SPOKE WITH THE SURGEON AFTER THE CASE AND SAID THIS HANDLE REPLACED IN CASE THERE WAS AN ISSUES WITH AIMING ARM. THE HANDLE WAS NOT LINING UP WITH THE PLATE HOLES. IT WAS CHECKED PRIOR TO INSERTION AND IT LINED UP, HOWEVER INTRA-OP, AFTER PLATE WAS BROUGHT TO BONE , IT DID NOT LINE UP. NO ADVERSE EVENTS OCCURRED. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. CONCOMITANT DEVICE REPORTED: UNK - PLATES: TRAUMA (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN) AND UNK - SCREWS: TRAUMA (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN). THIS REPORT IS FOR ONE (1) PERIARTIC AIM ARM 4.5 LCP PROX TIB PL-RT. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15796777 ↗
  • MalfunctionSep 24, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT, WHEN THE CURVED CONDYLAR PLATE AND THE AIMING ARM WAS ATTACHED, IT WAS NOT ABLE TO ATTACH DUE TO THE STRIPPAGE OF THE SCREW HOLE OF THE PLATE. WHILE USING THE AIMING ARM ON THE NEW UNKNOWN PLATE, THE NEUTRAL SLEEVES WERE NOT CONNECTING TO THE HARPOON. THE PEDIATRIC AIMING ARM WAS OUTRIGGER. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH (10) MINUTES DELAY. NO PATIENT CONSEQUENCE. THIS REPORT IS FOR (1) PERIARTICULAR AIMING ARM FOR 4.5MM LCP PROX TIBIA PL-RT. THIS IS REPORT 7 OF 8 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12525825 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page