Skip to content

LCP 03.120.004Recalls & reported problems

Synthes GmbH · PERIARTICULAR AIMING ARM FOR 4.5MM LCP PROX TIBIA PL-LT · Class 1

All models of LCP (367)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982074027), received between Sep 6, 2019 and Jan 10, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Component Missing1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 10, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2023 DURING A PROCEDURE THE SURGEON WAS USING A 4.5 PROXIMAL TIBIA AIMING ARMS. ONE OF THE GUIDE SLEEVES WOULD NOT PROPERLY SNAP INTO THE AIMING ARM. THE AIMING ARM ALSO DID NOT PERFECTLY ALIGN WITH THE PLATE. THE SURGEON RECOMMENDED REPLACING THE INSERTION HANDLE AND AIMING ARM BECAUSE HE FELT THEY WERE NOT FUNCTIONING PROPERLY. THE SURGEON WAS ABLE TO COMPLETE THE SURGERY WITH NO DELAYS. THIS REPORT IS FOR ONE (1) AIM-ARM F/LCP PROX-TIB-PL4.5/5 LE THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18483616 ↗
  • MalfunctionFeb 8, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT THE AIMING ARM SLEEVE MECHANISM FEATURE DOES NOT WORK PROPERLY. THE RECESS (MECHANISM TO ATTACH THE SLEEVE) ARE USED AND DOESN'T WORK PROPERLY THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) PERIARTIC AIM ARM 4.5 LCP PROX TIB PL-LT. THIS IS REPORT 1 OF 1 FOR COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16329339 ↗
  • MalfunctionSep 6, 2019Component Missing; Component Missing

    IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) PROCEDURE ON AN UNKNOWN DATE TO TREAT A TIBIA FRACTURE, WHILE INSERTING AN UNKNOWN SCREWS THROUGH THE LOCKING COMPRESSION PLATE (LCP) TIBIA LEFT AIMING ARM, THE SCREWS WERE MISSING THE LCP TIBIAL PROXIMAL PLATE UPON INSERTION WITH A GUIDE SLEEVE WHEREIN THE GUIDE SLEEVE WAS FINE. IT WAS NOTICED THAT THE REPORTED AIMING ARM HAS TOO MUCH QUAINT IN IT DUE TO 20 YEARS OF USE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH TEN (10) MINUTES OF SURGICAL DELAY. PATIENT STATUS WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: LCP TIBIAL PROXIMAL PLATE (PART # 240.045, LOT # UNKNOWN, QUANTITY 1), GUIDE SLEEVE (PART # 03.120.014, LOT # UNKNOWN, QUANTITY 2), UNKNOWN SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN).

    Patient problems: No Patient Involvement

    MAUDE report 8971070 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page