LCP 03.122.007Recalls & reported problems

Synthes GmbH · AIMING ARM F/LCP PROXIMAL HUMERUS PLATE · Class 1

All models of LCP (367)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982074553), received between Jun 8, 2018 and Jul 29, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility6
  • Failure to Align1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Code Available1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE PHILOS AIMING ARM DOESN'T MATCH THE HOLES OF THE PHILOS PLATE. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. A NEW INSTRUMENT SET WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22641847 ↗
  • MalfunctionJun 9, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT TRAUMA SURGERY ON (B)(6) 2025. DURING THE PROCEDURE, THE PHILOS AIMING ARM DID NOT MEET THE SCREW HOLES OF THE PHILOS PLATE. ANOTHER AIMING ARM WAS USED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A DELAY OF TEN MINUTES. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22167647 ↗
  • MalfunctionMar 3, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE PHILOS AIMING ARM DID NOT WORK PROPERLY. THE SHAFT HOLES DID NOT MATCH THE PLATE HOLES. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21505769 ↗
  • MalfunctionJul 26, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM THE NETHERLANDS REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A MINIMALLY INVASIVE PERCUTANEOUS PLATING (MIPO) PROCEDURE ON (B)(6) 2023, THE PHILOS AIMING ARM DIDN'T AIM AT THE HOLES ON THE PHILOS PLATE. PROCEDURE WAS COMPLETED WITH A FIFTEEN MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR AN AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 2 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397401 ↗
  • MalfunctionJul 26, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM THE NETHERLANDS REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A MINIMALLY INVASIVE PERCUTANEOUS PLATING (MIPO) PROCEDURE ON (B)(6) 2023 THE PHILOS AIMING ARM DIDN'T AIM AT THE HOLES ON THE PHILOS PLATE. PROCEDURE WAS COMPLETED WITH A FIFTEEN MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR AN AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 1 OF 3 FOR PC-(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397151 ↗
  • MalfunctionApr 25, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PROXIMAL HUMERUS AIMING ARM MISALIGNED WITH THE PLATE CAUSING TROCARS TO MISFIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCES/OUTCOME. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16811582 ↗
  • InjuryJun 8, 2018Failure to Align; Failure to Align

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING THE SURGERY ON (B)(6) 2018, PROPERLY LOCKING OF IMPLANTS WITH AIMING ARM WAS NOT POSSIBLE. THE GUIDING DEVICE AND AIMING ARM WERE AFFECTED. IN A FIRST STEP GUIDING DEVICE HAD BEEN CHANGED. THIS DID NOT IMPROVE THE SITUATION. AIMING ARM WAS CHANGED IN A SECOND STEP. THE SURGERY WAS PROLONGED ABOUT 15 MINUTES AS THEY SWITCHED FROM LAPAROSCOPIC TO OPEN. SURGERY WAS COMPLETED SUCCESSFULLY. WHEN DEVICES WERE TESTED PRE-OPERATIVE, DRILL GUIDE IN GUIDING DEVICE AND HOLE DID NOT CORRESPOND. CONCOMITANT DEVICE REPORTED: PHILOS PLATE (PART# 441.903, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Code Available

    MAUDE report 7584313 ↗

Other Guide, surgical, instrument devices