LCP 03.122.007Recalls & reported problems
Synthes GmbH · AIMING ARM F/LCP PROXIMAL HUMERUS PLATE · Class 1
All models of LCP (367)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982074553), received between Jun 8, 2018 and Jul 29, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility6
- Failure to Align1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PHILOS AIMING ARM DOESN'T MATCH THE HOLES OF THE PHILOS PLATE. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. A NEW INSTRUMENT SET WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22641847 ↗MalfunctionJun 9, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATIENT UNDERWENT TRAUMA SURGERY ON (B)(6) 2025. DURING THE PROCEDURE, THE PHILOS AIMING ARM DID NOT MEET THE SCREW HOLES OF THE PHILOS PLATE. ANOTHER AIMING ARM WAS USED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A DELAY OF TEN MINUTES. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167647 ↗MalfunctionMar 3, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PHILOS AIMING ARM DID NOT WORK PROPERLY. THE SHAFT HOLES DID NOT MATCH THE PLATE HOLES. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21505769 ↗MalfunctionJul 26, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM THE NETHERLANDS REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A MINIMALLY INVASIVE PERCUTANEOUS PLATING (MIPO) PROCEDURE ON (B)(6) 2023, THE PHILOS AIMING ARM DIDN'T AIM AT THE HOLES ON THE PHILOS PLATE. PROCEDURE WAS COMPLETED WITH A FIFTEEN MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR AN AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17397401 ↗MalfunctionJul 26, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM THE NETHERLANDS REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A MINIMALLY INVASIVE PERCUTANEOUS PLATING (MIPO) PROCEDURE ON (B)(6) 2023 THE PHILOS AIMING ARM DIDN'T AIM AT THE HOLES ON THE PHILOS PLATE. PROCEDURE WAS COMPLETED WITH A FIFTEEN MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR AN AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 1 OF 3 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17397151 ↗MalfunctionApr 25, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PROXIMAL HUMERUS AIMING ARM MISALIGNED WITH THE PLATE CAUSING TROCARS TO MISFIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCES/OUTCOME. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16811582 ↗InjuryJun 8, 2018Failure to Align; Failure to Align
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING THE SURGERY ON (B)(6) 2018, PROPERLY LOCKING OF IMPLANTS WITH AIMING ARM WAS NOT POSSIBLE. THE GUIDING DEVICE AND AIMING ARM WERE AFFECTED. IN A FIRST STEP GUIDING DEVICE HAD BEEN CHANGED. THIS DID NOT IMPROVE THE SITUATION. AIMING ARM WAS CHANGED IN A SECOND STEP. THE SURGERY WAS PROLONGED ABOUT 15 MINUTES AS THEY SWITCHED FROM LAPAROSCOPIC TO OPEN. SURGERY WAS COMPLETED SUCCESSFULLY. WHEN DEVICES WERE TESTED PRE-OPERATIVE, DRILL GUIDE IN GUIDING DEVICE AND HOLE DID NOT CORRESPOND. CONCOMITANT DEVICE REPORTED: PHILOS PLATE (PART# 441.903, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) AIMING ARM F/LCP PROXIMAL HUMERUS PLATE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7584313 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page