NA 03.130.250Recalls & reported problems
Synthes GmbH · DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Synthes depth gauge is a surgical measuring instrument used to check the depth of holes drilled for small screws. The 43 reports mostly describe the tip bending, warping or snapping off, sometimes breaking into pieces; in most cases nothing happened to the patient, though a few mention surgery plans being changed or the patient being involved. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 10886982075420), received between May 1, 2017 and May 1, 2026 — classified as 42 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (31 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation31
- Break10
- Bent1
- Device Operates Differently Than Expected1
- Incorrect Measurement1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions35
- No Consequences Or Impact To Patient5
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2026Material Deformation
IT WAS REPORTED THAT DEPTH GAUGE F/SCR Ø1.3 - 2 MEAS-RANGE AND DEPTH GAUGE F/SCR Ø2+2.4 MEAS-RANGE UP-T EXPERIENCED FAILURES. THE EVENT INVOLVED THE TIPS BREAKING OFF TWO DEPTH GAUGES AND A THIRD BEING SEVERELY BENT WHILE ATTEMPTING TO FIND THE FAR HOLE DURING A CASE. THE DEVICES WERE SEPARATED INTO TWO OR MORE PIECES, AND THE EVENT WAS IDENTIFIED AS INVOLVING NO PATIENT OR USER IMPACT, WITH NO OBSERVABLE CLINICAL SYMPTOMS OR CHANGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25056363 ↗MalfunctionJan 8, 2026Material Deformation
IT WAS REPORTED THAT THE DEPTH GAUGE HAD A BENT TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24002516 ↗MalfunctionDec 15, 2025Break; Break
THIS DEVICE IS BROKEN WITHOUT CONSEQUENCE FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23802410 ↗MalfunctionOct 22, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE DISTAL END OF THE 1.3MM/1.5MM DEPTH GAUGE WAS BENT AND WAS ASKED TO BE REPLACED BY THE SURGEON. THERE WAS NO DELAY IN SURGERY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23359615 ↗MalfunctionOct 15, 2025Break; Break
IT WAS REPORTED THAT ON (B)(6) 2025, SURGEON NEEDED TO USE THE VARIABLE ANGLE HAND TRAY FOR A METACARPAL FIXATION CASE. DEPTH GAUGE BROKE WHEN SCRUB TECH WAS CHECKING IT. HAD TO CHANGE PLANS AND USE A DIFFERENT PLATE AND USE A DIFFERENT DEPTH GAUGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23292522 ↗MalfunctionSep 10, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2025, A DEPTH GAUGE WAS BENT AND SURGEON WAS UNABLE TO BEND IT BACK. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23019928 ↗MalfunctionSep 2, 2025Break; Break
IT WAS REPORTED THAT THE DEPTH GAUGE WAS BROKEN. IT WAS NOTICED DURING A CASE AND PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22938525 ↗MalfunctionAug 29, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DEPTH GAUGE TIPS SEVERELY BENT. NO SURGICAL DELAY WAS REPORTED. NO PATIENT HARM WAS THERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22921294 ↗MalfunctionAug 27, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DEPTH GAUGE TIPS SEVERELY BENT. NO SURGICAL DELAY WAS REPORTED. NO PATIENT HARM WAS THERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22902344 ↗MalfunctionJun 17, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT ON (B)(6) 2025, THERE WERE DEVICE MALFUNCTIONS.; KNOB BROKE, BENT ROD, STRIPPED ROD. IT IS UNKNOWN WHERE EVENT OCCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22238205 ↗MalfunctionJun 13, 2025Material Deformation; Material Deformation
IT WAS REPORTED THE TIP PART OF THE DEPTH GAUGE IS BENT. ISSUE IS UNREPAIRABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22205774 ↗MalfunctionJun 3, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE NEEDLE ON THE BLACK DEPTH GAUGE HAS BEEN BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22133410 ↗MalfunctionJan 29, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE INSTRUMENT WAS TOO BENT TO BE USED. THESE ISSUES WERE IDENTIFIED DURING PRE-OP TESTING OF THE EQUIPMENT AND DID NOT IMPACT PATIENT OUTCOMES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21266426 ↗MalfunctionOct 4, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, IT WAS NOTICED THE ITEMS WERE BENT AT THE START OF THE SURGERY. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20371241 ↗MalfunctionAug 15, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON (B)(6) 2024 THE ARM HAD GOTTEN BENT WHILE INSTALLING THE TWO PIECES AFTER CLEANING. THERE WAS NO PATIENT OUTCOME REPORTED. THIS REPORT INVOLVES ONE (1) DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19992862 ↗MalfunctionMay 1, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CLAMP WOULDN'T STAY CLOSED, AND THE DEPTH GAUGE WAS BENT. THE EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19224076 ↗MalfunctionFeb 12, 2024Break; Break
IT WAS REPORTED THAT ON JAN 31, 2024, THE HOOK OF THE DEVICE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 1.3/1.5 AND 2.0 SCREWS THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18691230 ↗MalfunctionDec 14, 2023Mechanical Jam; Mechanical Jam
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN NEW ZEALAND AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGEON WAS USING THE VA HAND SET CONSIGNED TO THE HOSPITAL, THE SCREW COULD NOT BE MEASURED BECAUSE THE DEPTH GAUGE WAS JAMMED. ANOTHER DEPTH GAUGE FROM A DIFFERENT SET WAS USED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE (1) DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18327318 ↗MalfunctionNov 28, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT ON (B)(6) 2023, WHILE INSPECTING THE INSTRUMENTATION, AFTER GOING THROUGH THE DECONTAMINATION PROCESS, SALES REP NOTICED THAT THERE WERE SEVERAL INSTRUMENTS THAT WERE DAMAGED. THERE WILL BE A TOTAL OF EIGHT PRODUCTS ON THIS COMPLAINT. THERE WAS A 5 MM FLEXIBLE SCREWDRIVER, AND THE TIP OF THE SCREWDRIVER IS BROKE OFF. THERE WAS A 17 MM FLEXIBLE MONOBLOC REAMER, AND THIS ITEM APPEARS TO BE TWISTED IN A MANNER THAT HAS HAD SEVERE TORQUE APPLIED TO IT. THERE WAS THE SELF RETAINING PIN FOR THE POWER SCREWDRIVER. THE TIP OF IT HAS THREADS ON IT, AND THE THREADS WERE BROKE OFF. THERE WAS A 2.5 MM SCREWDRIVER, AND THE TIP OF THE SCREWDRIVER WAS STRIPPED. THERE WAS A 2.8 MM DRILL GUIDE, AND THE THREADS ON IT WERE STRIPPED. THERE WAS A HANDLE FOR THE 2.8 MM DRILL GUIDE AND THE INNER THREADS ON THIS HANDLE ARE STRIPPED. THE TIP OF THE DEPTH GAUGE IS BROKE OFF. THERE WAS ANOTHER DEBT GAUGE IN THE TIP OF THE DEPTH GAUGE IS BENT. ALL OF THESE ITEMS WERE SC STOCK AND BELONG IN FIELD INVENTORY SETS. ALL OF THESE ITEMS NEEDED TO BE REPLACED SO THAT THE FIELD INVENTORY SETS WOULD BE FULL. THERE WAS NO PATIENT INVOLVEMENT IN ANY OF THESE ITEMS. ALL ITEMS WERE ALSO AVAILABLE TO BE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18221427 ↗MalfunctionAug 22, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2023, THE DEPTH GAUGE WAS BENT AND NOT FUNCTIONAL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO PATIENT CONSEQUENCE WAS REPORTED. THIS REPORT INVOLVES ONE DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17595993 ↗MalfunctionMar 22, 2023Break; Material Deformation; Break; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE DEPTH GAUGE IN QUESTION WAS DEFORMED WHEN THE DISTRIBUTOR INSPECTED BEFORE DELIVERY TO HOSPITAL. THE DISTRIBUTOR INQUIRED WHETHER THE DEVICE APPEARED TO HAVE BEEN BENT INTENTIONALLY AND WHETHER IT WOULD PASS YOUR INSPECTION STANDARDS. THE DELIVERY WAS MADE IN THE DEFORMED STATE BECAUSE CLOSING TIME FOR DELIVERY TO THE HOSPITAL WAS APPROACHING. DURING CLEANSING IN THE OPERATION ROOM, THE TIP OF THE DEPTH GAUGE WAS BROKEN OFF. A DIFFERENT STANDARD DEPTH GAUGE WILL BE USED FOR SURGERY. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16590947 ↗MalfunctionMar 9, 2023Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHESE REPORTS AN EVENT IN UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING REPROCESSING DEPTH GAUGE WAS BENT. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16513091 ↗MalfunctionFeb 13, 2023Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT THE TIP PART WAS BENT ON THE DEPTH GAUGE. ISSUE AN UNREPAIRABLE LETTER. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) DEPTH GAUGE F/SCR Ø1.3 - 2 MEAS-RANGE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16363932 ↗MalfunctionFeb 2, 2023Material Deformation; Material Deformation
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PIN ON THE DEPTH GAUGE WAS BENT. THIS REPORT IS FOR A DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16287361 ↗MalfunctionJan 12, 2023Material Deformation; Material Deformation
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIP OF THE DEPTH GAUGE IN QUESTION WAS FOUND TO BE DEFORMED. THERE WAS NO REPORTED PATIENT INTERACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16150389 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 03.130.250
Has the NA 03.130.250 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982075420), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 03.130.250?
The FDA's MAUDE database lists 42 adverse event reports for this device, received between May 1, 2017 and May 1, 2026. In those reports, the most common specific device problems are “Material Deformation” (31), “Break” (10) and “Bent” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 03.130.250?
None of the reports are classified as deaths, 0 as injuries and 42 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 03.130.250 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 03.130.250, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.130.250). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.091Synthes GmbH · GAUGE, DEPTH64 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record