NA 04.005.436Recalls & reported problems
Synthes GmbH · 4.0MM TI LOCKING SCREW W/T25 STARDRIVE 46MM FOR IM NAILS · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982085481), received between Dec 28, 2017 and Jul 22, 2022 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K000089, K103002); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Break1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Loss of Range of Motion2
- Fatigue1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A RELATED RESEARCH ACTIVITY DATABASE (DDRA) [SUMMARY OF RETROSPECTIVE DATA FOR PATIENTS TREATED WITH THE LATERAL FEMORAL NAIL, ADOLESCENT LATERAL FEMORAL NAIL, AND THE FEMORAL RECON NAIL]. BETWEEN JANUARY 2014 AND OCTOBER 2021, A TOTAL OF 300 PATIENTS (219 PATIENTS TREATED WITH LFN, 56 PATIENTS TREATED WITH ALFN, AND 25 PATIENTS TREATED WITH FRN) WERE INCLUDED IN THE REPORT IF THEY HAD EITHER 9 MONTHS OF FOLLOW-UP OR EVIDENCE OF HEALING. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED IN THE BRONSON METHODIST HOSPITAL¿S ORTHOPEDIC INPATIENT DATABASE: LFN GROUP: PATIENT 5: A 32-YEAR-OLD MALE PATIENT HAD A WEAKNESS IN LEG, WHICH RESOLVED WITH PHYSICAL THERAPY. PATIENT 6: AN 81-YEAR-OLD MALE PATIENT HAD A PERSISTENT SWELLING ABOUT KNEE 1.5 YEARS POSTOPERATIVELY, COMPRESSION. PATIENT 8: A 58-YEAR-OLD FEMALE PATIENT HAD A PULMONARY EMBOLISM. PATIENT 9: A 35-YEAR-OLD FEMALE PATIENT HAD DIFFICULT REDUCTION INTRAOPERATIVELY AND HAD RESIDUAL SORENESS AT INCISION SITE POSTOPERATIVELY. PATIENT 13: A 16-YEAR-OLD MALE PATIENT HAD PAIN WITH IMPLANTS LEADING TO REHOSPITALIZATION FOR ADJUNCTIVE PLATE REMOVAL, L
Patient problems: Fatigue
MAUDE report 15084761 ↗InjuryJul 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A RELATED RESEARCH ACTIVITY DATABASE (DDRA) [SUMMARY OF RETROSPECTIVE DATA FOR PATIENTS TREATED WITH THE LATERAL FEMORAL NAIL, ADOLESCENT LATERAL FEMORAL NAIL, AND THE FEMORAL RECON NAIL]. BETWEEN JANUARY 2014 AND OCTOBER 2021, A TOTAL OF 300 PATIENTS (219 PATIENTS TREATED WITH LFN, 56 PATIENTS TREATED WITH ALFN, AND 25 PATIENTS TREATED WITH FRN) WERE INCLUDED IN THE REPORT IF THEY HAD EITHER 9 MONTHS OF FOLLOW-UP OR EVIDENCE OF HEALING. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED IN THE BRONSON METHODIST HOSPITAL¿S ORTHOPEDIC INPATIENT DATABASE: LFN GROUP: PATIENT 5: A 32-YEAR-OLD MALE PATIENT HAD A WEAKNESS IN LEG, WHICH RESOLVED WITH PHYSICAL THERAPY. PATIENT 6: AN 81-YEAR-OLD MALE PATIENT HAD A PERSISTENT SWELLING ABOUT KNEE 1.5 YEARS POSTOPERATIVELY, COMPRESSION. PATIENT 8: A 58-YEAR-OLD FEMALE PATIENT HAD A PULMONARY EMBOLISM. PATIENT 9: A 35-YEAR-OLD FEMALE PATIENT HAD DIFFICULT REDUCTION INTRAOPERATIVELY AND HAD RESIDUAL SORENESS AT INCISION SITE POSTOPERATIVELY. PATIENT 13: A 16-YEAR-OLD MALE PATIENT HAD PAIN WITH IMPLANTS LEADING TO REHOSPITALIZATION FOR ADJUNCTIVE PLATE REMOVAL, L
Patient problems: Loss of Range of Motion
MAUDE report 15084654 ↗InjuryJul 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A RELATED RESEARCH ACTIVITY DATABASE (DDRA) [SUMMARY OF RETROSPECTIVE DATA FOR PATIENTS TREATED WITH THE LATERAL FEMORAL NAIL, ADOLESCENT LATERAL FEMORAL NAIL, AND THE FEMORAL RECON NAIL]. BETWEEN JANUARY 2014 AND OCTOBER 2021, A TOTAL OF 300 PATIENTS (219 PATIENTS TREATED WITH LFN, 56 PATIENTS TREATED WITH ALFN, AND 25 PATIENTS TREATED WITH FRN) WERE INCLUDED IN THE REPORT IF THEY HAD EITHER 9 MONTHS OF FOLLOW-UP OR EVIDENCE OF HEALING. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED IN THE BRONSON METHODIST HOSPITAL¿S ORTHOPEDIC INPATIENT DATABASE: LFN GROUP: PATIENT 5: A 32-YEAR-OLD MALE PATIENT HAD A WEAKNESS IN LEG, WHICH RESOLVED WITH PHYSICAL THERAPY. PATIENT 6: AN 81-YEAR-OLD MALE PATIENT HAD A PERSISTENT SWELLING ABOUT KNEE 1.5 YEARS POSTOPERATIVELY, COMPRESSION. PATIENT 8: A 58-YEAR-OLD FEMALE PATIENT HAD A PULMONARY EMBOLISM. PATIENT 9: A 35-YEAR-OLD FEMALE PATIENT HAD DIFFICULT REDUCTION INTRAOPERATIVELY AND HAD RESIDUAL SORENESS AT INCISION SITE POSTOPERATIVELY. PATIENT 13: A 16-YEAR-OLD MALE PATIENT HAD PAIN WITH IMPLANTS LEADING TO REHOSPITALIZATION FOR ADJUNCTIVE PLATE REMOVAL, L
Patient problems: Loss of Range of Motion
MAUDE report 15084637 ↗InjurySep 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, THE HARDWARE FOR A FRACTURE OF RIGHT TIBIAL IMPLANTS WAS REMOVED DUE TO THE INFECTION. THE FOLLOWING IMPLANTS WERE REMOVED, ONE (1) 4.0MM TITANIUM LOCKING SCREWS, TWO (2) 4.0MM TITANIUM LOCKING SCREWS, AND ONE (1) TITANIUM CANNULATED TIBIAL NAIL AND ALL WERE INTACT(NO BROKEN IMPLANTS AND NO FRAGMENTS). PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS DOING WELL POST-OPERATIVELY. CONCOMITANT DEVICE REPORTED: 4.0MM TITANIUM LOCKING SCREWS (PART:04.005.434;LOT:UNKNOWN; QUANTITY:1), 4.0MM TITANIUM LOCKING SCREWS (PART:04.005.436;LOT:UNKNOWN; QUANTITY:1). THIS REPORT IS FOR ONE (1) 4.0MM TI LOCKING SCREW W/T25 STARDRIVE 46MM FOR IM NAILS. THIS IS REPORT 3 OF 4 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12521460 ↗InjurySep 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, THE HARDWARE FOR A FRACTURE OF RIGHT TIBIAL IMPLANTS WAS REMOVED DUE TO THE INFECTION. THE FOLLOWING IMPLANTS WERE REMOVED, ONE (1) 4.0MM TITANIUM LOCKING SCREWS, TWO (2) 4.0MM TITANIUM LOCKING SCREWS, AND ONE (1) TITANIUM CANNULATED TIBIAL NAIL AND ALL WERE INTACT(NO BROKEN IMPLANTS AND NO FRAGMENTS). PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT WAS DOING WELL POST-OPERATIVELY. CONCOMITANT DEVICE REPORTED: 4.0MM TITANIUM LOCKING SCREWS (PART:04.005.434;LOT:UNKNOWN; QUANTITY:1), 4.0MM TITANIUM LOCKING SCREWS (PART:04.005.436;LOT:UNKNOWN; QUANTITY:1). THIS REPORT IS FOR ONE (1) 4.0MM TI LOCKING SCREW W/T25 STARDRIVE 46MM FOR IM NAILS. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12521325 ↗InjurySep 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2019, DURING AN UNKNOWN PROCEDURE, IT WAS FOUND OUT THAT AN UNKNOWN HUMERAL NAIL SET DO NOT HAVE AN UNKNOWN CADDY WITH AN UNKNOWN BLADES, UNKNOWN 4.0 TITANIUM SCREWS AND UNKNOWN END CAPS. ORIGINALLY, THE SALES CONSULTANT GETS THE SET AND THERE¿S GOING TO BE A TOTE WITH IMPLANTS FOR BLADES AND SCREWS AND END CAPS. THE SALES CONSULTANT CALLED THE FSL AND EVENTUALLY DISCOVERED THAT THERE WAS SUPPOSED TO BE A CADDY INSIDE THE SET. THE SALES CONSULTANT THEN GO AND DRIVE TO ANOTHER HOSPITAL AND RETURN WITH A TOTE FOR SAID IMPLANTS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED; HOWEVER, THE PATIENT SPENT AN EXTRA 100 MINUTES UNDER ANESTHESIA. ALSO, THE END CAP AVAILABLE FROM THE OTHER HOSPITAL WAS NOT THE CORRECT SIZE, SO IT PROTRUDED FROM THE PATIENT¿S HUMERAL HEAD. RE-OPERATION IS MOST LIKELY. PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. THIS COMPLAINT INVOLVES THIRTY-EIGHT (37) DEVICES. (B)(4).
Patient problems: No Code Available
MAUDE report 9076158 ↗MalfunctionDec 28, 2017Break; Break
PART NUMBER IDENTIFIED AS 03.010.109 FOR CONCOMITANT PART.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7152443 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K000089, K103002. One submission often covers many models, so check the product and lot details in each record.
- Z-2060-2012 · May 22, 2012 · Terminated4.5mm TI Multiloc Screw Length 28MM-Sterile Expiration: 04/29/2012 REF 04.019.028S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu
- Z-2059-2012 · May 22, 2012 · Terminated4.5mm TI Multiloc Screw Length 38MM-Sterile Expiration: 04/29/2012 REF 04.019.038S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page