EXPert Nail 04.013.049Recalls & reported problems
Synthes GmbH · TI SPIRAL BLADE 85MM FOR TI RETROGRADE FEMORAL NAILS-EX · Class 2 · Implant
All models of EXPert Nail (151)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982087522), received between Apr 16, 2021 and Apr 16, 2021 — classified as 1 injury.
1 other recall cover devices cleared under the same FDA submission (K033618); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Nonunion/Delayed-union Bone Fracture1
- Post Operative Wound Infection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021 A RETROGRADE NAIL EXCHANGE WAS DONE BECAUSE OF A POSSIBLE INFECTION AND NONUNION. THE ORIGINAL INJURY WAS DONE DUE TO A GUNSHOT TO THE DISTAL FEMUR. THE NAIL AND HARDWARE WERE SUCCESSFULLY EXCHANGED. THERE WAS ONE (1) 5.0MM BROKEN SCREW PROXIMALLY THAT WAS SUCCESSFULLY REMOVED. WHILE INSERTING THE NEW NAIL, THE SURGEON HIT THE NAIL WITH THE SECOND MOST DISTAL SCREW DRILL BIT. HOWEVER, ADJUSTMENTS WERE MADE AND THE SURGEON WAS ABLE TO PLACE THE SCREW SUCCESSFULLY. THE SURGEON WAS LEANING ON THE AIMING ARM CAUSING IT TO BEND AND MISS THE NAIL. THERE WAS A TWENTY (20) MINUTES SURGICAL DELAY. PATIENT OUTCOME IS REPORTED AS GOOD. THIS REPORT CAPTURES THE POST-OP EVENT OF POSSIBLE INFECTION AND NONUNION WHILE RELATED COMPLAINT (B)(4) CAPTURES THE INTRA-OP EVENT DURING INSERTION OF THE NEW NAIL THE SURGEON HIT THE NAIL WITH THE SECOND MOST DISTAL SCREW DRILL BIT. THIS REPORT IS FOR ONE (1) TI SPIRAL BLADE 85MM FOR TI RETROGRADE FEMORAL NAILS-EX THIS IS REPORT 3 OF 6 FOR COMPLAINT (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture; Post Operative Wound Infection
MAUDE report 11682383 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033618. One submission often covers many models, so check the product and lot details in each record.
- Z-2058-2013 · Nov 2, 2012 · TerminatedSynthes Trauma Nail System. The devices are indication for bone fixation.
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page