Skip to content

MULTILOC 04.016.039SRecalls & reported problems

Synthes GmbH · 9.5MM TI MULTILOC PROX HUMERAL NAIL/LT/CANN/160MM-STER · Class 2 · Implant

All models of MULTILOC (48)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982089212), received between Mar 18, 2022 and Sep 22, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K103002); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Manufacturing, Packaging or Shipping Problem1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862692 ↗
  • MalfunctionApr 11, 2026Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE WHEN THE 9.5MM RIGHT SIDE MULTILOC PROX HUMERUS NAIL PACKAGE WAS OPENED IT WAS FOUND THAT A LEFT SIDE NAIL WAS PACKAGED IN THE BOX BUT HAD A RIGHT SIDE LABEL. THE PATIENT WAS NOT AFFECTED. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24851719 ↗
  • InjuryMar 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023 THE PATIENT UNDERWENT MULTI LOCKING FIXATION FOR HUMERUS SURGICAL NECK FRACTURE. THE DISTAL LOCKING SCREW LOOSENED, AND THE FRACTURE SITE BEGAN TO OPEN. IT WAS PLANNED TO REMOVE THE NAIL AND REPLACE WITH A REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2024. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 9.5MM TI MULTILOC PROX HUMERAL NAIL/LT/CANN/160MM-STER. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 18980444 ↗
  • MalfunctionFeb 12, 2024Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2024 THE PATIENT UNDERWENT AN ORIF SURGERY WITH MHN IMPLANTS FOR A FRACTURE OF THE LEFT PROXIMAL HUMERUS. THE SURGEON USED A DRILL SLEEVE FOR THE DISTAL WHEN HE DRILLED THE PROXIMAL AP HOLE. GAPS BETWEEN THE DRILL SLEEVE AND THE HOLE CAUSED THE DRILL HOLE TO SHIFT AND THE DRILL BIT INTERFERED WITH THE NAIL. THE SURGEON DRILLED MANUALLY AT THE SURGEON¿S DISCRETION AND SUCCEEDED IN MAKING A HOLE. THE SCREW WAS INSERTED SUCCESSFULLY. WHEN THE SURGEON WAS INSERTING DISTAL LOCKING SCREWS, DRILLING WAS DONE SMOOTHLY FOR ONE OF THE LOCKING SCREWS, AND THE DRILL BIT WAS LEFT FOR FIXATION. HOWEVER, ANOTHER DRILL BIT ALSO INTERFERED WITH THE NAIL WHEN IT WAS USED TO DRILL A HOLE FOR A 2ND DISTAL LOCKING SCREW. THE SURGEON DRILLED MANUALLY FOR THE HOLE AND INSERTED THE DISTAL LOCKING SCREW SUCCESSFULLY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. NO FRAGMENTS WERE LEFT IN THE PATIENT, THE SURGEON CONFIRMED THIS VIA X-RAY. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 9.5MM TI MULTILOC PROX HUMERAL NAIL/LT/CANN/160MM-STER. THIS IS RE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18687586 ↗
  • MalfunctionMar 18, 2022Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE POST OP INSPECTION OF SCREW RACK IS PEELING, ALONG WITH NON LOCKING PLATE SET CUSHIONS. MULTILOCK NAIL PACKAGING WAS ALSO TORN DURING TRANSPORT. PATIENTS INVOLVEMENT IS THERE THIS COMPLAINT INVOLVES (1) DEVICE 9.5 TI MULTILOC PRX HUM NL/LT/CANN/160-S. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13804466 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K103002. One submission often covers many models, so check the product and lot details in each record.

  • Z-2059-2012 · May 22, 2012 · Terminated
    4.5mm TI Multiloc Screw Length 38MM-Sterile Expiration: 04/29/2012 REF 04.019.038S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu
  • Z-2060-2012 · May 22, 2012 · Terminated
    4.5mm TI Multiloc Screw Length 28MM-Sterile Expiration: 04/29/2012 REF 04.019.028S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page