MULTILOC 04.019.040Recalls & reported problems
Synthes GmbH · 4.5MM TI MULTILOC SCREW LENGTH 40MM · Class 2 · Implant
All models of MULTILOC (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982089823), received between Sep 5, 2018 and May 17, 2019 — classified as 2 injuries.
2 other recalls cover devices cleared under the same FDA submission (K103002); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- No Code Available2
- Pain2
- Necrosis1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT HAD HARDWARE REMOVAL OF ONE (1) MULTILOC HUMERAL NAIL, ONE (1) MULTILOC SCREW LENGTH 50MM, ONE (1) MULTILOC SCREW LENGTH 42MM, ONE (1) MULTILOC SCREW LENGTH 48MM, ONE (1) MULTILOC SCREW LENGTH 40MM, ONE (1) TITANIUM LOCKING SCREW LENGTH 34MM, AND ONE (1) TITANIUM LOCKING SCREW LENGTH 36MM DUE TO SHOULDER PAIN AND LIMITED MOBILITY OF HUMERUS. THERE WERE FOUR PART PROXIMAL HUMERUS FRACTURE FIXED WITH THE NAIL BUT THE BONE HAD GONE ON TO DIE. THE SURGEON REMOVED ALL THE IMPLANTS BECAUSE THE HEAD JUST FELL INTO VARIOUS COLLAPSE AND TAKE THE DEAD BONY FRAGMENTS. INITIAL SURGERY WAS DONE ON (B)(6) 2018. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 4.5MM TI MULTILOC SCREW 40MM. THIS IS REPORT 5 OF 7 FOR COMPLAINT (B)(4).
Patient problems: Necrosis; No Code Available; Pain
MAUDE report 8621281 ↗InjurySep 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT HARDWARE REMOVAL OF THREE (3) TITANIUM (TI) MULTILOC SCREWS DUE TO LEFT SHOULDER PAIN. ALL OF THE SCREWS WERE REMOVED INTACT WITHOUT ANY PROBLEM. THE MULTILOC HUMERAL NAIL WAS LEFT IN THE PATIENT. THE SURGEON DECIDED TO LEAVE THE NAIL IN PLACE AS IT WAS NOT STICKING OUT NOR WAS THERE A MALFUNCTION WITH AND HE FELT THE SCREWS WERE THE CAUSE OF THE PAIN. IT WAS NOT PROVIDED IF THERE WAS ANY OTHER MEDICAL INTERVENTION DONE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. PATIENT OUTCOME WAS NOT PROVIDED. CONCOMITANT DEVICE: MULTILOC HUMERAL NAIL (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A 4.5MM TI MULTILOC SCREW LENGTH 40MM. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Code Available; Pain
MAUDE report 7847416 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103002. One submission often covers many models, so check the product and lot details in each record.
- Z-2059-2012 · May 22, 2012 · Terminated4.5mm TI Multiloc Screw Length 38MM-Sterile Expiration: 04/29/2012 REF 04.019.038S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu
- Z-2060-2012 · May 22, 2012 · Terminated4.5mm TI Multiloc Screw Length 28MM-Sterile Expiration: 04/29/2012 REF 04.019.028S Manufactured or Distributed by Synthes (USA) 1101 Synthes Avenue, Monument CO 80132 4.5MM TI Mu
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page