TFN-ADVANCE 04.037.024SRecalls & reported problems
Synthes GmbH · 10MM/125 DEG TI CANN TFNA 340MM/RIGHT - STERILE · Class 2 · Implant
All models of TFN-ADVANCE (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982095824), received between Jan 16, 2019 and Feb 11, 2022 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- No Code Available2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Impaired Healing1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 11, 2022Break; Break
IT WAS REPORTED ON (B)(6) 2022, THE PATIENT UNDERWENT REMOVAL OF A BROKEN TITANIUM CANNULATED PROXIMAL FEMORAL NAILING SYSTEM (TFNA). THE ORIGINAL SURGERY WAS DONE IN JUNE AND POST-OP IMAGES WERE IN GOOD CONDITION. THIS IS THE SECOND FAILURE THAT THE SURGEON REPORTED ON (B)(6) 2022. THE OTHER FAILURE WAS A SUB-TROCHANTERIC NONUNION. THE NAIL WAS IN FOR OVER A YEAR BEFORE IT FAILED. THE PATIENT WAS REPORTED AS GOOD. THIS REPORT INVOLVES ONE (1) 10MM/125 DEG TI CANN TFNA 340MM/RIGHT - STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4). THIS PRODUCT COMPLAINT, (B)(4), IS RELATED TO (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13513916 ↗InjuryJan 16, 2019Break; Break
01/16/2019. UPDATED EVENT DESCRIPTION: IT WAS REPORTED THAT THE CASE WAS COMPLETED ON (B)(6) 2017. THE TROCHANTERIC FEMORAL NAIL ADVANCE (TFNA) LAG SCREW AND UNKNOWN DISTAL LOCKING SCREW BROKE IN THE TFNA NAIL DURING AN UNKNOWN PROCEDURE. A PROPOSED REMOVAL PROCEDURE FOR A TOTAL HIP ARTHROPLASTY IS YET TO BE SCHEDULED. THERE WAS AN UNKNOWN SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVED THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) 10MM/125 DEG TI CANN TFNA 340MM/RIGHT - STERILE. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 8254067 ↗InjuryJan 16, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT HARDWARE REMOVAL DUE TO NONUNION AND A BROKEN TITANIUM CANNULATED TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) OF THE RIGHT HIP. ALL HARDWARE WAS REMOVED AND REVISED. FRAGMENTS WERE GENERATED BUT WERE EASILY REMOVED. THE DEVICES WERE ORIGINALLY IMPLANTED ON (B)(6) 2018. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY. PATIENT STATUS IS STABLE. CONCOMITANT DEVICES: TFNA SCREW 80MM (PART: 04.038.080S, LOT: 9840313, QUANTITY: 1). 5.0MM LOCKING SCREW 42MM (PART: 04.005.532, LOT: UNKNOWN, QUANTITY: 1). 5.0MM LOCKING SCREW 38MM (PART: 04.005.528, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A 10MM/125 DEGREE TITANIUM (TI) CANNULATED TFNA NAIL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Impaired Healing; No Code Available
MAUDE report 8252555 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.
- Z-1239-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
- Z-1240-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
- Z-1246-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
- Z-1248-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
- Z-1247-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
- Z-0126-2016 · Sep 21, 2015 · TerminatedSynthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
- Z-2091-2015 · Jun 10, 2015 · TerminatedHelical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
- Z-1600-2015 · Apr 9, 2015 · TerminatedTFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
- Z-1352-2015 · Feb 24, 2015 · TerminatedSynthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page