Skip to content

TFN-ADVANCE 04.037.033SRecalls & reported problems

Synthes GmbH · 10MM/125 DEG TI CANN TFNA 420MM/LEFT - STERILE · Class 2 · Implant

All models of TFN-ADVANCE (195)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982095916), received between May 10, 2019 and Apr 1, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2024No Apparent Adverse Event; No Apparent Adverse Event

    DEVICE REPORT FROM FRANCE REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGERY TO IMPLANT THE TFNA WITH BLADE ON (B)(6) 2023. THE SUBTROCHANTERIC FRACTURE WAS REDUCED AND OSTEOSYNTHESISED WITH A LONG TFNA NAIL. THE PATIENT WAS SEEN AGAIN AS PART OF HIS FOLLOW-UP CONSULTATION AND HAD RESUMED WEIGHT-BEARING WITHOUT TOO MUCH DIFFICULTY FOR SEVERAL WEEKS. IT WAS NOTED ON IMAGING ON (B)(6) 2024 THAT THE BLADE HAS MIGRATED INTO THE ACETABULUM DESPITE LOCKING THE SCREW DURING THE INITIAL SURGERY. THE PATIENT UNDERWENT A REVISION AND WAS CONVERTED TO AN HSP WITH FEMORAL CERCLAGE (NOT SOLID AT TWO MONTHS). THE DEVICE WAS EXPLANTED ON (B)(6) 2024. THIS REPORT IS FOR A 10MM/125 DEG TI CANN TFNA 420MM/LEFT - STERILE. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19012674 ↗
  • InjuryMay 10, 2019Break; Break

    IT WAS REPORTED ON (B)(6) 2019, PATIENT UNDERWENT DEVICE REMOVAL DUE TO A BROKEN TROCHANTERIC FIXATION NAIL ADVANCE (TFNA) NAIL. THERE WERE NO FRAGMENTS GENERATED FROM THE BROKEN DEVICE. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICES: LOCKING SCREW (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 2), TFNA FENESTRATED SCREW 95MM (PART: 04.038.195, LOT: H691521, QUANTITY: 1). THIS REPORT IS FOR A 10MM/125 DEGREE TITANIUM (TI) CANNULATED TFNA NAIL. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Code Available

    MAUDE report 8601310 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.

  • Z-1240-2024 · Jan 24, 2024 · Open, Classified
    TFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
  • Z-1239-2024 · Jan 24, 2024 · Open, Classified
    TFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
  • Z-1248-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
  • Z-1247-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
  • Z-1246-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
  • Z-0126-2016 · Sep 21, 2015 · Terminated
    Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
  • Z-2091-2015 · Jun 10, 2015 · Terminated
    Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
  • Z-1600-2015 · Apr 9, 2015 · Terminated
    TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
  • Z-1352-2015 · Feb 24, 2015 · Terminated
    Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page