TFN-ADVANCE 04.037.055SRecalls & reported problems
Synthes GmbH · 10MM/130 DEG TI CANN TFNA 340MM/LEFT - STERILE · Class 2 · Implant
All models of TFN-ADVANCE (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982096111), received between Feb 6, 2019 and Oct 15, 2021 — classified as 4 injuries.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility2
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
- Pain1
- Unspecified Tissue Injury1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 15, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN PROCEDURE, WHILE INSERTING A TFNA NAIL, THE SURGEON PERFORATED THE FEMORAL NECK WITH THE HELICAL BLADE. THE NAIL AND HELICAL BLADE WERE EXTRACTED, AND REINSERTED. THE HELICAL BLADE WAS REPOSITIONED IN THE FEMORAL NECK AND THE SURGERY WAS COMPLETED SUCCESSFULLY. DURING EXTRACTION, ONE OF THE PRONGS OF THE CONNECTION INTERFACE WAS SLIGHTLY BENT; HOWEVER THE NAIL WAS SUCCESSFULLY RECONNECTED AND INSERTED. THERE WAS A (15) MINUTE SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCES. THIS REPORT IS FOR (1) 10/130 DEG TI CANN TFNA 340/LEFT - SILE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 12639940 ↗InjuryMay 5, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2021, PATIENT UNDERWENT REMOVAL OF TITANIUM CANNULATED PROXIMAL FEMORAL NAILING SYSTEM (TFNA) NAIL, TFNA FENESTRATED HELICAL BLADE AND TITANIUM LOCKING SCREW DUE TO PAIN. THE SURGEON IMPLANTED A TOTAL HIP REPLACEMENT. SURGERY WAS SUCCESSFUL AND IMPLANTS WERE REMOVED WITH NO ISSUES. NO PATIENT CONSEQUENCES WERE REPORTED. THIS REPORT IS FOR ONE (1) 10MM/130 DEG TI CANN TFNA 340MM/LEFT - STERILE. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: Pain
MAUDE report 11776725 ↗InjuryJan 15, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT A REVISION OF A TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) NAIL DUE TO BREAKAGE AT THE PROXIMAL END WHERE THE HELICAL BLADE IS PLACED. THE NON-UNION OF THE FRACTURE WAS THE REASON THE HARDWARE IS BELIEVED TO HAVE FAILED. THE DATE OF THE INITIAL IMPLANT WAS SIX (6) MONTHS PRIOR. THE PROCEDURE OUTCOME AND PATIENT STATUS WERE UNKNOWN. CONCOMITANT DEVICES: UNKNOWN TFNA HELICAL BLADE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1), UNKNOWN LOCKING SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN), UNKNOWN TFNA END CAP (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE 10 MM/130 DEG TI CANN TFNA 340MM/LEFT - STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 9591055 ↗InjuryFeb 6, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2019, DURING INSERTION OF TROCHANTERIC FEMORAL NAILING ADVANCED (TFNA) NAIL, THE SURGEON TRIED TO COMPRESS THE HELICAL BLADE BUT THERE WAS A BACK OUT. THE SURGEON DECIDED NOT TO COMPRESS THE BLADE, BUT RATHER LOCKED THE BLADE INSTEAD. THIS ALLOWED THE BLADE TO SLIDE AS NORMAL (GUIDED COLLAPSE). IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 10MM/130 DEG TI CANN TFNA 340MM/LEFT - STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8310134 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.
- Z-1240-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
- Z-1239-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
- Z-1247-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
- Z-1246-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
- Z-1248-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
- Z-0126-2016 · Sep 21, 2015 · TerminatedSynthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
- Z-2091-2015 · Jun 10, 2015 · TerminatedHelical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
- Z-1600-2015 · Apr 9, 2015 · TerminatedTFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
- Z-1352-2015 · Feb 24, 2015 · TerminatedSynthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page