TFN-ADVANCE 04.037.056SRecalls & reported problems
Synthes GmbH · 10MM/130 DEG TI CANN TFNA 360MM/RIGHT - STERILE · Class 2 · Implant
All models of TFN-ADVANCE (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982096128), received between Dec 5, 2017 and Aug 19, 2026 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Adverse Event Without Identified Device or Use Problem1
- Failure to Align1
- Mechanical Jam1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Insufficient Information1
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Unspecified Tissue Injury1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 19, 2026Break
Patient problems: Insufficient Information
MAUDE report 50320872 ↗MalfunctionJun 26, 2025No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE TFNA NAIL WAS REVISED DUE TO SCREW BACK OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22336078 ↗MalfunctionMar 17, 2023Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING A PROCEDURE, IT APPEARED THAT THE FENESTRATED LAG SCREW MISSED THE TFNA NAIL WHEN USING THE 130 DEGREE AIMING ARM. THE LAG SCREW ENDED UP MISSING THE NAIL ANTERIOR CAUSING THE SURGEON TO FREE HAND THE LAG SCREW IN THE CORRECT PLACE. ALSO, THE TFNA HELICAL BLADE AND SCREW INSERTER BROKE WHILE TURNING THE SCREW. THE THREADED TIP SNAPPED OFF WHEN INSERTING. THIS COMPLAINT INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) 10/130 DEG TI CANN TFNA 360/RIGHT - SILE THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16565408 ↗InjuryJul 11, 2022Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, A LONG TFNA NAIL WITH LAG SCREW BROKE AT THE JUNCTION OF THE LAG SCREW AND THE NAIL. THE HARDWARE WAS REMOVED AND THE PATIENT WAS REVISED TO A TOTAL HIP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY, PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS REPORT IS FOR A 10MM/130 DEG TI CANN TFNA 360MM/RIGHT - STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 14976943 ↗InjuryMar 16, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, THE SURGEON REMOVED A BROKEN TFNA FROM A PATIENT. THE ORIGINAL IMPLANTATION DATE IS UNKNOWN. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME WAS SUCCESSFUL. CONCOMITANT DEVICES REPORTED: UNKNOWN NAIL HEAD ELEMENTS: TFNA HELICAL BLADE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# 1); UNKNOWN SCREWS: LOCKING (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# UNKNOWN). THIS REPORT IS FOR ONE (1) 10MM/130 DEG TI CANN TFNA 360MM/RIGHT - STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 11495085 ↗MalfunctionApr 18, 2018Failure to Align; Mechanical Jam; Failure to Align; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2018, PATIENT UNDERWENT SURGERY. DURING THE SURGERY, WHILE USING THE HIP TFNA NAIL, THE SURGEON WENT TO TAKE THE LOCKING MECHANISM OF THE NAIL DOWN AND IT WOULDN¿T BUDGE. THEY WERE UNABLE TO SEE ON THE X-RAYS WHETHER THE NAIL WAS UP OR DOWN, SO THEY HAD TO TAKE THE BLADE AND NAIL OUT AND PUT THEM BACK IN AGAIN. THIS TIME, THE LOCKING MECHANISM MOVED WITHOUT A PROBLEM. DUE TO THIS, TWENTY-FIVE (25) MINUTES WERE ADDED TO THE PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO HARM TO THE PATIENT. CONCOMITANT DEVICES REPORTED: HELICAL BLADE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 10MM/130 DEG TI CANN TFNA 360MM/RIGHT ¿ STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7440255 ↗InjuryDec 5, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6)2017 PATIENT PRESENTED TO THE EMERGENCY ROOM WITH A INTERTROCHANTERIC HIP FRACTURE. THE FRACTURE WAS CONFIRMED WITH X-RAYS. THE SURGEON TOOK THE PATIENT TO THE OPERATING ROOM ON (B)(6) 2017 AND PATIENT WAS IMPLANTED WITH A LONG TROCHANTERIC FIXATION NAIL ADVANCE (TFNA) NAIL WITH HELICAL BLADE AND DISTAL LOCKING SCREWS. ON (B)(6) 2017, AN X-RAY DURING NORMAL FOLLOW UP SHOWED THE FEMORAL HEAD STARTING TO SLIGHTLY COLLAPSE. THE SURGEON ELECTED TO WAIT AND WATCH THE FRACTURE. X-RAY ON (B)(6) 2017 SHOWED FURTHER COLLAPSE OF THE HEAD INTO VARUS. ON (B)(6) 2017, SURGEON TOOK THE PATIENT BACK TO THE OPERATING ROOM TO REMOVE THE TFNA NAIL, THE TFNA HELICAL BLADE, AND TWO LOCKING SCREWS AND TOOK CULTURES. NONE OF THE IMPLANTS WERE BROKEN AND THE REMOVAL WAS A SUCCESS. A HEMI-ARTHROPLASTY WAS PERFORMED IN PLACE OF THE HARDWARE REMOVAL. CONCOMITANT PARTS: 5.0MM TI LOCKING SCREW WITH T25 STRARDRIVE 40MM F/IM NAIL (PART NUMBER: 04.005.530S, LOT NUMBER: H359595, QUANTITY 1), 5.0MM TI LOCKING SCREW WITH T25 STARDRIVE 38MM F/IM NAIL (PART NUMBER: 04.005.528S, LOT NUMBER: H329932, QUANTITY 1). THIS REPORT IS FOR ONE (1) 10MM/130 DEGREE TI CANNULATED TFNA 360MM/RIGHT - S
Patient problems: No Code Available
MAUDE report 7084307 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.
- Z-1240-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
- Z-1239-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
- Z-1247-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
- Z-1246-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
- Z-1248-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
- Z-0126-2016 · Sep 21, 2015 · TerminatedSynthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
- Z-2091-2015 · Jun 10, 2015 · TerminatedHelical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
- Z-1600-2015 · Apr 9, 2015 · TerminatedTFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
- Z-1352-2015 · Feb 24, 2015 · TerminatedSynthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page