Skip to content

TFN-ADVANCE 04.037.233SRecalls & reported problems

Synthes GmbH · 12MM/125 DEG TI CANN TFNA 420MM/LEFT - STERILE · Class 2 · Implant

All models of TFN-ADVANCE (195)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982097194), received between Apr 2, 2019 and Apr 2, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, A LONG TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) NAIL BROKE AT THE HELICAL BLADE/NAIL JUNCTION SEVERAL MONTHS AFTER BEING IMPLANTED. ORIGINAL DATE OF IMPLANTATION WAS ON (B)(6) 2018. IMPLANTATION WAS DONE TO TREAT AN INTERTROCHANTERIC FRACTURE WITH SUBTROCHANTERIC EXTENSION. DURING IMPLANTATION, THE INTERNAL LOCKING MECHANISM WAS LOCKED TO CREATE A FIXED ANGLE CONSTRUCT BETWEEN THE NAIL AND THE HELICAL BLADE AND A TORQUE LIMITING ATTACHMENT (TLA) DRIVER WAS USED TO LOCK THE HELICAL BLADE. THERE WAS NO ADVERSE EVENT OR SURGICAL DELAY DURING IMPLANTATION. SIX (6) WEEKS AFTER IMPLANTATION, THE LOCKING MECHANISM FAILED TO RESIST GUIDED COLLAPSE. THREE (3) MONTHS AFTER IMPLANTATION, THE DISTAL LOCKING SCREW FAILED. FIVE (5) MONTHS AFTER IMPLANTATION, THERE WAS OBVIOUS NAIL BREAKAGE. IT IS UNKNOWN IF THERE WAS A REVISION/REMOVAL PROCEDURE. PATIENT STATUS IS UNKNOWN. CONCOMITANT DEVICES REPORTED: STERILE TFNA HELICAL BLADE (PART # 04.038.305S, LOT # UNKNOWN, QUANTITY 1); UNKNOWN LOCKING SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN 1); UNKNOWN DRIVER (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 12MM/125 DEG TI

    Patient problems: No Code Available

    MAUDE report 8473805 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.

  • Z-1240-2024 · Jan 24, 2024 · Open, Classified
    TFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
  • Z-1239-2024 · Jan 24, 2024 · Open, Classified
    TFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
  • Z-1247-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
  • Z-1246-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
  • Z-1248-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
  • Z-0126-2016 · Sep 21, 2015 · Terminated
    Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
  • Z-2091-2015 · Jun 10, 2015 · Terminated
    Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
  • Z-1600-2015 · Apr 9, 2015 · Terminated
    TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
  • Z-1352-2015 · Feb 24, 2015 · Terminated
    Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page