Skip to content

TFN-ADVANCE 04.037.243SRecalls & reported problems

Synthes GmbH · 12MM/130 DEG TI CANN TFNA 200MM - STERILE · Class 2 · Implant

All models of TFN-ADVANCE (195)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982097279), received between Sep 25, 2018 and Jan 5, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • Failure of Implant1
  • Insufficient Information1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 5, 2026Break

    IT WAS REPORTED THAT THE PATIENT HAD A TFNA IMPLANTED IN (B)(6) 2024. A TFNA NAIL BROKE NEAR THE TFNA BLADE. THIS WAS NOTICED AFTER A REPEAT X-RAY TAKEN IN ORDER TO DETERMINE THE TIMING FOR METAL REMOVAL.

    Patient problems: Insufficient Information

    MAUDE report 23967892 ↗
  • InjurySep 25, 2018Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN INTRAMEDULLARY NAIL OF THE LEFT HIP ON (B)(6) 2016, AFTER PATIENT EXPERIENCED A FALL AND BROKE HER LEFT HIP ON AN UNKNOWN DATE. THE PATIENT HAD A CHRONIC NON-UNION DUE TO A MALIGNANT LESION. THE INTERNAL NAIL BROKE THROUGH THE DISTAL LOCKING SCREW ON AN UNKNOWN DATE. THE PATIENT UNDERWENT HARDWARE REMOVAL ON (B)(6) 2018. THE SYNTHES TROCHANTERIC FIXATION NAIL ADVANCE (TFNA) WAS REMOVED COMPLETELY WITH NO REMAINING HARDWARE LEFT BEHIND. ALL REMOVED HARDWARE WAS INTACT EXCEPT FOR THE (TFNA) NAIL. THERE WAS NO SURGICAL DELAY REPORTED. THE PATIENT WAS STABLE. CONCOMITANT DEVICE REPORTED: HELICAL BLADE (PART #04.038.320S, LOT # UNKNOWN, QUANTITY # 1) LOCKING SCREW (PART # 458.944, LOT # UNKNOWN, QUANTITY # 1). THIS REPORT IS FOR ONE (1) TFNA NAIL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 7907286 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.

  • Z-1240-2024 · Jan 24, 2024 · Open, Classified
    TFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
  • Z-1239-2024 · Jan 24, 2024 · Open, Classified
    TFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
  • Z-1246-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
  • Z-1247-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
  • Z-1248-2023 · Feb 1, 2023 · Open, Classified
    TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
  • Z-0126-2016 · Sep 21, 2015 · Terminated
    Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
  • Z-2091-2015 · Jun 10, 2015 · Terminated
    Helical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
  • Z-1600-2015 · Apr 9, 2015 · Terminated
    TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
  • Z-1352-2015 · Feb 24, 2015 · Terminated
    Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page