TFN-ADVANCE 04.037.261SRecalls & reported problems
Synthes GmbH · 12MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE · Class 2 · Implant
All models of TFN-ADVANCE (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886982097415), received between Aug 30, 2018 and Nov 20, 2025 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (7 reports).
9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Crack1
- Difficult to Advance1
Patient problems
- Insufficient Information3
- Failure of Implant2
- No Code Available2
- Nonunion/Delayed-union Bone Fracture2
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryNov 20, 2025Break; Break
IT WAS REPORTED ON AN UNKNOWN DATE THAT THE SAME PATIENT HAS EXPERIENCED 3 TFNA NAILS BREAKING IN THE SAME AREA OF THE NAIL, RESULTING IN A FINAL NAIL REMOVAL AND PROXIMAL FEMUR REPLACEMENT. THE NAIL WAS EXAMINED UPON REMOVAL AND DIDN¿T SEEM TO HAVE ANY SIGNS OF DRILL DAMAGE FROM INITIAL NAIL PLACEMENT.
Patient problems: Insufficient Information
MAUDE report 23603750 ↗InjuryNov 19, 2025Break; Break
IT WAS REPORTED ON AN UNKNOWN DATE THAT THE SAME PATIENT HAS EXPERIENCED 3 TFNA NAILS BREAKING IN THE SAME AREA OF THE NAIL, RESULTING IN A FINAL NAIL REMOVAL AND PROXIMAL FEMUR REPLACEMENT. THE NAIL WAS EXAMINED UPON REMOVAL AND DIDN¿T SEEM TO HAVE ANY SIGNS OF DRILL DAMAGE FROM INITIAL NAIL PLACEMENT.
Patient problems: Insufficient Information
MAUDE report 23592605 ↗InjuryNov 13, 2025Break; Break
IT WAS REPORTED THAT THE SAME PATIENT HAS EXPERIENCED 3 TFNA NAILS BREAKING IN THE SAME AREA OF THE NAIL, RESULTING IN A FINAL NAIL REMOVAL AND PROXIMAL FEMUR REPLACEMENT. THE NAIL WAS EXAMINED UPON REMOVAL AND DIDN¿T SEEM TO HAVE ANY SIGNS OF DRILL DAMAGE FROM INITIAL NAIL PLACEMENT.
Patient problems: Insufficient Information
MAUDE report 23541553 ↗InjuryAug 20, 2025Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED FEMUR NON-UNION WITH BROKEN TFNA AND SCREWS. IMPLANTS WERE REMOVED AND REPLACED WITH NEW TFNA.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 22848809 ↗MalfunctionDec 9, 2021Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT ON (B)(6) 2021, DURING A PROCEDURE, WHEN SURGEON WAS ADVANCING THE SET SCREW IN THE PROXIMAL PORTION OF THE NAIL, IT WOULD NOT ADVANCE DISTAL ENOUGH WITHIN THE NAIL TO ENGAGE WITH THE TFNA LAG SCREW. IT SEEMED LIKE IT WAS NOT CENTERED PROPERLY WITHIN THE NAIL OR WAS CROSS THREADED. AFTER TURNING THE SET SCREW COUNTER CLOCKWISE AND HAMMERING ON THE 5.0MM FLEXIBLE SCREWDRIVER SEVERAL TIMES THE SET SCREW ADVANCED DISTAL ENOUGH TO ENGAGE WITH THE TFNA LAG SCREW IN THE STATIC POSITION. THIS DELAYED THE CASE FOR ABOUT FIFTEEN (15) MINUTES. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) 12MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12969631 ↗InjurySep 2, 2021Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN POSTOPERATIVE DATE THE TROCHANTERIC FIXATION NAIL-ADVANCED (TFNA) NAIL BROKE. THE PATIENT HAD A NON UNION OF THE LEFT. THE PATIENT UNDERWENT THE REVISION SURGERY ON (B)(6) 2021 TO REMOVE THE BROKEN NAIL AND REPLACE IT WITH A LARGER NAIL. PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) 12MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 12418073 ↗InjuryJan 4, 2021Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 THAT THE PATIENT UNDERWENT A REVISION OPERATION DUE TO A BROKEN LONG 400MM TROCHANTERIC FIXATION NAIL ADVANCED (TFNA). THE BREAK OCCURRED APPROXIMATELY SIX (6) MONTHS AFTER IMPLANTATION NEAR THE PROXIMAL BOD AT THE HELICAL BLADE OPENING. THE IMPLANT WAS REPLACED WITH AN ADDITIONAL TFNA IMPLANT WITH A SCREW INSTEAD OF A HELICAL BLADE. THE SURGERY WAS REVISED TO REMOVE AND REPLACE THE FAILED IMPLANT. CONCOMITANT DEVICES: UNK - END CAPS: TFNA (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1); UNK - SCREWS: LOCKING (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1); UNK - NAIL HEAD ELEM: TFNA HELICAL BLADE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 12MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 11108620 ↗InjuryJul 31, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2017, THE PATIENT UNDERWENT A REVISION OSTEOTOMY SURGERY DURING WHICH THE FAILED AND FRACTURED HARDWARE WAS REMOVED AND REPLACED WITH NEW HARDWARE. ON (B)(6) 2016, THE PATIENT UNDERWENT AN INTRAMEDULLARY NAILING SURGICAL PROCEDURE AFTER SUSTAINING A LEFT SUBTROCHANTERIC FEMUR FRACTURE. ON OR ABOUT (B)(6) 2017, THE PATIENT INFORMED THAT THE HARDWARE HAD FAILED AND THAT ITS CANNULATED TITANIUM NAIL HAD FRACTURED. CONCOMITANT DEVICES REPORTED: TFNA 105MM HELICAL BLADE (PART# 04.038.305S, LOT# 7971267, QUANTITY (B)(4)), 5X48MM INTERLOCKING SCREW (PART# UNKNOWN, LOT# 651785, QUANTITY (B)(4)).
Patient problems: Failure of Implant; No Code Available
MAUDE report 10353001 ↗InjuryAug 30, 2018Crack; Crack
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT REVISION SURGERY DUE TO BROKEN TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) NAIL. ORIGINAL IMPLANT DATE WAS (B)(6) 2018 AND PATIENT OUTCOME POST-SURGERY WAS UNEVENTFUL. THE PATIENT REPORTED INCREASING PAIN AND A CRACK WAS IDENTIFIED IN THE TFNA NAIL AT THE PROXIMAL SCREW HOLE ON AN UNKNOWN DATE. THERE WERE INDICATIONS THAT THERE WAS FURTHER TRAUMA THAT MIGHT HAVE LED TO THE NAIL FRACTURE. THE NAIL WAS REMOVED, AND THE FRACTURE WAS FIXED WITH BLADE PLATE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY AND IF THE SURGERY WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICES: LOCKING SCREWS (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: UNKNOWN), TFNA HELICAL BLADE (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A 12MM/130 DEGREE TITANIUM (TI) CANNULATED TFNA 400MM/LEFT-STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available; Pain
MAUDE report 7832417 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.
- Z-1239-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
- Z-1240-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
- Z-1247-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
- Z-1248-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
- Z-1246-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
- Z-0126-2016 · Sep 21, 2015 · TerminatedSynthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
- Z-2091-2015 · Jun 10, 2015 · TerminatedHelical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
- Z-1600-2015 · Apr 9, 2015 · TerminatedTFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
- Z-1352-2015 · Feb 24, 2015 · TerminatedSynthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page