TFN-ADVANCE 04.037.461SRecalls & reported problems
Synthes GmbH · 14MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE · Class 2 · Implant
All models of TFN-ADVANCE (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982097651), received between Aug 16, 2019 and Feb 10, 2025 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
9 other recalls cover devices cleared under the same FDA submission (K131548, K160167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Material Discolored1
- Material Twisted/Bent1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- Insufficient Information3
- No Code Available2
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
- Nonunion/Delayed-union Bone Fracture1
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2025Break; Break
IT WAS REPORTED THAT POSTOPERATIVELY, IT WAS OBSERVED THAT A TFNA NAIL AND DISTAL LOCKING SCREW HAD BROKEN.
Patient problems: Insufficient Information
MAUDE report 21349421 ↗MalfunctionMar 27, 2024Material Discolored; Tear, Rip or Hole in Device Packaging; Detachment of Device or Device Component; Material Twisted/Bent; Material Discolored; Tear, Rip or Hole in Device Packaging; Detachment of Device or Device Component; Material Twisted/Bent
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE IMPLANTS IN QUESTION WERE DAMAGED. THE BOXES THAT CONTAINED THE ITEMS HAD GOTTEN WET, CAUSED THE BOXES TO DISCOLOR AND FALL APART. THE INSTRUMENTS WERE STRIPPED. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 14MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18985409 ↗InjuryAug 1, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, PATIENT CAME TO ER STATING THAT SHE HAD A GROUND LEVEL FALL AND WAS COMPLAINING OF INCREASE LEFT HIP PAIN. PATIENT HAD ORIGINAL TROCHANTERIC FIXATION NAIL-ADVANCED (TFNA) IMPLANTED ON (B)(6) 2022. SURGEON SCHEDULED A REVISION SURGERY ON (B)(6) 2022 FOR REMOVAL OF BROKEN HARDWARE WHICH WAS CONFIRMED BY ER X-RAYS ON (B)(6) 2022. HARDWARE WAS REMOVED AND A TROCHANTERIC FIXATION NAI (TFN) WAS CHOSEN FOR THE REPAIR BECAUSE THE TFN HAD A LARGER DIAMETER PROXIMAL BODY AND LAG SCREW. IMPLANTS WERE REMOVED AND DOCUMENTED AND TFN WAS IMPLANTED WITH SUCCESS. IT WAS DESCRIBED THAT IMPLANT FAILED AT PROXIMAL END OF NAIL, PROXIMAL TO LAG SCREW INSERTION HOLE. PATIENT EXPERIENCED AN ADVERSE EVENT. THE SURGERY WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) 14MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: Pain
MAUDE report 15140420 ↗InjuryNov 3, 2021Break; Break
IT WAS REPORTED THAT, THE PATIENT UNDERWENT A PROCEDURE 14/130 DEG TI CANN TFNA 400/LEFT-STERILE BROKE INTO TFNA NAIL. PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: UNKNOWN HELICAL BLADE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1); UNKNOWN LOCKING SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) 14/130 DEG TI CANN TFNA 400/LEFT-STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 12746157 ↗InjuryDec 31, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, TROCHANTERIC FIXATION NAIL ADVANCED (TFNA) LEFT NAIL 14 X 400MM, 110MM FENESTRATED SCREW, AND TWO (2) DISTAL INTER-LOCKING SCREWS WERE REMOVED DUE TO BONE NON-UNION. PATIENT HAS CANCER. DEVICES WERE ORIGINALLY IMPLANTED ON (B)(6) 2019. IMPLANTS WERE REMOVED SUCCESSFULLY WITHOUT COMPLICATIONS. THIS REPORT IS FOR ONE (1) 14MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 11100556 ↗InjurySep 10, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT A REMOVAL OF A BROKEN TFNA HIP NAIL THAT BROKE INSIDE THE PATIENT. THE PROCEDURE OUTCOME AND PATIENT STATUS IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN HELICAL BLADE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1), UNKNOWN LOCKING SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) 14MM/130 DEG TI CANN TFNA 400MM/LEFT-STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant; No Code Available
MAUDE report 10516466 ↗InjuryAug 16, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, A PATIENT UNDERWENT REMOVAL OF A BROKEN PROXIMAL FEMORAL NAILING SYSTEM (TFNA) NAIL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT STATUS IS UNKNOWN. CONCOMITANT DEVICES: TFNA FENESTRATED HELICAL BLADE (PART: 04.038.410 LOT: H54340, QUANTITY: 1), TITANIUM (TI) LOCKING SCREW (PART: 04.005.536, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A 14MM/130 DEGREE TI CANNULATED TFNA NAIL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 8901052 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131548, K160167. One submission often covers many models, so check the product and lot details in each record.
- Z-1239-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
- Z-1240-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
- Z-1248-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
- Z-1247-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
- Z-1246-2023 · Feb 1, 2023 · Open, ClassifiedTFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
- Z-0126-2016 · Sep 21, 2015 · TerminatedSynthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head eleme
- Z-2091-2015 · Jun 10, 2015 · TerminatedHelical Blade Inserter for Trochanteric Fixation Nail-ADVANCED (TFNA); intended to secure the head element to the head element inserter to prevent disengagement during head element
- Z-1600-2015 · Apr 9, 2015 · TerminatedTFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
- Z-1352-2015 · Feb 24, 2015 · TerminatedSynthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proxima
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page