TFN-ADVANCE 04.038.375SRecalls & reported problems
Synthes GmbH · TFNA FENESTRATED HELICAL BLADE 75MM - STERILE · Class 2 · Implant
All models of TFN-ADVANCE (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982099198), received between Oct 15, 2021 and Aug 5, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K160167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility2
- Device Slipped1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Tissue Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2025Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility
IT WAS REPORTED THAT ON JUL 23, 2025, THE PATIENT UNDERWENT AN UNKNOWN SURGERY FOR TROCHANTERIC FRACTURE WITH THE NAIL AND THE BLADE. AFTER THE SURGEON INSERTED THE BLADE AND INJECTED CEMENT, HE WAS ABOUT TO INSERT THE LATERAL LOCKING SCREW WHEN HE NOTICED THAT THE AIMING ARM AND INSERTION HANDLE WERE LOOSE. JUST TO BE SURE, THE SURGEON CHECKED THE BLADE AND FOUND THAT IT HAD NOT PASSED THROUGH THE NAIL. THE SURGEON REMOVED THEM, CHANGED THE NAIL FROM F10MM TO F9MM, AND THE BLADE FROM 75MM TO 80MM, AND INSERTED AGAIN. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THE SURGEON GAVE THE FOLLOWING COMMENTS ABOUT THE CAUSE: 1) THE DEVICE WAS LOOSE, AND 2) WHEN INSERTING THE GUIDE WIRE TO POSITION THE BLADE, THE SURGEON LIFTED THE REDUCTION TOOL WITH FORCE, WHICH PUT STRESS ON THE DEVICE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22719538 ↗InjuryOct 15, 2021Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN PROCEDURE, WHILE INSERTING A TFNA NAIL, THE SURGEON PERFORATED THE FEMORAL NECK WITH THE HELICAL BLADE. THE NAIL AND HELICAL BLADE WERE EXTRACTED, AND REINSERTED. THE HELICAL BLADE WAS REPOSITIONED IN THE FEMORAL NECK AND THE SURGERY WAS COMPLETED SUCCESSFULLY. DURING EXTRACTION, ONE OF THE PRONGS OF THE CONNECTION INTERFACE WAS SLIGHTLY BENT; HOWEVER THE NAIL WAS SUCCESSFULLY RECONNECTED AND INSERTED. THERE WAS A FIFTEEN (15) MINUTE SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) TFNA FENESTRATED HELICAL BLADE 75 - SILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 12639890 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160167. One submission often covers many models, so check the product and lot details in each record.
- Z-1239-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 9mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
- Z-1240-2024 · Jan 24, 2024 · Open, ClassifiedTFNA Femoral Nail 11mm, right, 130, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.