NA 05.001.045Recalls & reported problems
Synthes GmbH · BURR ATTACHMENT-SHORT FOR PEN DRIVE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982128669), received between May 7, 2018 and Jan 31, 2023 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Excessive Heating” (4 reports).
6 other recalls cover devices cleared under the same FDA submission (K043310); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Excessive Heating4
- Overheating of Device2
- Appropriate Device Problem Term/Code Not Available1
- Device-Device Incompatibility1
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Burn(s)1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 31, 2023Excessive Heating; Excessive Heating
IT WAS REPORTED BY NETHERLANDS THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE BURR ATTACHMENT DEVICE WAS GENERATING HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR GENERAL FUNCTION IN RUNNING MODE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16271367 ↗MalfunctionMay 9, 2022Appropriate Device Problem Term/Code Not Available; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Excessive Heating
IT WAS REPORTED BY (B)(4) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE BURR ATTACHMENT DEVICE WAS GENERATING HEAT. IT WAS FURTHER OBSERVED THAT THE DEVICE HAD VIBRATION AND ILLEGIBLE ETCH. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR GENERAL CONDITION AND MARKINGS. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14336270 ↗MalfunctionJul 29, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING PRE-LAB TESTING, IT WAS OBSERVED THAT THE BURR ATTACHMENT-SHORT DEVICE PRODUCED AN EXCESS HEAT AND UNUSUAL NOISE. THE EVENT WAS NOT RELATED TO SURGERY. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12247177 ↗MalfunctionMay 18, 2021Device-Device Incompatibility; Premature Separation; Device-Device Incompatibility; Premature Separation
IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE BURR ATTACHMENT DEVICE 'GOT DISABLE TO CONNECT TO THE HANDPIECE DEVICE'. ACCORDING TO THE REPORTER, THE HANDPIECE CONNECTION GOT POOR JUST AFTER SEVERAL TIMES OF NORMAL USE. THERE WERE NO REPORTS OF DELAYS IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11840758 ↗InjuryOct 22, 2019Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS GETTING HOT WHICH LEAD TO THE PATIENT'S LIP BEING BURNED. THERE WERE NO REPORTS OF DELAYS IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION WERE REQUIRED. THE DATE OF EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Burn(s)
MAUDE report 9220931 ↗MalfunctionMay 7, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED FROM (B)(6) THAT THE BURR ATTACHMENT DEVICE WAS MAKING NOISE, SMOKING AND OVERHEATING. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8587394 ↗MalfunctionMay 7, 2018Overheating of Device; Overheating of Device
IT WAS REPORTED FROM (B)(6) THAT THE BURR ATTACHMENT-SHORT DEVICE WAS MAKING A STRANGE SOUND AND BECAME HOT DURING USE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7491921 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K043310. One submission often covers many models, so check the product and lot details in each record.
- Z-0681-2017 · Nov 1, 2016 · TerminatedSynthes 0.76 mm Drill Bit/Stryker J-Latch/with 5mm Stop/44.5 mm Non-Sterile
- Z-2783-2016 · Jun 24, 2016 · TerminatedLight Adaptor for Small Battery Drive and Small Battery Drive II.
- Z-2784-2016 · Jun 24, 2016 · TerminatedAdaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in
- Z-2199-2016 · May 31, 2016 · TerminatedPart # 313.250, 1.1mm Drill Bit, Mini QC with 4mm Stop/44.5mm, lot # 4326989, mfg. 9/28/2001 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments ar
- Z-1313-2015 · Dec 19, 2014 · TerminatedSynthes Electric Pen Drive (EPD) System, The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticati
- Z-0615-2014 · Apr 15, 2013 · TerminatedSynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and met
Other DRILL, BONE, POWERED devices
- NA 03.505.004Synthes GmbH · DRILL, BONE, POWERED69 reports
- Nobel Biocare® 37470Nobel Biocare AB · DRILL, BONE, POWERED1 recall0 reports
- MatrixMANDIBLE 03.503.476Synthes GmbH · DRILL, BONE, POWERED26 reports
- NA 310.510Synthes GmbH · DRILL, BONE, POWERED17 reports
- NA 317.82Synthes GmbH · DRILL, BONE, POWERED16 reports
- MatrixMANDIBLE 03.503.479Synthes GmbH · DRILL, BONE, POWERED11 reports
- NA 05.001.120Synthes GmbH · DRILL, BONE, POWERED9 reports
- NA 310.520Synthes GmbH · DRILL, BONE, POWERED9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page