Skip to content

NA 05.001.183Recalls & reported problems

Synthes GmbH · SAW ATTACHMENT/IN-LINE FOR EPD AND APD · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982129567), received between Aug 31, 2018 and Feb 20, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Loss of Power2
  • Material Fragmentation2
  • Appropriate Device Problem Term/Code Not Available1
  • Detachment of Device or Device Component1
  • Excessive Heating1
  • Noise, Audible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2025Appropriate Device Problem Term/Code Not Available; Noise, Audible; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Noise, Audible; Excessive Heating

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE SAW ATTACHMENT DEVICE HAD EXCESSIVE NOISE, HEAT AND THE MOVING PARTS DID NOT MOVE SMOOTHLY. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK GENERAL FUNCTION IN THE RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS RATTLING WHEN IN USE. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE FACILITY IS A VETERINARY MEDICAL HOSPITAL, AND THEREFORE, THE DEVICE IS MOST LIKELY USED FOR VETERINARY PURPOSES ONLY. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21426030 ↗
  • MalfunctionMay 24, 2024Material Fragmentation; Complete Loss of Power; Material Fragmentation; Complete Loss of Power

    IT WAS REPORTED THAT DURING AN OSTEOTOMY PROCEDURE IT WAS OBSERVED THAT THE SAW BLADE OSCILLATION STOPPED AS SOON AS PRESSURE WAS APPLIED, AND SMALL AMOUNTS OF DEBRIS WERE NOTED SIMULTANEOUSLY. IT WAS NOT REPORTED IF THERE WAS A DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED AS THIS WAS A VETERINARY PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19397178 ↗
  • MalfunctionMay 24, 2024Material Fragmentation; Complete Loss of Power; Material Fragmentation; Complete Loss of Power

    IT WAS REPORTED THAT DURING AN OSTEOTOMY PROCEDURE IT WAS OBSERVED THAT THE SAW BLADE OSCILLATION STOPPED AS SOON AS PRESSURE WAS APPLIED, AND SMALL AMOUNTS OF DEBRIS WERE NOTED SIMULTANEOUSLY. IT WAS NOT REPORTED IF THERE WAS A DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED AS THIS WAS A VETERINARY PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19397098 ↗
  • MalfunctionAug 31, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE COUPLING TOOL SIDE OF THE SAW ATTACHMENT/IN-LINE DEVICE WAS BROKEN AND TORN OFF. IT WAS FURTHER DETERMINED THAT THE SAW HEAD BASE WAS BROKEN. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR STEEL HOUSING VERSION, TICK ¿PASS¿ FOR STEEL HOUSING AND CHECK THE SAW BLADE COUPLING. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK ANYMORE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2018. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7838546 ↗

Other MOTOR, DRILL, PNEUMATIC devices

All MOTOR, DRILL, PNEUMATIC devices →

Sources and updates