NA 05.001.213Recalls & reported problems
Synthes GmbH · LARGE QUICK COUPLING FOR DRILL BITS FOR TRAUMA RECON SYSTEM · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Synthes large quick coupling is the drill bit attachment for the Trauma Reconstruction System. The 72 reports mostly describe the coupling jamming or freezing so it would not move, coming apart or slipping off the drill, and losing power. All72 are malfunctions, with no deaths and no injuries. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 71
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 10886982129819), received between Dec 14, 2017 and Apr 6, 2026 — classified as 71 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (44 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam44
- Complete Loss of Power19
- Detachment of Device or Device Component19
- Appropriate Device Problem Term/Code Not Available13
- Premature Separation4
- Device-Device Incompatibility3
- Break1
- Compatibility Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions54
- No Consequences Or Impact To Patient16
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2026Mechanical Jam; Complete Loss of Power
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE SMALL BATTERY DRIVE DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE WOULD NOT START. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24794280 ↗MalfunctionApr 1, 2026Mechanical Jam; Complete Loss of Power
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LARGE QUICK COUPLING DEVICE WAS FROZEN/WILL NOT MOVE, MOVING PARTS OF THE DEVICE DID NOT MOVE SMOOTHLY, AND THE COUPLING WAS WORN. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR GENERAL CONDITION AND CHECK THE TOOL COUPLING. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24752713 ↗MalfunctionMar 4, 2026Mechanical Jam; Complete Loss of Power
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LARGE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT FUNCTION. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24506719 ↗MalfunctionJan 21, 2026Appropriate Device Problem Term/Code Not Available; Premature Separation
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LARGE QUICK COUPLING DEVICE HAD AN ISSUE WITH DEVICE INTERACTION (2+ DEVICES): UNINTENDED/UNEXPECTED DISENGAGEMENT. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24131604 ↗MalfunctionNov 4, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE LARGE QUICK COUPLING FOR TRS DEVICE WAS FROZEN/WOULD NOT MOVE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23467236 ↗MalfunctionOct 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT POST SURGERY, IT WAS OBSERVED THAT THE LARGE QUICK COUPLING FOR TRS DEVICE HAD A DROPPED PIN. IT WAS FURTHER REPORTED THAT COMPONENTS OF THE DEVICE HAD SEPARATED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23363603 ↗MalfunctionSep 15, 2025Break; Unintended System Motion; Device-Device Incompatibility; Premature Separation; Break; Unintended System Motion; Device-Device Incompatibility; Premature Separation
IT WAS REPORTED THAT THE QUICK COUPLING DEVICE ATTACHMENT WAS SLIPPING. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION." THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23049472 ↗MalfunctionJul 9, 2025Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT THE QUICK COUPLING DEVICE WAS NOT CLAMPING/HOLDING THE REAMER DEVICES, CAUSING THEM TO CONTINUALLY FALL OUT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22461645 ↗MalfunctionApr 30, 2025Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS NOTED IN THE SERVICE ORDER THAT THE QUICK COUPLING DEVICE DID NOT WORK. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER DETERMINED THAT THE DEVICE HAD COSMETIC DAMAGE AND COULD ENGAGE THE ATTACHMENT - COUPLING. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK PINS LENGTH OF HANDPIECE COUPLING, CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21934075 ↗MalfunctionDec 11, 2024Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20901429 ↗MalfunctionNov 19, 2024Mechanical Jam; Appropriate Device Problem Term/Code Not Available; Mechanical Jam; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE COUPLING WOULD NOT ENGAGE THE ATTACHMENT DEVICE, WAS FROZEN/WOULD NOT MOVE AND HAD COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETESTS FOR GENERAL CONDITION, CHECK THE TOOL COUPLING, CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS NOT WORKING PROPERLY. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20724719 ↗MalfunctionAug 26, 2024Mechanical Jam; Appropriate Device Problem Term/Code Not Available; Mechanical Jam; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED FROM INDIA THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE AND THE COUPLING COULD NOT BE ENGAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK THE TOOL COUPLING, CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN THE RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS NOT WORKING PROPERLY. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, AN ADDITIONAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20066920 ↗MalfunctionApr 1, 2024Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED BY SINGAPORE THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER OBSERVED THAT THE MECHANICS SEIZED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE ATTACHMENT WAS NOT WORKING WHEN CONNECTED TO THE DRILL DEVICE. IT WAS REPORTED THAT THE DRILL WORKED WITH ANOTHER ATTACHMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19013289 ↗MalfunctionDec 18, 2023Appropriate Device Problem Term/Code Not Available; Premature Separation; Appropriate Device Problem Term/Code Not Available; Premature Separation
IT WAS REPORTED FROM JAPAN THAT DURING AN OPEN REDUCTION AND INTERNAL FIXATION (ORIF) SURGERY FOR FEMORAL INTERTROCHANTERIC FRACTURE ON PROXIMAL FEMUR, IT WAS OBSERVED THAT THE QUICK COUPLING DEVICE AND REAMER TIP COULD NOT BE CONNECTED AND WOULD SPIN OUT IF TORQUE WAS APPLIED. THERE WAS A TEN MINUTE DELAY IN THE SURGICAL PROCEDURE. A SPARE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18344782 ↗MalfunctionDec 14, 2023Mechanical Jam; Appropriate Device Problem Term/Code Not Available; Mechanical Jam; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED FROM NETHERLANDS THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE LARGE QUICK COUPLING DEVICE WAS WORN AND WOULD NOT MOVE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETESTS FOR GENERAL CONDITION, CHECK THE TOOL COUPLING, CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNSPECIFIED MALFUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18328998 ↗MalfunctionOct 11, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED BY JAPAN THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER OBSERVED THAT THE ETCHING ON THE LABELING WAS ILLEGIBLE, THE MECHANICS SEIZED, AND THE DEVICE HAD COSMETIC DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR MARKING AND LABELING, CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17914817 ↗MalfunctionOct 2, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED FROM SINGAPORE THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE HAD SEIZED MECHANICS AND WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR MECHANICAL FREE MOVEMENT, CHECK GENERAL FUNCTION IN THE RUNNING MODE AND CHECK FOR RUN-OUT INSPECTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS NOT WORKING AND HAD A CEASED CENTRAL BEARING AND TRIED TO LUBRICATE BUT THE ATTACHMENT STILL FAILED TO WORK. HAD AN UNSPECIFIED MALFUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, AN ADDITIONAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17853871 ↗MalfunctionSep 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED FROM DENMARK THAT THE QUICK COUPLING ATTACHMENT DEVICE KNOB WAS BROKEN OFF. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17787060 ↗MalfunctionJul 19, 2023Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION PROCEDURE FOR A FEMORAL FRACTURE, THE QUICK COUPLING DEVICE WOULD NOT ROTATE AND THE SURGEON REPLACED THE REAMER SEVERAL TIMES BUT IT WOULD NOT MOVE. IT WAS REPORTED THAT THE SURGEON SWITCHED TO THE JACOBS CHUCK DEVICE AND IT WORKED FINE. THEREFORE, THIS WAS THOUGHT THAT THE PROBLEM WAS WITH THE QUICK COUPLING. IT WAS REPORTED THAT THEIR WERE NO DELAYS IN THE SURGICAL PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17349995 ↗MalfunctionJun 8, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS DETERMINED THAT THE MECHANICS SEIZED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17086676 ↗MalfunctionJun 2, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED FROM JAPAN THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE HAD ILLEGIBLE ETCHING, THIN PRINT, WAS FROZEN/WILL NOT MOVE, THE SHAFT WAS STIFF, WAS NOT WORKING, HAD COSMETIC DAMAGE, AND WORN ETCH. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR MARKING AND LABELING, CHECK FOR MECHANICAL FREE MOVEMENT, AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17050268 ↗MalfunctionJun 2, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED BY SWITZERLAND, THAT DURING SERVICE AND EVALUATION. IT WAS DETERMINED, THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER OBSERVED, THAT THE DEVICE HAD A SEIZED GEAR, CORROSION/RUSTING/PITTING. AND THE LABELING HAD AN ILLEGIBLE ETCH. IT WAS FURTHER DETERMINED, THAT THE DEVICE FAILED PRETEST FOR MARKING AND LABELING, CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER, THAT THE DEVICE WAS NOT WORKING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED, THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049698 ↗MalfunctionMay 19, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power
IT WAS REPORTED BY INDIA THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE QUICK COUPLING DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECKING FOR MECHANICAL FREE MOVEMENT AND CHECKING GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16963843 ↗MalfunctionMay 19, 2023Mechanical Jam; Mechanical Jam
IT WAS REPORTED FROM CANADA THAT THE INNER MECHANISM OF LARGE QUICK COUPLING DEVICE DID NOT SPIN THE DRILL AT ALL. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16963832 ↗MalfunctionMay 4, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED FROM ISRAEL THAT THE QUICK COUPLING DEVICE WAS BROKEN. IT WAS REPORTED THAT THE BUTTON FELL OFF DURING CLEANING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16867551 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page