Skip to content

NA 05.001.217Recalls & reported problems

Synthes GmbH · HUDSON DRILLING ATTACHMENT FOR TRAUMA RECON SYSTEM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982129895), received between Nov 28, 2022 and May 1, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Loss of Power4
  • Mechanical Jam4
  • Excessive Heating1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2024Excessive Heating; Excessive Heating

    IT WAS REPORTED FROM THE NETHERLANDS THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD HEAT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, AN ADDITIONAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219668 ↗
  • MalfunctionSep 6, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DRILLING ATTACHMENT DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER OBSERVED THAT THE DEVICE HAD SEIZED MECHANICS. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECKING FOR MECHANICAL FREE MOVEMENT AND CHECKING GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17693412 ↗
  • MalfunctionJan 21, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED BY SINGAPORE THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE HUDSON DRILLING ATTACHMENT DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER OBSERVED THAT THE DEVICE HAD SEIZED MECHANICS AND ATTACHMENT WAS UNABLE TO ROTATE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECKING FOR MECHANICAL FREE MOVEMENT, AND CHECKING GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16220583 ↗
  • MalfunctionDec 2, 2022Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED HUDSON DRILLING ATTACHMENT DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER OBSERVED THAT THE DEVICE HAD SEIZED MECHANICS. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECKING FOR MECHANICAL FREE MOVEMENT AND CHECKING GENERAL FUNCTION IN RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15911099 ↗
  • MalfunctionNov 28, 2022Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED FROM GERMANY THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD A WORN BEARING AND WAS FROZEN/WOULD NOT MOVE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR MECHANICAL FREE MOVEMENT AND CHECK GENERAL FUNCTION IN THE RUNNING MODE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE KNOT BLOCKED THE HANDPIECE AND THE HANDPIECE WOULD NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, AN ADDITIONAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15873406 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates