LCP 223.411Recalls & reported problems
Synthes GmbH · 3.5MM LCP® METAPHYSEAL PLATE 11 HOLES · Class 2 · Implant
All models of LCP (367)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982157638), received between Dec 19, 2017 and Jan 28, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- Nonunion/Delayed-union Bone Fracture1
- Post Operative Wound Infection1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT A HARDWARE REMOVAL WITH 3.5 LCP METAPHYSEAL PLATE AND 3.5 CORTEX SCREWS IN QUESTION. THE PATIENT HAD AN INITIAL PROCEDURE PERFORMED AT AN OUTSIDE HOSPITAL IN (B)(6) OF 2020 FOR A HUMERUS NON-UNION. THE PATIENT WAS TRANSFERRED TO A SURGEON ON (B)(6) 2021 AND UPON LABORATORY DIAGNOSIS WAS DETERMINED TO HAVE AN INFECTION AND THE HARDWARE WAS REMOVED WITHOUT ISSUE ON (B)(6) 2021. THIS REPORT IS FOR ONE (1) 3.5MM LCP® METAPHYSEAL PLATE 11 HOLES. THIS IS REPORT 1 OF 10 FOR (B)(4).
Patient problems: Post Operative Wound Infection
MAUDE report 11244005 ↗InjuryDec 19, 2017Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION OF A PLATE AND SCREWS ON (B)(6) 2017 DUE TO BROKEN PLATE, NON-UNION, AND SEPTIC INFECTION. THE PATIENT PRESENTED WITH A BROKEN 3.5 MM LOCKING COMPRESSION PLATE (LCP) METAPHYSEAL PLATE, 11 HOLE PLATE IN THE LEFT MID-SHAFT ULNA. THE ORIGINAL IMPLANT DATE IS UNKNOWN. THE PLATE AND NINE (9) SCREWS (SIX (6) 3.5 MM CORTEX SCREWS IN THE PLATE, ONE (1) INDEPENDENT LAG SCREW, AND TWO (2) 3.5 MM LOCKING SCREWS) WERE SUCCESSFULLY REMOVED, THERE WERE NO FRAGMENTS. A CHISEL WAS USED TO REMOVE BONE GROWTH FROM OVER THE SCREWS AND PLATE. THE AREA WAS DEBRIDED AND IRRIGATED. A NEW PLATE WAS IMPLANTED; THE SURGEON CUT THE PROXIMAL TIP OF THE PLATE TO ACCOMMODATE THE PATIENT ANATOMY. RELEVANT SCREWS WERE USED FOR FIXATION: VARIABLE ANGLE ELBOW SET AND LCP SMALL FRAGMENT SET. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR A 3.5 MM LCP METAPHYSEAL PLATE. (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 7130842 ↗
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE devices
- LOCK SCREW SQUARE DRIVE 131827118Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE1 recall0 reports
- LOCK SCREW SQUARE DRIVE 131827120Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE1 recall0 reports
- IJS IJS-E Base Plate AssemblySkeletal Dynamics L.L.C. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE28 reports
- NA 390.008Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE21 reports
- NA 294.56Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 390.003Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 394.85Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- TRUE/LOK MONOLATERAL/BILATERAL FIXATOR 1Orthofix Srl · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page