DHS 281.131Recalls & reported problems
Synthes GmbH · 135 DEG DHS® PLATE-STANDARD BARREL 3 HOLES/62MM · Class 2 · Implant
All models of DHS (57)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982176660), received between Mar 19, 2018 and Mar 19, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2018Break; Break
IT WAS REPORTED THAT AN INITIAL SURGERY FOR IMPLANTATION OF A DYNAMIC HIP SYSTEM (DHS) INTO THE PROXIMAL FEMUR WAS PERFORMED ON (B)(6) 2018. DURING THE SURGERY, THE LAG SCREW FOR THE DHS PLATE WAS PUT IN CORRECTLY, HOWEVER, THE PLATE WAS PUT IN OFF-AXIS. THEREFORE, THE PLATE HAD TO BE RE-POSITIONED, AND AS THE SURGEON WAS REMOVING THE CONNECTING SCREW AND GUIDE SHAFT FROM THE LAG SCREW, THE TWO INSTRUMENTS BECAME BENT (IT WAS REPORTED THAT THE SURGEON DID NOT DISCONNECT THE INSTRUMENTS PRIOR TO TAKING THEM OUT) AND THE SURGEON HAD TO USE PLIERS TO TAKE THE INSTRUMENTS OUT. THE SAME PLATE WAS PUT BACK INTO THE PATIENT WITHOUT ISSUE. A GUIDE WIRE WAS ALSO REPORTED TO BE BENT DURING THE SURGERY. THE PATIENT WAS IMPLANTED WITH A LAG SCREW, PLATE AND THREE 4.5 MM CORTICAL SCREWS (SIZE 36, 42 AND 48 MM). THERE WAS NO REPORTED DELAY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. IT WAS FURTHER REPORTED THAT IN STERILE PROCESSING THE CONNECTING SCREW BROKE INTO TWO (2) PIECES AND THE GUIDE SHAFT, STILL IN ONE PIECE, WAS BENT TRYING TO GET THEM APART. THIS COMPLAINT IS LINKED WITH (B)(4) FOR THE POST-OPERATIVE EVENT OF BROKEN/BENT (CONNECTING SCREW AND GUIDE SHAFT) INSTRUMENTS. CONCOMITANT DEV
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7351462 ↗
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page