DHSRecalls & reported problems across all models
Synthes GmbH · 57 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 296
- Deaths reported
- 0
- Injuries reported
- 44
- Recalls
- 0
We found 57 models of DHS by Synthes GmbH in the FDA's device registry (GUDID). Together they are named in 296 adverse event reports in the FDA's MAUDE database, received between Dec 21, 2016 and May 21, 2026 — classified as 44 injuries, 252 malfunctions.
The most frequently reported specific device problem is “Break” (166 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported DHS problems
Device problems
- Break166
- Device-Device Incompatibility44
- Material Deformation32
- Adverse Event Without Identified Device or Use Problem21
- Material Twisted/Bent13
- Failure to Cut11
- Bent9
- Crack6
Patient problems
- No Clinical Signs, Symptoms or Conditions120
- No Consequences Or Impact To Patient60
- No Patient Involvement36
- No Code Available25
- No Known Impact Or Consequence To Patient19
- Insufficient Information16
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
- Nonunion/Delayed-union Bone Fracture5
DHS models
57 models, most reported first. Each model has its own page with its reports, recalls and device details.
- DHS 338.31EXTRACTOR64 reports
- DHS 338.20ORTHOPEDIC MANUAL SURGICAL INSTRUMENT42 reports
- DHS 338.26IMPACTOR35 reports
- DHS 338.23Guide, surgical, instrument26 reports
- DHS 338.19Guide, surgical, instrument11 reports
- DHS 338.28IMPACTOR11 reports
- DHS 338.10BIT, DRILL10 reports
- DHS 338.21Guide, surgical, instrument9 reports
- DHS 338.302WRENCH8 reports
- DHS 280.900APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT7 reports
- DHS 338.08WRENCH5 reports
- DHS 280.850APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT4 reports
- DHS 280.950APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT3 reports
- DHS 280.990APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT3 reports
- DHS 338.06WRENCH3 reports
- DHS 338.11REAMER3 reports
- DHS 280.000APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 280.100APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 280.280APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 280.290APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 281.021APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 281.102APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 281.140APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 281.540SAPPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 reports
- DHS 338.13REAMER2 reports
- DHS 338.22ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2 reports
- DHS 338.750Guide, surgical, instrument2 reports
- DHS 280.050APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 280.100SAPPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 280.285SDEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 280.295DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 280.295SAPPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 280.305DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 280.310DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 280.800APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 280.800SAPPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 280.900SAPPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 281.021SDEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 281.031DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 281.031SDEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 281.060SPLATE, FIXATION, BONE1 report
- DHS 281.100SAPPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 281.130APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 281.131APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 281.160APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 281.220APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 281.440PLATE, FIXATION, BONE1 report
- DHS 281.450PLATE, FIXATION, BONE1 report
- DHS 281.504DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 281.520SDEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT1 report
- DHS 281.840APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 report
- DHS 338.005Guide, surgical, instrument1 report
- DHS 338.01Guide, surgical, instrument1 report
- DHS 338.02Guide, surgical, instrument1 report
- DHS 338.047Guide, surgical, instrument1 report
- DHS 338.32Guide, surgical, instrument1 report
- DHS 338.44REAMER1 report
Recent DHS adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2026Difficult to Advance; Material Deformation
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE LAG SCREW. THE SURGEON DESCRIBED IT AS NOT FEELING QUITE RIGHT WHEN SCREWING IT IN BUT WONDERED IT JUST DUE TO THE GOOD BONE QUALITY, THEY THEN TRIED TO FIT THE DHS PLATE OVER THE SCREW AND IT WOULDN¿T FIT, THEY TOOK THE SCREW OUT AND THE TOP WHERE THEY INSERT THE SCREW DRIVER LOOKED WARPED. THEY REPLACED THE SCREW WITH A NEW ONE AND IT WORKED FINE.
Patient problems: Insufficient Information
MAUDE report 25240560 ↗MalfunctionApr 30, 2026Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE COUPLING SCREW BROKE OFF INSIDE THE BACK OF THE LAG SCREW WHILE INSIDE THE PATIENT. THE BROKEN PIECE WAS RETRIEVED. THERE WAS NO REPORTED PATIENT HARM OR SIGNIFICANT SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25046863 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT DHS COUPLING SCREW THREADED TIP BROKE OFF INTO THE THREADED PORTION OF THE LAG SCREW.
Patient problems: Insufficient Information
MAUDE report 24944409 ↗MalfunctionApr 10, 2026Break
IT WAS REPORTED ON (B)(6), 2026 THAT THE AWL HAS BROKEN TIP, AND THE INSERTION INSTRUMENT WAS BENT. NO OPERATION INVOLVED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24839267 ↗MalfunctionApr 6, 2026Failure to Cut
IT WAS REPORTED THAT THERE WAS A DULL REAMER AND BARREL REAMERS. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24793740 ↗MalfunctionApr 6, 2026Failure to Cut
IT WAS REPORTED THAT THERE WAS A DULL REAMER AND BARREL REAMERS. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24793729 ↗MalfunctionApr 6, 2026Failure to Cut
IT WAS REPORTED THAT THE REAMER AND BARREL REAMERS WERE DULL. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24793589 ↗MalfunctionApr 6, 2026Failure to Cut
IT WAS REPORTED THAT THE REAMER AND BARREL REAMERS WERE DULL. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24793587 ↗MalfunctionMar 24, 2026Break; Crack; Device-Device Incompatibility
IT WAS REPORTED THAT THE INSTRUMENTS WERE UNABLE TO STERILIZE/ASSEMBLE. THERE WAS CHIPPING AND CRACKING. IT WAS UNABLE TO BE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24672974 ↗MalfunctionMar 24, 2026Break; Crack; Device-Device Incompatibility
IT WAS REPORTED THAT THE INSTRUMENTS WERE UNABLE TO STERILIZE/ASSEMBLE. THERE WAS CHIPPING AND CRACKING. IT WAS UNABLE TO BE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24672969 ↗MalfunctionMar 24, 2026Break; Crack; Device-Device Incompatibility
IT WAS REPORTED THAT THE INSTRUMENTS WERE UNABLE TO STERILIZE/ASSEMBLE. THERE WAS CHIPPING AND CRACKING. IT WAS UNABLE TO BE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24672967 ↗MalfunctionMar 10, 2026Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE DHS SCREW INSERTER WAS BENT SO IT DID NOT CAPTURE THE SCREW. THEREFORE IT COULD NOT BE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24558348 ↗MalfunctionMar 10, 2026Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE DHS SCREW INSERTER WAS BENT SO IT DID NOT CAPTURE THE SCREW. THEREFORE IT COULD NOT BE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24558347 ↗MalfunctionFeb 6, 2026Break
IT WAS REPORTED THAT THE COUPLING SCREW TIP BROKE INTRAOPERATIVELY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT DELAY. NO FRAGMENTS WERE GENERATED, AND THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24280772 ↗MalfunctionJan 13, 2026Break; Material Twisted/Bent
IT WAS REPORTED THAT MULTIPLE TRAYS HAD COMPONENTS/INSTRUMENTS THAT WERE STARTING TO BREAK APART AND SHEDDING PIECES WITH NOT IN USABLE CONDITION DURING THE SPD STERILIZATION PROCESS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24059233 ↗
Other Synthes GmbH products
All Synthes GmbH devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page