NA 309.035Recalls & reported problems
Synthes GmbH · HOLLOW REAMER COMPLETE FOR 3.5MM & 4.0MM SCREWS · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886982185549), received between Sep 15, 2017 and Feb 5, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Failure to Cut1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2024Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE IS BROKEN. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) HOLLOW REAMER COMPLETE FOR 3.5MM & 4.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18646472 ↗MalfunctionJan 20, 2023Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: THIS PC IS RELATED TO (B)(4) WHICH REPORTS THAT THE CORTEX SCREW WAS BROKEN AFTER THE PRIMARY SURGERY. THIS PC REPORTS THAT THE REAMER TUBE WAS BROKEN DURING THE REVISION SURGERY. IT WAS REPORTED THAT ON UNKNOWN DATE, THE PATIENT UNDERWENT THE REVISION SURGERY FOR THE PLATEAU PSEUDOARTHROSIS. THE SURGEON USED THE HOLLOW REAMER TO REMOVE THE BROKEN CORTEX SCREW IN BONE. DURING USING THE HOLLOW REAMER, THE REAMER TUBE WAS BROKEN. THE SURGEON CONFIRMED BY THE IMAGE AND DIRECT OBSERVATION AND DETERMINED THAT THERE WAS NO BROKEN BLADE IN THE BODY. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) HOLLOW-REAMER-COMPL ANTICLOCKWISE CUTTIN THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16214032 ↗MalfunctionJul 20, 2022Failure to Cut; Failure to Cut
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT AT AN UNKNOWN DATE IN 2022, THE REAMER WAS FOUND TO BE UNABLE TO CUT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A HOLLOW REAMER COMPLETE FOR 3.5MM & 4.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15065612 ↗MalfunctionMay 21, 2021Break; Break
THIS IS REPORT 1 OF 6 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11867294 ↗InjuryMar 16, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT A SURGERY FOR DISTAL TIBIA WITH THE LOCKING SCREWS. ON AN UNKNOWN DATE, THE SURGEON FOUND THAT ONE SCREW HAD BEEN BROKEN. ON (B)(6) 2020, THE PATIENT UNDERWENT A REMOVAL SURGERY WITH THE HOLLOW REAMER DUE TO THE SCREW BREAKAGE. DURING THE SURGERY, THE SURGEON FOUND THAT THE OTHER 2 SCREWS HAD BEEN BROKEN AND HE HAD DIFFICULTY IN REMOVING THE BROKEN SCREW WITH THE HOLLOW REAMER BECAUSE THE PATIENT¿S BONE QUALITY WAS GOOD. HE TRIED MANY TIMES, AND WHEN HE PRESSURED THE REAMER AXIALLY, THE TIP OF THE REAMER BROKE. HE REMOVED THE FRAGMENT OF THE BROKEN REAMER, BUT THE FRAGMENT OF THE SCREWS REMAINED IN THE PATIENT. HE COMPLETED THE SURGERY WITH ANOTHER HOLLOW REAMER WITHIN 30 MINUTES DELAY. THE TOTAL OF THE BROKEN SCREWS IS 3. THE PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICES REPORTED: UNKNOWN SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1), SPARE CENTERING PIN (PART # 309.370, LOT # UNKNOWN, QUANTITY # 1). THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS IS 1 OF 2 FOR REPORT (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 9837169 ↗MalfunctionMar 3, 2020Break; Break
IT WAS REPORTED ON (B)(6) 2020, THE THREADED PORTION OF A HOLLOW REAMER COMPLETE WAS BROKEN OFF INTO THE SPARE REAMER TUBE FOR HOLLOW REAMER. THE ISSUE WAS FOUND OUT IN THE STERILE PROCESSING. THERE WAS NO PATIENT INVOLVEMENT. THIS IS REPORT 01 OF 01 OF (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9780422 ↗MalfunctionFeb 8, 2019Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, A HOLLOW REAMER HAD BREAKAGE OF SCREWS. IT IS UNKNOWN IF THERE WERE A PROCEDURE OR PATIENT INVOLVEMENT. THIS REPORT IS FOR (1) HOLLOW REAMER COMPLETE FOR 3.5MM & 4.0MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8323473 ↗MalfunctionSep 15, 2017Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2017, AND EXTRACTION BOLT FOR 4.5MM SCREWS BROKE AND THE THREAD OF A HOLLOW REAMER COMPLETE FOR 3.5MM & 4.0MM SCREWS IS BROKEN AND HAS A SCREW STUCK INSIDE OF IT. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. CONCOMITANT DEVICES REPORTED: SCREW (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) HOLLOW REAMER COMPLETE FOR 3.5MM & 4.0MM SCREWS. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6870514 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page