NA 319.701Recalls & reported problems
Synthes GmbH · CANNULATED SCREW MEASURING DEVICE · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982190321), received between Apr 18, 2018 and Oct 16, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
- Incorrect Measurement1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2020Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT DURING A STANDARD OPEN REDUCTION INTERNAL FIXATION (ORIF) OF THE DISTAL FEMUR ON (B)(6) 2020, THE CANNULATED MEASURING DEVICE WAS NOT WORKING PROPERLY. WHEN THE SURGEON USED THE RULER TO MEASURE LENGTH AFTER INSERTING THE GUIDEWIRE INTO DRILL GUIDE, THE MEASUREMENT DIDN¿T COINCIDE WITH THE MEASUREMENT OF THE BONE AND GAVE A VALUE THAT WAS SHORTER THAN EXPECTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE USE OF ANOTHER SCREW MEASURING DEVICE. THERE WAS A TWO (2) MINUTE SURGICAL DELAY REPORTED. PATIENT STATUS WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: WIRE GUIDE (PART # 324.174, LOT # UNKNOWN, QUANTITY 1), DRILL TIP GUIDEWIRE (PART # 310.243, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) CANNULATED SCREW MEASURING DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10693020 ↗MalfunctionApr 18, 2018Break; Break
IT WAS REPORTED THE CANNULATED SCREW MEASURING DEVICE WAS DISCOVERED TO BE BROKEN. NO PATIENT OR SURGICAL INVOLVEMENT WAS REPORTED. THIS REPORT IS FOR ONE (1) CANNULATED SCREW MEASURING DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7441697 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 319.701
Has the NA 319.701 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982190321), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 319.701?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 18, 2018 and Oct 16, 2020. In those reports, the most common specific device problems are “Break” (1) and “Incorrect Measurement” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 319.701?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 319.701 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 319.701, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 319.701). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.091Synthes GmbH · GAUGE, DEPTH64 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page