NA 324.202Recalls & reported problems
Synthes GmbH · 3.2MM PERCUTANEOUS THREADED DRILL GUIDE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982191199), received between Oct 13, 2021 and Apr 18, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 18, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT DURING AN OPEN REDUCTION/INTERNAL FIXATION PROCEDURE ON (B)(6) 2023, THE SURGEON WAS THREADING THE H&LE FOR PERCUTANEOUS THRD DRILL GUIDES INTO THE THREADED DRILL GUIDES, AND THE HANDLE FOR PERCUTANEOUS THREADED DRILL GUIDES WOULD THREAD BUT NOT RETAIN THE RIGIDNESS THAT THE SURGEON WAS USED TO ACHIEVING. THE SURGERY WAS DELAYED BY FOUR MINUTES DUE TO THE EVENT. A SECOND SET OF INSTRUMENTS WAS OPENED IN ORDER TO OBTAIN ANOTHER PERCUTANEOUS THREADED DRILL GUIDE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE 3.2MM PERCUTANEOUS THREADED DRILL GUIDE. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16762245 ↗MalfunctionOct 13, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, THE THREADED GUIDES FOR THE 4.5 LCP AIMING ARM HAVE BEEN REPLACED. THE 4.0MM HEX WRENCH THAT HAS TYPICALLY BEEN USED TO TIGHTEN/LOOSEN THE GUIDES FROM THE PLATE WILL NOT BE ACCEPTED NOW. UPON INSPECTION, THE DIAMETER OF THE PROXIMAL HOLES ARE SMALLER THAN THE GUIDES THAT WERE REPLACED. THIS REPORT IS FOR ONE (1) 3.2MM PERCUTANEOUS THREADED DRILL GUIDE. THIS IS REPORT 4 OF 8 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12625196 ↗MalfunctionOct 13, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, THE THREADED GUIDES FOR THE 4.5 LCP AIMING ARM HAVE BEEN REPLACED. THE 4.0MM HEX WRENCH THAT HAS TYPICALLY BEEN USED TO TIGHTEN/LOOSEN THE GUIDES FROM THE PLATE WILL NOT BE ACCEPTED NOW. UPON INSPECTION, THE DIAMETER OF THE PROXIMAL HOLES ARE SMALLER THAN THE GUIDES THAT WERE REPLACED. THIS REPORT IS FOR ONE (1) 3.2MM PERCUTANEOUS THREADED DRILL GUIDE. THIS IS REPORT 5 OF 8 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12625168 ↗MalfunctionOct 13, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, THE THREADED GUIDES FOR THE 4.5 LCP AIMING ARM HAVE BEEN REPLACED. THE 4.0MM HEX WRENCH THAT HAS TYPICALLY BEEN USED TO TIGHTEN/LOOSEN THE GUIDES FROM THE PLATE WILL NOT BE ACCEPTED NOW. UPON INSPECTION, THE DIAMETER OF THE PROXIMAL HOLES ARE SMALLER THAN THE GUIDES THAT WERE REPLACED. THIS REPORT IS FOR ONE (1) 3.2MM PERCUTANEOUS THREADED DRILL GUIDE. THIS IS REPORT 3 OF 8 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12625150 ↗MalfunctionOct 13, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, THE THREADED GUIDES FOR THE 4.5 LCP AIMING ARM HAVE BEEN REPLACED. THE 4.0MM HEX WRENCH THAT HAS TYPICALLY BEEN USED TO TIGHTEN/LOOSEN THE GUIDES FROM THE PLATE WILL NOT BE ACCEPTED NOW. UPON INSPECTION, THE DIAMETER OF THE PROXIMAL HOLES ARE SMALLER THAN THE GUIDES THAT WERE REPLACED. THIS REPORT IS FOR ONE (1) 3.2MM PERCUTANEOUS THREADED DRILL GUIDE. THIS IS REPORT 1 OF 8 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12624989 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page