DHS 338.08Recalls & reported problems
Synthes GmbH · DHS®/DCS® T-HANDLE QC · Class 1
All models of DHS (57)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982193148), received between Jun 21, 2019 and Sep 26, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility5
- Break2
- Contamination /Decontamination Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 26, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2023, THE DHS 130 DEGREE GUIDE AND T HANDLE WERE USED DURING A PROCEDURE. THE GUIDE KEPT DISENGAGING. WHILE INSERTING THE CSS+ LARGE SCREW, THE HANDLE QUIT RATCHETING AND BOTH SCREWDRIVER HEADS WERE STRIPPED. THE SCREW WAS INSERTED AND THE HEAD SNAPPED OFF DUE TO TORQUE. THERE WERE NO REPORTED PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR A DHS®/DCS® T-HANDLE QC. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17818759 ↗MalfunctionFeb 18, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, UPON ROUTINE INSPECTION WHILE RESTOCKING THE INSTRUMENT TRAY IT WAS NOTICED THAT THE T-HANDLE FOR QUICK COUPLING WOULD NOT DISENGAGE FROM THE ANGLED GUIDE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A T-HANDLE FOR QUICK COUPLING. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13559254 ↗MalfunctionFeb 25, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, BOTH DHS T-HANDLES WOULD NOT HOLD THE INSTRUMENTS AND THE SPRING MECHANISMS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS REPORT INVOLVES ONE (1) DHS®/DCS T-HANDLE QC. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11377727 ↗MalfunctionFeb 25, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, BOTH DHS T-HANDLES WOULD NOT HOLD THE INSTRUMENTS AND THE SPRING MECHANISMS WERE BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS REPORT INVOLVES ONE (1) DHS®/DCS T-HANDLE QC. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11377716 ↗MalfunctionJun 21, 2019Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A HIP SURGERY. DURING THE SURGERY, THE DYNAMIC HIP AND CONDYLAR SCREW SYSTEM (DHS/DCS) T-HANDLE DID NOT CONNECT AND WAS IMPOSSIBLE TO CLEAN PROPERLY, AND A PINNACLE STRAIGHT CUP IMPACTOR BROKE INTO PIECES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR A DHS®/DCS® T-HANDLE QC. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8721870 ↗
Other WRENCH devices
- MONARCH 277040510Medos International Sàrl · WRENCH199 reports
- PINNACLE 2217-50-008Depuy (Ireland) · WRENCH108 reports
- SIGMA 96-1673Depuy (Ireland) · WRENCH102 reports
- P.F.C. 86-5189Depuy (Ireland) · WRENCH92 reports
- Veptr 03.641.004Synthes GmbH · WRENCH86 reports
- RECLAIM 2975-00-300Depuy (Ireland) · WRENCH51 reports
- PINNACLE 2217-50-006Depuy (Ireland) · WRENCH34 reports
- NA 388.261Synthes GmbH · WRENCH32 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page