Skip to content

NA 357.370Recalls & reported problems

Synthes GmbH · RETAINING RING FOR 357.369 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982196040), received between Apr 3, 2019 and Apr 3, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2019Break; Break

    IT WAS REPORTED ON AN UNKNOWN DATE, THE (TROCHANTERIC FIXATION NAIL) TFN HELICAL BLADE INSERTER ATTACHMENT SCREW BROKE WHILE INSERTING THE HELICAL BLADE. REPORTEDLY, NUT FELL OFF THE INSTERTER CAUSING IT TO COME APART AND HAD TO MALLET IT TO GET IT DISCONNECTED FROM HELICAL BLADE. IT WAS NOTED ALSO THAT WHILE INSERTING THE HELICAL BLADE, THE END OF THE HELICAL BLADE COUPLING SCREW BROKE OFF LEAVING THE SHAFT IN THE INSERTER STILL ATTACHED TO THE HELICAL BLADE. THE HELICAL BLADE WAS SUCCESSFULLY INSERTED AND IT TOOK 20 MINUTES TO TAKE OFF THE INSERTER. FINALLY, THE BOLT CUTTERS WERE USED TO CUT THE SHAFT ABOVE WHERE IT WAS ATTACHED TO THE HELICAL BLADE. THE AIMING ARM WAS THEN ATTACHED BACK AND DISTALLY LOCKED TO FINISH THE CASE. WHILE MALLETING THE HELICAL BLADE, THE END OF THE COUPLING SCREW BROKE LEAVING THE SHAFT OF THE CONNECTING SCREW IN THE HELICAL BLADE INSERTER. THE FOLLOWING DEVICES WERE DAMAGED WHEN TRYING/ATTEMPTING TO REMOVE THE INSERTER: ONE (1) BUTTRESS/COMPRESSION NUT, ONE (1) BLADE GUIDE SLEEVE, AND ONE (1) AIMING ARM. THERE WERE FRAGMENTS GENERATED FROM A BROKEN DEVICE. THE FRAGMENTS WERE REMOVED EASILY WITHOUT ANY ADDITIONAL INTERVENTION. THERE WERE THIRTY (30) MI

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8479621 ↗

Other INSTRUMENT, COMPRESSION devices

All INSTRUMENT, COMPRESSION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page