Skip to content

NA 357.411Recalls & reported problems

Synthes GmbH · INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10886982196347), received between Jan 17, 2017 and Feb 7, 2022 — classified as 3 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility6
  • Failure to Align5
  • Bent2
  • Device Operates Differently Than Expected1
  • Sticking1

Patient problems

  • No Known Impact Or Consequence To Patient7
  • No Consequences Or Impact To Patient4
  • No Code Available2
  • Failure of Implant1
  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • Not Applicable1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 7, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE A PATIENT UNDERWENT THE SURGERY WITH A SHORT TITANIUM TROCHANTERIC FIXATION NAIL SYSTEM (TFN) TO TREAT A HIP FRACTURE. DURING THE PROCEDURE WHEN SURGEON TRIED TO INSERT THE DISTAL LOCKING SCREW IT MISSED THE NAIL. HE REMOVED THE NAIL AND RECONNECTED THE INSTRUMENTS BUT IT CONTINUED TO MISS THE NAIL. THE SURGEON FINALLY INSERTED THE SCREW WITH A FREEHAND TECHNIQUE. THE SURGERY WAS DELAYED FOR FIVE (5) MINUTES DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS. THIS IS REPORT 3 OF 5 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13476815 ↗
  • MalfunctionFeb 21, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2020, A INCIDENT WAS REPORTED BY THE SURGEON THAT INVOLVED A TROCHANTERIC FIXATION NAIL (TFN) INSERTION HANDLE AND CONNECTING SCREWS. WHILE CONNECTING THE NAIL, THE SCREW ENGAGED IN THE BACK OF THE NAIL, HOWEVER, THE NAIL ITSELF TUSSLED. IT WAS TIGHTENED. THE SURGEON ATTEMPTED TO INSERT IT. THE DEVICE STILL FELT LOOSE AND BACKED IT OUT, RE-TIGHTENED IT BUT NAIL STILL TUSSLED. THE SURGERY WAS DELAYED WHILE NAIL WAS ATTEMPTED TO BE TIGHTEN AGAIN. SURGICAL DELAY TIME IS UNKNOWN. THE PROCEDURE AND PATIENT OUTCOMES ARE UNKNOWN. THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9739273 ↗
  • MalfunctionDec 13, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A TROCHANTERIC FIXATION NAIL (TFN) INSERTION ON (B)(6) 2019, THE TARGETING DEVICE MISSED THE NAIL. THE DISTAL SCREW DID NOT GO THROUGH THE NAIL, IT WENT ANTERIOR TO THE NAIL. THE SURGEON FREE-HAND LOCKED THE NAIL. THERE WAS A SURGICAL DELAY OF TWENTY (20) MINUTES. PATIENT STATUS WAS STABLE. CONCOMITANT DEVICES: TFN NAIL (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1), DISTAL SCREW (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A INSERTION HANDLE FOR TROCHANTERIC FIXATION NAILS. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9468394 ↗
  • InjurySep 23, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2019, DURING TROCHANTERIC NAIL PROCEDURE, THE INSERTION HANDLE & CANNULATED CONNECTING SCREW WOULD NOT DISENGAGE FROM THE NAIL. ALLEGATIONS OF DEVICES INSERTION HANDLE AND CONNECTING SCREW WOULD NOT DISENGAGE FROM NAIL. BROKE WRENCH AND TWO (357.397 AND ONE 521.30) BALL HEX SCREWDRIVERS TRYING TO REMOVE INSERTION HANDLE FROM NAIL. FRAGMENTS WERE GENERATED FROM BROKEN DEVICES. SURGEON ENDED UP REMOVING THE ENTIRE CONSTRUCT AND REDID THE OPERATION WITH A DIFFERENT SET AND NEW NAIL. OTHER MEDICAL INTERVENTION WAS TO REMOVE IMPLANTED PRODUCT AND START PROCEDURE OVER WITH NEW SETS AND IMPLANTS. PROCEDURE ENDED UP BEING DONE SUCCESSFULLY WITH NEW SYSTEM AND NAIL SURGERY WAS DELAYED BY 45 MINUTES WITH NO LESS DESIRABLE OUTCOME. CONCOMITANT DEVICE REPORTED: UNK - NAILS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS IS REPORT 1 FOR 4 (B)(4).

    Patient problems: Failure of Implant; No Code Available; Not Applicable

    MAUDE report 9106873 ↗
  • InjurySep 10, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2019, DURING INSERTION OF CANNULATED TROCHANTERIC FIXATION NAIL (TFN), DIFFICULTY WAS ENCOUNTERED UPON DRILLING THROUGH THE 130-DEGREE AIMING ARM. THE DRILL BIT COLLIDED WITH THE TFN IMPLANT. THE RESULT WAS ADDITIONAL DRILL HOLES IN THE FEMUR. THE PROCEDURE WAS EXTENDED APPROXIMATELY 20 MINUTES. ADDITIONAL C-ARM TIME WAS REQUIRED. THE LOCKING SCREW WAS SUCCESSFULLY INSERTED IN THE LOCKING OPTION. THE INNER DRILL SLEEVES WERE REMOVED, AND THE DRILL BIT WAS THEN USED TO DRILL THROUGH BOTH CORTICES IN A FREEHAND TECHNIQUE. THERE WERE NO FRAGMENTS GENERATED. THERE WAS A SURGICAL DELAY APPROXIMATELY TWENTY (20) MINUTES. PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN LOCKING SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS IS REPORT 1 FOR (B)(4).

    Patient problems: Injury

    MAUDE report 8983851 ↗
  • InjuryAug 21, 2018Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT REVISION ON (B)(6) 2018 DUE TO A TITANIUM TROCHANTERIC FIXATION NAIL (TFN) DISTAL SCREW MISSING THE NAIL. ORIGINAL HARDWARE WAS IMPLANTED ON (B)(6) 2018 TO TREAT A REVERSE OBLIQUE HIP FRACTURE; IT WAS NOTED DURING THAT PROCEDURE THERE WAS A TWO-MINUTE DELAY. DURING THAT PROCEDURE, ONLY ANTERIOR X-RAYS WERE TAKEN; LATERAL X-RAYS WERE NOT DONE. X-RAYS TAKEN ON (B)(6) 2018 DISCOVERED THAT THE SCREW HAD MISSED THE NAIL; THE DISTAL SCREW WAS LODGED IN THE PATIENT¿S BONE. A POSSIBLE MISALIGNMENT OF THE INSTRUMENTATION MAY HAVE CAUSED THE DISTAL SCREW TO BE MISGUIDED AND MISS THE NAIL. THERE WAS NO REPORTED ISSUE WITH THE HELICAL BLADE. THE INSTRUMENTATION THAT MAY HAVE CONTRIBUTED TO THE MISALIGNMENT, CONSISTED OF AN INSERTION HANDLE, AIMING ARM, CONNECTING SCREW AND CANNULA. UPON REVIEWING THE X-RAYS ON (B)(6) 2018, IT WAS ALSO NOTED A SMALL HAIRLINE FRACTURE OF THE FEMUR BONE WAS PRESENT AT THE BOTTOM OF THE NAIL AND THE PATIENT REPORTED PAIN. PATIENT UNDERWENT REVISION SURGERY THE SAME DAY TO REMOVE THE SHORT NAIL ALONG WITH THE HELICAL BLADE AND DISTAL SCREW THAT WAS LODGED IN THE PATIENT¿S BONE. PATIENT WAS REVISED TO A LONGER NAIL, ANOTHER HEL

    Patient problems: No Code Available

    MAUDE report 7801357 ↗
  • MalfunctionApr 19, 2018Bent; Bent

    (B)(6) REPORTED AS FOLLOW: IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING THE SURGERY, INSERTION HANDLE FOR TROCHANTERIC FIXATION NAIL WAS BENT OR DAMAGED AND DID NOT ALLOW THE AIMING ARM TO ALIGN WITH THE HOLE IN THE NAIL. THERE WERE NO FRAGMENTS GENERATED, AND THERE ARE NO MISSING PIECES. THE SURGERY WAS COMPLETED USING THE DEVICE, WITH NO REPORTED PATIENT HARM. CONCOMITANT DEVICE REPORTED: TFN AIMING ARM (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) INSERTION HANDLE FOR TROCHANTERIC FIXATION NAILS. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7443037 ↗
  • MalfunctionMar 22, 2018Failure to Align; Failure to Align

    IT WAS REPORTED THAT DURING AN INITIAL IMPLANT OF A TROCHANTERIC FIXATION NAIL (TFN) SYSTEM ON (B)(6) 2018, ONE (1) DISTAL SCREW MISSED THE TFNA NAIL. WHILE DRILLING FOR THE DISTAL SCREW, THE DRILL BIT HIT THE TFN NAIL AND A SOUND WAS HEARD. THE SURGEON CONTINUED TO DRILL AT THIS POINT AND THEN INSERTED THE DISTAL SCREW INTO THE NAIL USING AN AIMING ARM, INSERTION HANDLE AND A CANNULATED CONNECTING SCREW. UPON TAKING A LATERAL RADIOGRAPHIC IMAGE, THE SURGEON REALIZED THAT THE DISTAL SCREW HAD MISSED THE NAIL AND WAS VISIBLE TO BE POSTERIOR TO THE NAIL IN THIS LATERAL IMAGE. THE SURGEON THEN REPORTEDLY MANUALLY INSERTED THE DISTAL SCREW INTO POSITION. THERE WAS NO VISIBLE MALFUNCTION ON ANY OF THE INSTRUMENTS. CONCOMITANT DEVICES REPORTED: 10MM 130 DEGREE CANNULATED TFN NAIL 170MM (PART 456.315S, LOT NUMBER UNKNOWN, QUANTITY 1), CANNULATED CONNECTING SCREW FOR TFN NAIL (PART 357.397, LOT 6365450, QUANTITY 1) THIS REPORT IS FOR ONE (1) INSERTION HANDLE FOR TFN NAILS THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7363231 ↗
  • MalfunctionFeb 2, 2018Failure to Align; Failure to Align

    CANADA REPORTED THAT THE INSERTION HANDLE FOR TROCHANTERIC FIXATION NAIL IS BENT OR DAMAGED AND DOES NOT ALLOW THE AIMING ARM TO ALIGN WITH THE HOLE IN THE NAIL. THERE WERE NO FRAGMENTS GENERATED, AND THERE ARE NO MISSING PIECES. THE SURGERY WAS COMPLETED USING THE DEVICE, WITH NO REPORTED PATIENT HARM. CONCOMITANT DEVICE REPORTED: UNKNOWN TFN AIMING ARM (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7238358 ↗
  • MalfunctionFeb 1, 2018Failure to Align; Failure to Align

    (B)(6) REPORTED AS FOLLOW: IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING THE SURGERY, INSERTION HANDLE FOR TROCHANTERIC FIXATION NAIL WAS BENT OR DAMAGED AND DID NOT ALLOW THE AIMING ARM TO ALIGN WITH THE HOLE IN THE NAIL. THERE WERE NO FRAGMENTS GENERATED, AND THERE ARE NO MISSING PIECES. THE SURGERY WAS COMPLETED USING THE DEVICE, WITH NO REPORTED PATIENT HARM. CONCOMITANT DEVICE REPORTED: TFN AIMING ARM (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7234158 ↗
  • MalfunctionOct 5, 2017Sticking; Sticking

    IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTY REMOVING THE INSERTION HANDLE AND CONNECTING SCREW FROM THE TROCHANTERIC FIXATION NAIL (TFN) NAIL DURING A PROCEDURE ON (B)(6) 2017. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO DELAY AND PATIENT OUTCOME WAS STABLE. THIS REPORT INVOLVES 2 DEVICES. CONCOMITANT DEVICES REPORTED: TROCHANTERIC FIXATION NAIL (TFN) (PART #UNKNOWN, LOT # UNKNOWN, QUANTITY #UNKNOWN). THIS REPORT IS 2 OF 2 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6916596 ↗
  • MalfunctionAug 18, 2017Failure to Align; Failure to Align

    IT WAS REPORTED THAT DURING A TROCHANTERIC FIXATION NAIL (TFN) SURGERY ON (B)(6) 2017, THE 130 DEGREE AIMING ARM AND THE INSERTION HANDLE DID NOT LINE UP PROPERLY WHEN INSERTING THE DISTAL LOCKING SCREW. THE NAIL AND THE HELICAL BLADE WERE BOTH ALREADY IN PLACE. THE SURGEON ADJUSTED AND TIGHTENED THE AIMING ARM, HANDLE AND THE DRIVING CAP, AND WAS THEN ABLE TO USE THE INSTRUMENTS TO IMPLANT THE DISTAL SCREW AS INTENDED. THERE WERE NO FRAGMENTS GENERATED, NO SURGICAL DELAY, AND NO ADDITIONAL MEDICAL INTERVENTION. THE SURGERY WAS COMPLETED, AND THE PATIENT OUTCOME WAS STABLE. CONCOMITANT DEVICES REPORTED: TFN NAIL (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1); HELICAL BLADE (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1); DISTAL LOCKING SCREW (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1); DRIVING CAP (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) INSERTION HANDLE (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6804893 ↗
  • MalfunctionJul 13, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED THAT THE DISTAL TARGETING FOR A 5MM X 36MM LOCKING SCREW MISSED THE NAIL DURING A TROCHANTERIC FIXATION NAIL (TFN) PROCEDURE ON (B)(6) 2017 DUE TO ISSUE WITH GETTING RIGHT TRAJECTORY USING INSERTION HANDLE AND AIMING ARM. THERE WAS NO ALLEGATION AGAINST LOCKING SCREW AND NAIL. THE SURGEON MADE A SECOND ATTEMPT AND FREE-HANDED THE DRILL BIT AND WAS ABLE TO GET THE RIGHT TRAJECTORY USING INSERTION HANDLE AND AIMING ARM TO GET THE SCREW IN. THERE WAS A REPORTED TEN MINUTE SURGICAL DELAY FROM THE SECOND ATTEMPT TO GET CONSTRUCT LINED UP CORRECTLY. NO ADDITIONAL X-RAYS OR OTHER MEDICAL INTERVENTION REQUIRED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE PATIENT IN STABLE CONDITION. CONCOMITANT DEVICES REPORTED: 5.0MM TI LOCKING SCREW 36MM- FOR IM NAILS (PART # 458.936, LOT # UNKNOWN, QUANTITY # 1), 4.0MM THREE-FLUTED DRILL BIT QC/260MM/65MM CALIBRATION (PART # 357.407, LOT # UNKNOWN, QUANTITY # 1), 10MM SHORT TROCHANTERIC FIXATION NAIL (10X170MM) (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6710469 ↗
  • MalfunctionJul 10, 2017Failure to Align; Failure to Align

    DEVICE REPORT FROM SYNTHES (B)(4) REPORTED THE FOLLOWING: IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INSERTION OF A HELICAL BLADE DURING A TROCHANTERIC FIXATION NAIL (TFN) PROCEDURE ON (B)(6) 2017 TO TREAT A PROXIMAL FEMUR FRACTURE. DURING THE PROCEDURE THE HELICAL INSERTION COUPLING DID NOT ALIGN PROPERLY MAKING IT HARD TO INSERT THE BLADE CORRECTLY. THERE WAS A TEN (10) MINUTE DELAY RELATED TO THIS EVENT. NO INFORMATION AVAILABLE ON WHETHER THE PROCEDURE WAS COMPLETED. NO HARM WAS REPORTED TO THE PATIENT. NO INFORMATION AVAILABLE ON PATIENT OUTCOME. CONCOMITANT DEVICE REPORTED: HELICAL BLADE INSERTION SCREW (PART 357.369, LOT 6581164, QUANTITY 1); COMPRESSION NUT (PART 357.371, LOT 6599561, QUANTITY 1); HELICAL BLADE COUPLING SCREW (UNKNOWN PART AND LOT NUMBERS, QUANTITY 1). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6696460 ↗
  • MalfunctionJan 17, 2017Bent; Bent

    IT WAS REPORTED THAT AN INITIAL TROCHANTERIC FIXATION NAIL (TFN) PROCEDURE FOR THE LEFT FEMUR TOOK PLACE ON (B)(6) 2016 WHERE SEVERAL DEVICE MALFUNCTIONS OCCURRED. THE HELICAL BLADE COUPLING SCREW BROKE INTO TWO PIECES WHILE BEING HAMMERED BY AN UNKNOWN MALLET. THOSE PIECES WERE RETRIEVED WITH DIFFICULTY AND UNSPECIFIED ADDITIONAL INTERVENTION RESULTING IN A THIRTY (30) MINUTE SURGICAL DELAY. ADDITIONALLY, AS THE HELICAL BLADE INSERTER WAS REMOVED IT WAS NOTED THE TIP WAS BROKEN IN PIECES AS A RESULT FROM THE HAMMERING. ALL PIECES WERE RETRIEVED FROM THE BROKEN INSERTER. IT WAS ALSO DETERMINED THE THREADS OF THE BLADE GUIDE SLEEVE BECAME DAMAGED WHILE THE SURGEON WAS TRYING TO REMOVE THE HELICAL BLADE INSERTER. THE BUTTRESS COMPRESSION NUT FOR THE BLADE GUIDE SLEEVE COULD NO LONGER BE TURNED, AND WAS NO LONGER USABLE BECAUSE OF THE DAMAGE TO THE THREADS. THIS PUSHED THE BLADE GUIDE SLEEVE INTO THE 130 DEGREE AIMING ARM. THE BUTTRESS COMPRESSION NUT AND THE BLADE GUIDE SLEEVE BECAME STUCK TOGETHER. DURING THE IMPLANTATION OF THE HELICAL BLADE AND THE NAIL THE SURGEON REPORTED DIFFICULTY IN IMPLANTING THE TWO DEVICES. A PAIR OF NON-SYNTHES PLIERS WAS USED TO REMOVE THE HELICAL BLADE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6255726 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates