NA 390.051Recalls & reported problems

Synthes GmbH · 4.0MM ADJUSTABLE CLAMP FOR DISTAL RADIUS FIXATOR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982198082), received between Sep 28, 2020 and Aug 27, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

2 other recalls cover devices cleared under the same FDA submission (K031973, K033158, K090658, K963618); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Device-Device Incompatibility2
  • Device Slipped1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2025Break; Break

    IT WAS REPORTED THAT ON (B)(6), 2025 THE DEVICE 4.0/3.0 SELF-DRILL SCHANZ SCR 20 THRD/80, WOULD NOT ENGAGE IN ADAPTER, THE 4.0 ADJUSTABLE CLAMP DSTL RAD FIXATOR, CLAMP BROKE AND THE SDDRIVE SCREWDRIVER SHAFT T8 105MM TIP SNAPPED OFF. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22908647 ↗
  • MalfunctionJun 12, 2025Device Slipped; Device Slipped

    IT WAS REPORTED THAT THE SCREW WON'T TURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22192658 ↗
  • MalfunctionDec 13, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE WRENCH IS BROKEN AND CANNOT BE REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 20924331 ↗
  • MalfunctionSep 17, 2024Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE SCREW CAME OFF OF THE CLAMP. THIS REPORT IS FOR A 4.0MM ADJUSTABLE CLAMP FOR DISTAL RADIUS FIXATOR.

    Patient problems: Insufficient Information

    MAUDE report 20237370 ↗
  • MalfunctionOct 26, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE WAS DIFFICULT TO MOVE. NO FURTHER INFORMATION IS AVAILABLE.THIS REPORT IS FOR ONE (1) 4.0MM ADJUSTABLE CLAMP FOR DISTAL RADIUS FIXATOR.THIS IS REPORT 1 OF 1 FOR COMPLAINT PC (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18016320 ↗
  • MalfunctionSep 28, 2020Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT THE EXTERNAL SKELETAL FIXATION FOR DISTAL RADIUS FRACTURES BY USING THE CLAMP IN QUESTION. DURING THE SURGERY, WHEN THE SURGEON WAS SETTING UP THE CLAMP, HE COULD NOT TIGHTEN THE SCREW OF THE CONNECTION PART OF THE CLAMP. BEFORE SETTING UP THE CLAMP, HE CONFIRMED THAT THE SCREW OF THE CLAMP COULD BE TIGHTENED, AND THE SCREW PART WORKED. HOWEVER, AFTER SETTING UP THE CLAMP AFTER SELDRILL WAS INSERTED, HE REALIZED THAT THE SCREW OF THE CLAMP PLACED AT PROXIMAL END COULD NOT BE TIGHTENED. HE TRIED TO TIGHTEN THE SCREW OF THE CLAMP BY USING A DRIVER SEVERAL TIMES AND HE COULD NOT TIGHTEN IT WHILE THE SCREW WAS ROTATING IDLE. ANOTHER CLAMP WAS NOT AVAILABLE, SO THE SURGEON USED A MEDIUM EXTERNAL FIXATOR TO PLACE THE CLAMP FOR FIXATION WHILE HE WAS ENGAGING THE ALREADY INSERTED SELDRILL WITH THE CLAMP. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH A LESS-THAN-30-MINUTE DELAY. NO FURTHER INFORMATION IS AVAILABLE. CONCOMITANT DEVICES REPORTED: UNKNOWN SELDRILL SCHANZ SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1), UNKNOWN DRIVER (PART # UNKNOWN, LOT # UNKNOWN, QU

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10597018 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K031973, K033158, K090658, K963618. One submission often covers many models, so check the product and lot details in each record.

  • Z-2451-2015 · Jul 29, 2015 · Terminated
    Synthes Trauma Ex-Fix System (Small, Medium, Distraction Osteogenesis (DO) and Large) ; (S) intended to stabilize and provide treatment for fractures of the small bones, such as th
  • Z-1833-2014 · Apr 10, 2014 · Terminated
    Synthes External Fixation Systems (Small, Medium, Large and DO), bone fixation appliance components.

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page