NA 399.48Recalls & reported problems
Synthes GmbH · PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE · Class 1
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10886982202833), received between Oct 19, 2017 and Oct 21, 2024 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Break” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break9
- Contamination /Decontamination Problem4
- Crack1
- Material Discolored1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient4
- No Patient Involvement4
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT ON (B)(6) 2024 THE WOODEN HANDLE IS LEECHING. THERE WAS OPERATION INVOLVED. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20493645 ↗MalfunctionDec 20, 2022Nonstandard Device; Nonstandard Device
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE SCREW HANDLE ARRIVED WITHOUT A PART AND THE DRILL BITS WERE VERY WORN. THERE WAS NO REPORTED PATIENT IMPACT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16015430 ↗MalfunctionSep 1, 2022Break; Break
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE DEVICE WAS OPENED AND IT WAS FOUND THAT THE HANDLE OF THE DISSECTOR WAS SPLIT IN HALF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THE PATIENT OUTCOME IS OK. THIS REPORT IS FOR A PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15340343 ↗MalfunctionAug 22, 2022Break; Break
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE HANDLE IN QUESTION WAS FOUND TO BE BROKEN. THERE WAS NO REPORTED PATIENT INTERACTION, NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15266973 ↗MalfunctionAug 1, 2022Break; Break
VETERINARY COMPLAINT: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING VETERINARY TPLO PROCEDURE THE HANDLE BROKE. NO FRAGMENTS WERE GENERATED. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) PERIOSTEAL ELEVATOR 3 CRVD BLD-STR EDGE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15137867 ↗MalfunctionJun 3, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, DURING STERILIZATION WAS DISCOVERED THAT PERIOSTEAL ELEVATOR 3MM HANDLE WAS BROKEN. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11929398 ↗MalfunctionNov 24, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CUSTOMER BOUGHT 2 PERIOSTEAL ELEVATORS. WHEN THEY RECEIVED THE INSTRUMENTS, BOTH INSTRUMENTS SEEMED TO HAVE BEEN SANDBLASTED ON THE HANDLE AND THE STERILIZATION DEPARTMENT REFUSED TO PROCESS THEM. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR 1 PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10889712 ↗MalfunctionNov 20, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE CUSTOMER BOUGHT TWO (2) PERIOSTEAL ELEVATORS. WHEN THEY RECEIVED THE INSTRUMENTS, BOTH INSTRUMENTS SEEMED TO HAVE BEEN SANDBLASTED ON THE HANDLE AND THE STERILIZATION DEPARTMENT REFUSED TO PROCESS THEM. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10877261 ↗MalfunctionJun 2, 2020Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PERIOSTEAL ELEVATOR 3MM CURVED BLADE- STRAIGHT EDGE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES 1 DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10111627 ↗MalfunctionMay 26, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING INSPECTION, A BENDING PLIERS WOULD NOT OPEN DUE TO RUST EMBEDDED IN THE GROOVES AND SPRING. THE THREADS OF AN EXTRACTION SCREW WERE STRIPPING, A STARDRIVE SCREWDRIVER SHAFT TIP IS WARPING DUE TO WEAR, 2 PERIOSTEAL ELEVATOR (ROUND AND STRAIGHT) HAVE STAINING BETWEEN THE WOODEN HANDLE AND ELEVATOR, A GRAPHIC CASE TOP COAT OF AUX BIN IS STARTING TO PEEL, AND THE DEPTH GAUGE HANDLE HAS A CRACK THROUGH IT. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THIS REPORT IS FOR 1 PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10088423 ↗MalfunctionSep 27, 2019Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT PERIOSTEAL ELEVATOR 3 MM CURVED BLADE-STRAIGHT EDGE IS BROKEN. NO OTHER INFORMATION HAS BEEN PROVIDED. THIS REPORT IS FOR ONE (1) PERIOSTEAL ELEVATOR 3 MM CURVED BLADE-STRAIGHT EDGE THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9131611 ↗MalfunctionAug 12, 2019Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A ROUTINE INCOMING INSPECTION OF A LOANER SET, THE PERIOSTEAL ELEVATOR 3 MM CURVED BLADE-STRAIGHT EDGE WAS OBSERVED BROKEN. THERE WAS NO KNOWN HOSPITAL OR PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8885941 ↗MalfunctionAug 15, 2018Break; Crack; Break; Crack
IT WAS REPORTED THAT ON (B)(6) 2018, THE CENTRAL PROCESSING MANAGER PRESENTED ONE (1) PERIOSTEAL ELEVATOR CURVED BLADE-ROUND EDGE, ONE (1) PERIOSTEAL ELEVATOR CURVED BLADE-STRAIGHT EDGE, ONE (1) CRUCIFORM SCREWDRIVER, AND ONE (1) REDUCTION FORCEPS WITH POINTS RATCHET THAT WERE CRACKED OR BROKEN. IT WAS UNKNOWN WHEN THE DEVICES BECAME DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) PERIOSTEAL ELEVATOR 3MM CURVED BLADE - STRAIGHT EDGE. THIS IS REPORT 2 OF 4 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7784457 ↗MalfunctionJan 5, 2018Material Discolored; Material Discolored
IT WAS REPORTED THAT SEVERAL INSTRUMENTS AT THE FACILITY ARE DISCOLORED. THE CANNULATED 4.0MM HEXAGONAL SCREWDRIVER, THE PERIOSTEAL ELEVATOR 3MM CURVED BLADE (STRAIGHT EDGE), THE PERIOSTEAL ELEVATOR 3MM CURVED BLADE (ROUND EDGE), AND THE HOLDING SLEEVE. WHEN PEEL PACKED, THE HANDLES ARE LEACHING A BROWN SUBSTANCE ON THE THE BACKING OF THE PEEL PACK. THE HANDLES HAVE A DULL, ROUGH FINISH. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR A CURVED BLADE (STRAIGHT EDGE) ELEVATOR. THIS REPORT IS 2 OF 4 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7167581 ↗MalfunctionOct 19, 2017Break; Break
IT WAS REPORTED THAT THE SERVICE AND REPAIR DEPARTMENT DOCUMENTED THAT A PERIOSTEAL ELEVATOR 3MM CURVED BLADE-STRAIGHT EDGE HAS A BROKEN HANDLE. THERE IS NO PATIENT/PROCEDURE INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 6963520 ↗
Other ELEVATOR devices
- NA 399.36Synthes GmbH · ELEVATOR15 reports
- NA 399.40Synthes GmbH · ELEVATOR9 reports
- NA 399.481Synthes GmbH · ELEVATOR8 reports
- NA 03.133.202Synthes GmbH · ELEVATOR3 reports
- NA 399.37Synthes GmbH · ELEVATOR3 reports
- NA 389.56Synthes GmbH · ELEVATOR2 reports
- Innomed, Inc. 3450Innomed, Inc. · Elevator2 reports
- NA SD399.035Synthes GmbH · ELEVATOR1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.