NA 399.55Recalls & reported problems
Synthes GmbH · CHISEL BLADE 10MM · Class 1
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886982202901), received between Oct 8, 2018 and Jan 16, 2019 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break10
- Failure to Cut1
- Material Twisted/Bent1
Patient problems
- No Patient Involvement10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2019Break; Failure to Cut; Material Twisted/Bent; Break; Failure to Cut; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, ONE (1) CHISEL BLADE W/ 10, ONE (1) CHISEL BLADE W/ 16 AND ONE (1) PERIOSTEAL ELEVATOR 6MM CURVED BLADE - STRAIGHT EDGE WERE MARKED AS BROKEN IN MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD). THE TIP OF THE TWO (2) CHISEL BLADES WERE UNEVEN AND DULL. THE TIP OF THE PERIOSTEAL ELEVATOR WAS DINGED AND DULL. THERE WAS NO PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) CHISEL BLADE 10MM. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8254988 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTY-EIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTY-EIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 8 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8039001 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTY-EIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTY-EIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 7 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038991 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTY-EIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTY-EIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 6 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038955 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTYEIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTYEIGHT (38) DEVICES ARE ADDRESSED IN PC-(B)(4) (10), PC-(B)(4) (10), PC-(B)(4) (10), AND PC-(B)(4) (8). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 5 OF 10 FOR PC-(B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038945 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTY-EIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTY-EIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 4 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038910 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTYEIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTYEIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 3 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038887 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTY-EIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTY-EIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 2 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038826 ↗MalfunctionNov 5, 2018Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THIRTYEIGHT (38) CHISEL BLADES WERE FOUND IN THE MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD) AND TAGGED AS BROKEN. THERE WAS NO PATIENT OR SURGICAL INVOLVEMENT. THE THIRTYEIGHT (38) DEVICES ARE ADDRESSED IN (B)(4). THIS REPORT IS FOR ONE (1) 10MM CHISEL BLADE. THIS IS REPORT 1 OF 10 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8038802 ↗MalfunctionOct 8, 2018Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE CONNECTING SCREW AND CHISEL BLADE WERE TAGGED AS BROKEN AT MEDICAL DEVICE REPROCESSING DEPARTMENT (MDRD). THERE WERE NO PATIENT AND SURGICAL INVOLVEMENT REPORTED. THIS REPORT IS FOR ONE (1) CHISEL BLADE 10MM. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 7945057 ↗
Other CHISEL, SURGICAL, MANUAL devices
- SIGMA 2024-85-000Depuy (Ireland) · CHISEL, SURGICAL, MANUAL63 reports
- NA 2421-01-000Depuy (Ireland) · CHISEL, SURGICAL, MANUAL36 reports
- NA 399.56Synthes GmbH · CHISEL, SURGICAL, MANUAL11 reports
- T.L.I.F. 274279000Depuy Spine, LLC · CHISEL, SURGICAL, MANUAL7 reports
- NA 332.121Synthes GmbH · CHISEL, SURGICAL, MANUAL5 reports
- INSTRUMENT 6541-4-709Howmedica Osteonics Corp. · CHISEL, SURGICAL, MANUAL4 reports
- NA 202953Depuy Spine, LLC · CHISEL, SURGICAL, MANUAL4 reports
- GENESIS II 71440144Smith & Nephew, Inc. · CHISEL, SURGICAL, MANUAL3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.