NA 511.902Recalls & reported problems

Synthes GmbH · RECIPROCATING SAW ATTACHMENT · Class 1

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982238078), received between Jul 17, 2020 and Jul 22, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Loss of Power3
  • Mechanical Jam3
  • Excessive Heating2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2024Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED FROM NETHERLANDS THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATOR SAW ATTACHMENT DEVICE WAS FROZEN (WOULD NOT MOVE). IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETESTS FOR CHECK FUNCTION IN RUNNING MODE AND CHECK THE OSCILLATION FREQUENCY WITH FREQUENCY METER. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DOES NOT WORK AT ALL DURING ROUTINE MAINTENANCE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19793180 ↗
  • MalfunctionOct 5, 2023Excessive Heating; Excessive Heating

    IT WAS REPORTED FROM NETHERLANDS THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE RECIPROCATING SAW ATTACHMENT DEVICE HEATED UP, WAS VIBRATING AND HE MOVING PARTS DID NOT MOVE SMOOTHLY. THE DEVICE ALSO FAILED PRETESTS FOR FUNCTION IN RUNNING MODE AND OSCILLATION FREQUENCY WITH FREQUENCY METER. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17876476 ↗
  • MalfunctionApr 26, 2023Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED BY RUSSIA THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATING SAW ATTACHMENT DEVICE WAS FROZEN/WOULD NOT MOVE. IT WAS OBSERVED THAT THE DEVICE HAD A COSMETIC DAMAGE ¿ WORN ETCH, AND THE LABELING HAD AN ILLEGIBLE ETCH. IT WAS FURTHER DETERMINED THAT THE DEVICE COULD NOT ENGAGE THE ATTACHMENT ¿ COUPLING. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR GENERAL CONDITION, MARKING AND LABELING, CHECK THE HANDPIECE COUPLING, CHECK SYMMETRY WITH HANDPIECE, CHECK FUNCTION IN RUNNING MODE AND CHECK THE OSCILLATION FREQUENCY WITH FREQUENCY METER. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. IT WAS REPORTED THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AND THE SURGERY WAS COMPLETED AS INTENDED. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16815768 ↗
  • MalfunctionApr 22, 2022Mechanical Jam; Complete Loss of Power; Mechanical Jam; Complete Loss of Power

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE WAS FROZEN/WOULD NOT MOVE, AND DID NOT FUNCTION. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR GENERAL CONDITION, CHECK THE HANDPIECE COUPLING, CHECK SYMMETRY WITH THE HANDPIECE, CHECK FUNCTION IN RUNNING MODE, AND CHECK OSCILLATION FREQUENCY. IT WAS NOTED IN THE SERVICE ORDER THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT WHEN THE DEVICE MADE CONTACT WITH BONE, IT WOULD LOSE STRENGTH. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN (B)(6) OF 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14188359 ↗
  • MalfunctionJul 17, 2020Excessive Heating; Excessive Heating

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE BEARING OF THE ATTACHMENT DEVICE WAS WORN OUT AND CORRODED, THE MOVING PARTS OF THE DEVICE DID NOT MOVE SMOOTHLY, AND THERE WAS HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR HEAT, AND CHECK OSCILLATION FREQUENCY. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10289927 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.